Operation Model
7600
Teledyne Analytical Instruments
70
Note: During auto zero calibration, the keys are inoperable other
than key lock ON/OFF and “Stop Auto Zero Calibration.”
When the key lock is set to ON, even the “Auto Zero
Calibration Stop” is locked out. To stop “Auto Zero
Calibration” forcedly, set the key lock to OFF and then
execute “Auto Zero Calibration Stop” as detailed above.
3.5.6 Peak Alarm Setting
The optional peak alarm is triggered whenever the CO
concentration exceeds the upper limit value for a preset number of times.
The peak number is set from within the Peak Alarm Setting menu.
Note: Only instruments with the peak alarm option installed will
have this menu.
The following items are set within the peak alarm menu:
•
OFF/ON status of peak alarm
•
Alarm setpoint value
•
Alarm count (number of peaks)
•
Hysteresis
To adjust the peak alarm settings:
1.
From the measurement mode, enter the user mode by
pressing the MODE key.
2.
Point the cursor to “Setting of Peak Alarm” by pressing
the UP or DOWN key.
Summary of Contents for 7600
Page 14: ...Model 7600 Teledyne Analytical Instruments xiv ...
Page 20: ...Introduction Model 7600 Teledyne Analytical Instruments 20 ...
Page 96: ...Maintenance Model 7600 Teledyne Analytical Instruments 96 Figure 4 3 Sample Cell ...
Page 108: ...Maintenance Model 7600 Teledyne Analytical Instruments 108 ...
Page 120: ...Appendix Model 7600 Teledyne Analytical Instruments 120 A 9 Description on Terminal Block ...
Page 121: ...Infrared Gas Analyzer Appendix Teledyne Analytical Instruments 121 ...
Page 122: ...Appendix Model 7600 Teledyne Analytical Instruments 122 ...
Page 123: ...Infrared Gas Analyzer Appendix Teledyne Analytical Instruments 123 ...
Page 124: ...Appendix Model 7600 Teledyne Analytical Instruments 124 ...
Page 126: ...Index Model 7600 Teledyne Analytical Instruments 126 ...