Infrared Gas Analyzer
Operation
Teledyne
Analytical
Instruments
55
CAUTION:
ALARMS ARE INACTIVE FOR 10 MINUTES AFTER
TURNING ON POWER.
3.5.3.1
C
ONFIGURING THE
A
LARMS
Note: To change an alarm setting, first set the alarm ON/OFF
setting to OFF, and then change the value.
1.
From the measurement mode, press the MODE key to
display the user mode.
2.
Point the cursor to “Alarm Setting” by pressing the UP or
DOWN key and press the ENT key.
3.
In the “Alarm Setting” screen that appears, point the
cursor to the alarm number you want to set (1 through 6)
using the UP or DOWN key. Press the ENT key when the
cursor is on the alarm number you desire.