Performance Tests
1- 10
RSA3408A Technical Reference
Prerequisites
The tests in this section make up an extensive, valid confirmation of performance
and functionality when the following requirements are met:
H
The cabinet must be installed on the analyzer.
H
The instrument must have passed the procedures under
Brief Procedures
found on page 1--3.
H
A signal-path compensation must have been done within the recommended
calibration interval and at a temperature within
±
3
_
C of the present
operating temperature. (If at the time you did the prerequisite
Brief Procedures,
the temperature was within the limits just stated, consider
this prerequisite met.)
H
The analyzer must have been last adjusted at an ambient temperature
b20
_
C and +30
_
C, must have been operating for a warm-up
period of at least 20 minutes, and must be operating at an ambient tempera-
ture of b10
_
C and +40
_
C.