Maintenance Model
7600
Teledyne Analytical Instruments
102
Figure 4-9: Beam Adjustment Plate
8.
Move the beam adjustment plate sidewise to change the
value displayed at (c) or (g).
9.
Move the beam adjustment plate sidewise to change the
value displayed at (d) or (h).
10.
Repeat the procedures in (6) to 9) to make all the
displayed values come close to 0 as possible within ±100
range.
Note: Adjust the beam adjustment plate which is the nearest to
the zero adjustment knob first, and sequentially.
11.
After the optical balance adjustment, remount the top
cover of the analyzer unit, then carry out a moisture
interference compensation adjustment, and perform zero
and span calibrations.
Note: Before moving the beam adjustment plate, loosen the
detector set bolts (just enough to make the plate movable
for snug adjustment.
4.5 Moisture Interference Compensation Adjustment
If the following operation is not performed correctly,
subsequent analysis may adversely be affected. This
Summary of Contents for 7600
Page 14: ...Model 7600 Teledyne Analytical Instruments xiv ...
Page 20: ...Introduction Model 7600 Teledyne Analytical Instruments 20 ...
Page 96: ...Maintenance Model 7600 Teledyne Analytical Instruments 96 Figure 4 3 Sample Cell ...
Page 108: ...Maintenance Model 7600 Teledyne Analytical Instruments 108 ...
Page 120: ...Appendix Model 7600 Teledyne Analytical Instruments 120 A 9 Description on Terminal Block ...
Page 121: ...Infrared Gas Analyzer Appendix Teledyne Analytical Instruments 121 ...
Page 122: ...Appendix Model 7600 Teledyne Analytical Instruments 122 ...
Page 123: ...Infrared Gas Analyzer Appendix Teledyne Analytical Instruments 123 ...
Page 124: ...Appendix Model 7600 Teledyne Analytical Instruments 124 ...
Page 126: ...Index Model 7600 Teledyne Analytical Instruments 126 ...