Performance Tests
1- 20
RSA3408A Technical Reference
Power reference output
Power meter
Figure 1- 6: Hookup for calibrating the power sensor
Baseband.
2.
Set the signal generator output:
a.
Hook up the instruments:
Connect the signal generator (Item 1) output through a 50
Ω
N-N coaxial
cable to the power sensor input. See Figure 1--7.
Output
Signal generator
Power meter
Power sensor
Figure 1- 7: Hookup for setting the signal generator output
b.
Set the signal generator controls:
Frequency
25 MHz
. . . . . . . . . . . . . . . .
Level
--10 dBm
. . . . . . . . . . . . . . . . . . . .
c.
Set the power meter control:
Frequency
10 MHz
. . . . . . . . . . . . . . . .
d.
Adjust the output level of the signal generator so that the power meter
reads --10 dBm
0.05 dBm.
e.
Disconnect the cable from the power sensor input.