Performance Tests
1- 34
RSA3408A Technical Reference
Equipment
Required
None
Prerequisites
Refer to page 1--10.
1.
Prepare for the test:
a.
Initialize the analyzer:
H
Press the
SYSTEM
key on the front panel.
H
Press the
Reset All to Factory Defaults
side key.
RSA3408A
b.
Modify the analyzer controls:
H
Press the
FREQUENCY/CHANNEL
key on the front panel.
H
With the
Center Freq
side key, set the frequency to 10 MHz using
the general purpose knob.
H
Press the
SPAN
key on the front panel.
H
With the
Span
side key, set the span to 1 MHz using the general
purpose knob.
H
Press the
AMPLITUDE
key on the front panel.
H
With the
Ref Level
side key, set the level to --30 dBm using the
general purpose knob.
H
Press the
RBW/FFT
key on the front panel.
H
Press the
RBW/FFT
side key to select
Man
.
H
With the
RBW
side key, set the value to 10 kHz using the general
purpose knob.
Check Displayed Average
Noise Level