Infrared Gas Analyzer
Appendix
Teledyne
Analytical
Instruments
117
Response time (90%): 15 sec electrical response.
Within 60 seconds including replacement
time of sampling gas (with 0.5L/min gas
flow rate). Gas replacement time depends
on the number of components,and
measuring range.
A.5 Standard Requirements for Sample Gas
Flowrate: 0.5L / min ±0.2 L/ min
Temperature: 0 to 50°C
Pressure: 10 kPa (1.45 psi) or less (Gas outlet side
should be open to the atmospheric air.)
Dust:
100
g/Nm
3
in particle size of 1
μ
m or less
Mist:
Unallowable
Moisture: Below saturation at 2°C (condensation un-
allowable).
Corrosive component: 1 ppm or less
Standard gas for calibration: Zero gas: Dry N
2
Span gas: Each sample gas having
concentration 90 to 100% of its measuring
range (recommended). Concentration
above 100% FS is unusable.
If an external zirconia O
2
sensor is
installed and calibration is carried out on
the same calibration gas line:
Zero gas: Dry air or atmospheric air
(provided without CO
2
sensor).
Span gas: For other than O
2
measurement,
each sample gas having concentration 90
to 100% of its measuring range.
For
O
2
measurement: O
2
gas of 1 to 2 vol
%.
Summary of Contents for 7600
Page 14: ...Model 7600 Teledyne Analytical Instruments xiv ...
Page 20: ...Introduction Model 7600 Teledyne Analytical Instruments 20 ...
Page 96: ...Maintenance Model 7600 Teledyne Analytical Instruments 96 Figure 4 3 Sample Cell ...
Page 108: ...Maintenance Model 7600 Teledyne Analytical Instruments 108 ...
Page 120: ...Appendix Model 7600 Teledyne Analytical Instruments 120 A 9 Description on Terminal Block ...
Page 121: ...Infrared Gas Analyzer Appendix Teledyne Analytical Instruments 121 ...
Page 122: ...Appendix Model 7600 Teledyne Analytical Instruments 122 ...
Page 123: ...Infrared Gas Analyzer Appendix Teledyne Analytical Instruments 123 ...
Page 124: ...Appendix Model 7600 Teledyne Analytical Instruments 124 ...
Page 126: ...Index Model 7600 Teledyne Analytical Instruments 126 ...