Infrared Gas Analyzer
Operation
Teledyne
Analytical
Instruments
51
performed only for the range displayed when calibration
of the selected channel is performed.
6.
Press the ENT
key after the selection, and the specified
calibration is set.
To close the calibration range selection screen or cancel this mode
midway, press the ESC key. The previous screen will return.
Note: To perform calibration for “both,” set the same calibration
gas concentration for both ranges.
After setting the calibration range to “both”, you will notice in
the Manual Calibration screen (see Section 3.7), that there are cursors
next to both ranges whereas there would only be a single cursor next to
the range for calibration if “current” was set. For example, if channel 1
and channel 4 were set to “both” then cursors will appear in both ranges
of Ch1 and Ch4 in the Manual Calibration screen shown below.
3.5.2.4
S
ETTING OF AUTO CALIBRATION COMPONENT
/
RANGE
This menu is used to select the channel and the range in which auto
calibration is to be performed. If a channel has been set to autoranging
Summary of Contents for 7600
Page 14: ...Model 7600 Teledyne Analytical Instruments xiv ...
Page 20: ...Introduction Model 7600 Teledyne Analytical Instruments 20 ...
Page 96: ...Maintenance Model 7600 Teledyne Analytical Instruments 96 Figure 4 3 Sample Cell ...
Page 108: ...Maintenance Model 7600 Teledyne Analytical Instruments 108 ...
Page 120: ...Appendix Model 7600 Teledyne Analytical Instruments 120 A 9 Description on Terminal Block ...
Page 121: ...Infrared Gas Analyzer Appendix Teledyne Analytical Instruments 121 ...
Page 122: ...Appendix Model 7600 Teledyne Analytical Instruments 122 ...
Page 123: ...Infrared Gas Analyzer Appendix Teledyne Analytical Instruments 123 ...
Page 124: ...Appendix Model 7600 Teledyne Analytical Instruments 124 ...
Page 126: ...Index Model 7600 Teledyne Analytical Instruments 126 ...