Performance Tests
1- 14
RSA3408A Technical Reference
4.
Check frequency accuracy in the RF2 band:
a.
Modify the generator control:
Frequency
5 GHz
. . . . . . . . . . . . . . . .
b.
Modify the analyzer controls:
H
Press the
FREQUENCY/CHANNEL
key on the front panel.
H
With the
Center Freq
side key, set the frequency to 5 GHz using the
numeric keypad.
c.
Measure peak frequency:
H
Press the
PEAK
key to place the marker on the peak signal.
H
Read the marker readout. Confirm that the value is
5 GHz
2.0 kHz.
5.
Check frequency accuracy in the RF3 band:
a.
Modify the generator control:
Frequency
7 GHz
. . . . . . . . . . . . . . . .
b.
Modify the analyzer controls:
H
Press the
FREQUENCY/CHANNEL
key on the front panel.
H
With the
Center Freq
side key, set the frequency to 7 GHz using the
numeric keypad.
c.
Measure peak frequency:
H
Press the
PEAK
key to place the marker on the peak signal.
H
Read the marker readout. Confirm that the value is
7 GHz
2.4 kHz.
6.
Disconnect the test equipment:
Disconnect the cable at the analyzer input.
Summary of Contents for RSA3408A