Performance Tests
RSA3408A Technical Reference
1- 45
Input attenuator setting uncertainty test
Attenuation
Measurement (P
X
)
Low limit
Test result (P
X
- P
0/1
)
High limit
0 dB
P
0
:
--
--
--
5 dB
--0.2 dB
+0.2 dB
10 dB
--0.2 dB
+0.2 dB
15 dB
--0.2 dB
+0.2 dB
20 dB
--0.2 dB
+0.2 dB
25 dB
--0.2 dB
+0.2 dB
30 dB
--0.2 dB
+0.2 dB
30 dB
P
1
:
--
--
--
35 dB
--0.2 dB
+0.2 dB
40 dB
--0.2 dB
+0.2 dB
45 dB
--0.2 dB
+0.2 dB
50 dB
--0.2 dB
+0.2 dB
55 dB
--0.2 dB
+0.2 dB
Level linearity test
Amplitude
Low limit
Test result
High limit
0 dBm
--0.2 dB
+0.2 dB
--10 dBm
--0.2 dB
+0.2 dB
--20 dBm
--0.2 dB
+0.2 dB
--30 dBm
--0.2 dB
+0.2 dB
--40 dBm
--0.2 dB
+0.2 dB
3
rd
order intermodulation distortion test
Low limit
Test result
High limit
3
rd
order intermodulation distortion
+5 dBm reference level, 2 GHz center frequency
--78 dBc
Summary of Contents for RSA3408A