Performance Tests
RSA3408A Technical Reference
1- 39
4.
Check residual response in the RF2 band:
a.
Modify the analyzer controls:
H
Press the
FREQUENCY/CHANNEL
key on the front panel.
H
With the
Center Freq
side key, set the frequency to 5 GHz using the
numeric keypad.
b.
Measure residual response:
Press the
PEAK
key to place the marker at the peak on the noise floor.
c.
Check against limits:
Read the marker readout. Confirm that the residual signal level is
--85 dBm or less within the whole bandwidth.
5.
Check residual response in the RF3 band:
a.
Modify the analyzer controls:
H
Press the
FREQUENCY/CHANNEL
key on the front panel.
H
With the
Center Freq
side key, set the frequency to 6.5 GHz using
the numeric keypad.
b.
Measure residual response:
Press the
PEAK
key to place the marker at the peak on the noise floor.
c.
Check against limits:
Read the marker readout. Confirm that the residual signal level is
--85 dBm or less within the whole bandwidth.