Operation Model
7600
Teledyne Analytical Instruments
50
zero calibration screen that all there are cursors on all
components where “at once” is set.
Note: When the cylinder air or atmospheric air is used for the
zero gas, select “at once.”
3.5.2.3
S
ETTING OF CALIBRATION RANGE
Use the “About Calibration Range” submenu to configure the
instrument for calibration on one or two ranges for a channel.
To set calibration one or both ranges:
1.
From the measurement mode, press the MODE key to
display the user mode.
2.
Point the cursor to “Calibration Parameters” by pressing
the UP or DOWN key and press the ENT key.
3.
In the “Calibration Parameters” screen that appears, point
the cursor to “About Calibration Range” by pressing the
UP or DOWN key. Press the ENT key.
4.
In the next screen that appears, point the cursor to the
channel you want to set by pressing the UP or DOWN
key then press the ENT key.
5.
On the “calibration range selection” screen that appears,
select “both” or “current” by pressing the UP or DOWN
key.
If “both” is selected, zero or span calibration is performed
with Range 1 and Range 2 of the selected channel.
If “current” is selected, zero or span calibration is
Summary of Contents for 7600
Page 14: ...Model 7600 Teledyne Analytical Instruments xiv ...
Page 20: ...Introduction Model 7600 Teledyne Analytical Instruments 20 ...
Page 96: ...Maintenance Model 7600 Teledyne Analytical Instruments 96 Figure 4 3 Sample Cell ...
Page 108: ...Maintenance Model 7600 Teledyne Analytical Instruments 108 ...
Page 120: ...Appendix Model 7600 Teledyne Analytical Instruments 120 A 9 Description on Terminal Block ...
Page 121: ...Infrared Gas Analyzer Appendix Teledyne Analytical Instruments 121 ...
Page 122: ...Appendix Model 7600 Teledyne Analytical Instruments 122 ...
Page 123: ...Infrared Gas Analyzer Appendix Teledyne Analytical Instruments 123 ...
Page 124: ...Appendix Model 7600 Teledyne Analytical Instruments 124 ...
Page 126: ...Index Model 7600 Teledyne Analytical Instruments 126 ...