Performance Tests
1- 24
RSA3408A Technical Reference
Equipment
Required
One signal generator (Item 1)
One 50
Ω
N-N coaxial cables (Item 10)
Prerequisites
Refer to page 1--10.
1.
Prepare for the test:
a.
Set the generator controls:
Frequency
100 MHz
. . . . . . . . . . . . . . . .
Amplitude
--25 dBm
. . . . . . . . . . . . . . . .
RF output
On
. . . . . . . . . . . . . . . .
b.
Hook up the generator:
Connect the generator output through a 50
Ω
N-N coaxial cable to the
analyzer
INPUT
. See the following figure.
50
Ω
N-N coaxial cable
Output
Signal generator
RSA3408A
c.
Initialize the analyzer:
H
Press the
SYSTEM
key on the front panel.
H
Press the
Reset All to Factory Defaults
side key.
Check Input Attenuator
Setting Uncertainty
Summary of Contents for RSA3408A
Page 4: ......
Page 8: ...Table of Contents iv RSA3408A Technical Reference...
Page 12: ...Preface viii RSA3408A Technical Reference...
Page 13: ...Performance Verification...
Page 14: ......
Page 61: ...Specifications...
Page 62: ......
Page 85: ......
Page 86: ......