Performance Tests
1- 28
RSA3408A Technical Reference
3.
Determine the reference power in +10 to --30 dBm:
a.
Measure the power and record it as Pr0.
b.
Set the signal generator amplitude to 0 to --30 dBm in 10 dB steps and
measure the power as Pr1 to 4, respectively.
Signal generator amplitude Reference power
+10 dBm
Pr0
0 dBm
Pr1
--10 dBm
Pr2
--20 dBm
Pr3
--30 dBm
Pr4
4.
Modify the hookup:
a.
Connect the generator output through a 50
Ω
N-N coaxial cable to the
spectrum analyzer (Item 4) input. See the following figure.
50
Ω
N-N coaxial cable
Output
Signal generator
Spectrum analyzer
b.
Set the signal generator control:
Amplitude
--30 dBm
. . . . . . . . . . . . . . . .
c.
Set the spectrum analyzer controls
:
Measurement mode
SA mode
. . . . . . . .
Reference level
--20 dBm
. . . . . . . . . . . .
Center frequency
100 MHz
. . . . . . . . . .
Span
1 MHz
. . . . . . . . . . . . . . . . . . . .
RF attenuation
10 dB
. . . . . . . . . . . .
RBW
Auto
. . . . . . . . . . . . . . . . . . . .