Performance Tests
RSA3408A Technical Reference
1- 35
H
Press
TRACE/AVG
>
Trace 1 Type...
>
Average
.
H
Press the
Number Of Averages
side key and set the value to 50.
H
Press
MEASURE
>
Channel Power
>
MEAS SETUP
.
H
Press the
Channel Bandwidth
side key and set the value to 1 MHz
using the general purpose knob.
2.
Check noise in the baseband:
Read the measurement result of Density at the bottom of the screen.
Check that the value is --151 dBm/Hz or less.
3.
Check noise at 2 GHz in the RF1 band:
a.
Modify the analyzer control:
H
Press the
FREQUENCY/CHANNEL
key on the front panel.
H
With the
Center Freq
side key, set the frequency to 2 GHz using the
general purpose knob.
H
Press the
AMPLITUDE
key on the front panel.
H
With the
Ref Level
side key, set the level to --50 dBm using the
general purpose knob.
H
Press the
RF Atten/Mixer
side key to select
RF Att
.
H
With the
RF Att
side key, set the attenuation to 0 dB using the
general purpose knob.
b.
Check against limits:
Read the measurement result of Density at the bottom of the screen.
Check that the value is --150 dBm/Hz or less.
4.
Check noise at 3 GHz in the RF1 band:
a.
Modify the analyzer control:
H
Press the
FREQUENCY/CHANNEL
key on the front panel.
H
With the
Center Freq
side key, set the frequency to 3 GHz using the
general purpose knob.
b.
Check against limits:
Read the measurement result of Density at the bottom of the screen.
Check that the value is --150 dBm/Hz or less.