![Teledyne 7600 Operating Instructions Manual Download Page 100](http://html.mh-extra.com/html/teledyne/7600/7600_operating-instructions-manual_1080467100.webp)
Maintenance Model
7600
Teledyne Analytical Instruments
100
introduced directly to the inlet of analyzer unit through
the gas cylinder.
3.
Proceed to an optical adjustment as described in the
maintenance mode. The display on the operation panel of
the main unit is illustrated in Figure 4-7. Balance adjust-
ment is not required if the display falls within ±100.
Figure 4-7: Optical Adjustment Display for Dual Optical System
Option
Table 4-2: Components of Optical Adjustment Screen
No. Of Components
to be measured
a b c d e f g h
1-component
meter
Main
Comp
– – – – – –
NO/SO2
NO
Main
NO
Comp
SO2
Main
SO2
Comp
– – – –
CO2/CO
CO2
Main
CO2
Comp
CO
Main
CO
Comp
–
2-
component
meter
NO/CO
NO
Main
NO
Comp
– –
CO
Main
CO
Comp
– –
3-component meter
NO/SO2/CO
NO
Main
NO
Comp
SO2
Main
SO2
Comp
CO
Main
CO
Comp
– –
4-component meter
NO/SO2/CO2/CO
NO
Main
NO
Comp
SO2
Main
SO2
Comp
CO2
Main
–
CO
Main
CO
Comp
Summary of Contents for 7600
Page 14: ...Model 7600 Teledyne Analytical Instruments xiv ...
Page 20: ...Introduction Model 7600 Teledyne Analytical Instruments 20 ...
Page 96: ...Maintenance Model 7600 Teledyne Analytical Instruments 96 Figure 4 3 Sample Cell ...
Page 108: ...Maintenance Model 7600 Teledyne Analytical Instruments 108 ...
Page 120: ...Appendix Model 7600 Teledyne Analytical Instruments 120 A 9 Description on Terminal Block ...
Page 121: ...Infrared Gas Analyzer Appendix Teledyne Analytical Instruments 121 ...
Page 122: ...Appendix Model 7600 Teledyne Analytical Instruments 122 ...
Page 123: ...Infrared Gas Analyzer Appendix Teledyne Analytical Instruments 123 ...
Page 124: ...Appendix Model 7600 Teledyne Analytical Instruments 124 ...
Page 126: ...Index Model 7600 Teledyne Analytical Instruments 126 ...