Operation Model
7600
Teledyne Analytical Instruments
90
3.7 Calibration
3.7.1 Zero Calibration
The zero calibration is used for zero point adjustment. A suitable
zero calibration gas is described in Section 2.5.5
Preparation of
Calibration Gas
.
To zero calibrate the analyzer:
1.
Press the ZERO key on the Measurement screen to
display the Manual Zero Calibration screen.
2.
Select the channel to be calibrated by pressing the UP or
DOWN key. After selection, press the ENT key, and zero
gas will be supplied.
Note: For channels that are set to “both” in the “Zero Calibration”
of the Calibration Setting mode, the zero calibrations are
carried out at the same time.
3.
Wait until the display has stabilized with the zero gas
flowing. After stabilization, press the ENT key. At this
point, the zero calibration on the range indicated by the
cursor is carried out.
Note: If the selected channel was set for autoranging “AR” (see
Section 3.5.1
Switch Ranges
. the cursor automatically
moves to the range selected in “
Setting of Auto Calibration
Component/ Range
” (Section 3.5.2.4), and calibration is
carried out within that range.
Summary of Contents for 7600
Page 14: ...Model 7600 Teledyne Analytical Instruments xiv ...
Page 20: ...Introduction Model 7600 Teledyne Analytical Instruments 20 ...
Page 96: ...Maintenance Model 7600 Teledyne Analytical Instruments 96 Figure 4 3 Sample Cell ...
Page 108: ...Maintenance Model 7600 Teledyne Analytical Instruments 108 ...
Page 120: ...Appendix Model 7600 Teledyne Analytical Instruments 120 A 9 Description on Terminal Block ...
Page 121: ...Infrared Gas Analyzer Appendix Teledyne Analytical Instruments 121 ...
Page 122: ...Appendix Model 7600 Teledyne Analytical Instruments 122 ...
Page 123: ...Infrared Gas Analyzer Appendix Teledyne Analytical Instruments 123 ...
Page 124: ...Appendix Model 7600 Teledyne Analytical Instruments 124 ...
Page 126: ...Index Model 7600 Teledyne Analytical Instruments 126 ...