Operation Model
7600
Teledyne Analytical Instruments
56
4.
The alarm item selection screen will appear where you
can select the alarm type, assign it to a channel, define
setpoints, and enable or disable the alarm. Use the UP or
DOWN keys until the cursor is aligned with the desired
function and press the ENT key.
Note: Setpoints are adjustable form 0 to 100% of fullscale
however they must be defined so that H-limit value > L-limit
value and that (H-limit value- L-limit value) > hysteresis.
5.
After setting the last feature, store the values by pressing
ENT again.
To close the Alarm Setting screen or cancel this mode midway,
press the ESC key. The previous screen will return.
3.5.3.2
H
YSTERESIS
S
ETTING
The hysteresis setting is used prevent chattering of an alarm output
near the alarm setpoint. The hysteresis value range is 0-20% of fullscale
on any analysis range. Figure 3-6 shows the hysteresis width (in %
fullscale) set for a high limit alarm.
Summary of Contents for 7600
Page 14: ...Model 7600 Teledyne Analytical Instruments xiv ...
Page 20: ...Introduction Model 7600 Teledyne Analytical Instruments 20 ...
Page 96: ...Maintenance Model 7600 Teledyne Analytical Instruments 96 Figure 4 3 Sample Cell ...
Page 108: ...Maintenance Model 7600 Teledyne Analytical Instruments 108 ...
Page 120: ...Appendix Model 7600 Teledyne Analytical Instruments 120 A 9 Description on Terminal Block ...
Page 121: ...Infrared Gas Analyzer Appendix Teledyne Analytical Instruments 121 ...
Page 122: ...Appendix Model 7600 Teledyne Analytical Instruments 122 ...
Page 123: ...Infrared Gas Analyzer Appendix Teledyne Analytical Instruments 123 ...
Page 124: ...Appendix Model 7600 Teledyne Analytical Instruments 124 ...
Page 126: ...Index Model 7600 Teledyne Analytical Instruments 126 ...