Specifications
RSA3408A Technical Reference
2- 5
Table 2- 4: Amplitude
Characteristics
Description
Reference level
Reference level setting range
Baseband: --30 to +20 dBm (5 dB step)
RF: --50 to +30 dBm (1 dB step)
Option 03 IQ input: --10 to +20 dBm (5 dB step)
Accuracy (--10 to --50 dBm)
±
0.2 dB (at 100 MHz, 10 dB attenuation, 20 to 30
_
C)
Frequency response (RF attenuation
≥
10 dB)
at 20 to 30
_
C
±
0.5 dB (100 kHz to 40 MHz)
±
1.2 dB (40 MHz to 3.5 GHz)
±
1.7 dB (3.5 GHz to 6.5 GHz)
±
1.7 dB (5 GHz to 8 GHz)
Typical
±
0.3 dB (100 kHz to 40 MHz)
±
0.5 dB (40 MHz to 3.5 GHz)
±
1.0 dB (3.5 GHz to 6.5 GHz)
±
1.0 dB (5 GHz to 8 GHz)
at 10 to 40
_
C
±
0.7 dB (100 kHz to 40 MHz)
±
1.5 dB (40 MHz to 3.5 GHz)
±
2.0 dB (3.5 GHz to 6.5 GHz)
±
2.0 dB (5 GHz to 8 GHz)
n
Absolute amplitude accuracy
at calibration point (20 to 30
_
C)
Baseband:
±
0.3 dB (at 25 MHz, --10 dBm signal)
RF:
±
0.5 dB (at 100 MHz, --20 dBm signal, 0 dB attenuation)
n
Input attenuator setting uncertainty
±
0.2 dB (at 100 MHz)
n
Level linearity in display range
±
0.2 dB (0 to --50 dBfs);
±
0.12 dB (0 to --50 dBfs,