Installation Model
7600
Teledyne Analytical Instruments
32
2.6 Electrical Connections
2.6.1 Power Inlet
The power inlet for the Model 7600 is located on the rear panel as
shown in Figure 2-9. The proper 3-prong power cord has been supplied
with your instrument. Connect the power cable to the power inlet.
Figure 2-9: Model 7600 Electrical Connections
The Model 7600 requires 100-240 VAC 50/60 Hz power.
Avoid installing this instrument near an electrical unit that
generates electrical noise such as a high frequency furnace or electric
welder. If using the instrument near such a noise-generating unit is
unavoidable, use a separate power line to avoid noise.
Mount a noise suppressor such as varister or spark killer as shown
in Figure 2-10 to the noise generating unit when noise is generated from
relays or solenoid valves. The suppressor must be mounted near the
noise generating source, or it will have no effect.
Summary of Contents for 7600
Page 14: ...Model 7600 Teledyne Analytical Instruments xiv ...
Page 20: ...Introduction Model 7600 Teledyne Analytical Instruments 20 ...
Page 96: ...Maintenance Model 7600 Teledyne Analytical Instruments 96 Figure 4 3 Sample Cell ...
Page 108: ...Maintenance Model 7600 Teledyne Analytical Instruments 108 ...
Page 120: ...Appendix Model 7600 Teledyne Analytical Instruments 120 A 9 Description on Terminal Block ...
Page 121: ...Infrared Gas Analyzer Appendix Teledyne Analytical Instruments 121 ...
Page 122: ...Appendix Model 7600 Teledyne Analytical Instruments 122 ...
Page 123: ...Infrared Gas Analyzer Appendix Teledyne Analytical Instruments 123 ...
Page 124: ...Appendix Model 7600 Teledyne Analytical Instruments 124 ...
Page 126: ...Index Model 7600 Teledyne Analytical Instruments 126 ...