Infrared Gas Analyzer
Operation
Teledyne
Analytical
Instruments
75
To close the “Setting of Auto Calibration” or cancel the mode
midway, press the ESC key. A previous screen will return.
Settings range:
Hold setting:
0 to 100% FS
Response time: 1 to 60 sec. (Initial value: 15 sec)
Average period: 1 to 59 min or 1 to 4 hours (Initial value: 1 hour)
Backlight Timer: 1 to 60 min (Initial value: OFF)
Maintenance mode: 0000 to 9999 (Initial value: 0000)
3.5.7.1
O
UTPUT
H
OLD
By setting an output hold to ON, the output signal of each channel is
held during the calibration (manual calibration and auto calibration) plus
the gas flow time as shown in Figures 3-10 through 3-12. Refer to Section
3.5.4,
Setting of Auto Calibration
for a description of the gas flow time.
Regardless of Hold ON/OFF setting, an output signal can always be held
via an external input.
Manual calibration
Figure 3-10: Output Hold for Manual Calibration
Summary of Contents for 7600
Page 14: ...Model 7600 Teledyne Analytical Instruments xiv ...
Page 20: ...Introduction Model 7600 Teledyne Analytical Instruments 20 ...
Page 96: ...Maintenance Model 7600 Teledyne Analytical Instruments 96 Figure 4 3 Sample Cell ...
Page 108: ...Maintenance Model 7600 Teledyne Analytical Instruments 108 ...
Page 120: ...Appendix Model 7600 Teledyne Analytical Instruments 120 A 9 Description on Terminal Block ...
Page 121: ...Infrared Gas Analyzer Appendix Teledyne Analytical Instruments 121 ...
Page 122: ...Appendix Model 7600 Teledyne Analytical Instruments 122 ...
Page 123: ...Infrared Gas Analyzer Appendix Teledyne Analytical Instruments 123 ...
Page 124: ...Appendix Model 7600 Teledyne Analytical Instruments 124 ...
Page 126: ...Index Model 7600 Teledyne Analytical Instruments 126 ...