Operation Model
7600
Teledyne Analytical Instruments
60
To set the flow time:
1.
Place the cursor next to “Flow Time,” then press the ENT
key.
2.
On the flow time setting screen that appears, use the UP
and DOWN keys to move the cursor to the gas whose
settings you want to change, then press ENT.
3.
The highlighted value can be changed. Change the value
by pressing the UP or DOWN key, and then move the
cursor to the right by pressing the SIDE key.
4.
After changing the value, press the ENT key.
5.
Press the ESC key to return to the automatic calibration
setting screen.
Note: Only the channels used are displayed on this screen. The
Ex. time is the output signal hold extension time after the
completion of calibration. It is valid only when the hold
setting is set to “ON.” The Ex. time set here is also used in
a manual calibration.
The auto calibration contacts are closed during an auto calibration
event.
3.5.4.3
C
YCLE
S
ETTING
R
ANGE
The cycle range is settable from 1 to 99 hours or 1 to 40 days. The
initial value is set at 7days.
Note: When an auto calibration starts, the measurement screen
automatically appears.
Summary of Contents for 7600
Page 14: ...Model 7600 Teledyne Analytical Instruments xiv ...
Page 20: ...Introduction Model 7600 Teledyne Analytical Instruments 20 ...
Page 96: ...Maintenance Model 7600 Teledyne Analytical Instruments 96 Figure 4 3 Sample Cell ...
Page 108: ...Maintenance Model 7600 Teledyne Analytical Instruments 108 ...
Page 120: ...Appendix Model 7600 Teledyne Analytical Instruments 120 A 9 Description on Terminal Block ...
Page 121: ...Infrared Gas Analyzer Appendix Teledyne Analytical Instruments 121 ...
Page 122: ...Appendix Model 7600 Teledyne Analytical Instruments 122 ...
Page 123: ...Infrared Gas Analyzer Appendix Teledyne Analytical Instruments 123 ...
Page 124: ...Appendix Model 7600 Teledyne Analytical Instruments 124 ...
Page 126: ...Index Model 7600 Teledyne Analytical Instruments 126 ...