Performance Tests
RSA3408A Technical Reference
1- 29
5.
Set the reference offset:
a.
Set the spectrum analyzer control:
Average
10
. . . . . . . . . . . . . . . . . .
b.
Measure the peak power Ppr using the maker peak.
c.
Modify the spectrum analyzer control:
Reference offset
Pr4 -- Ppr
. . . . . . . . . . .
6.
Determine the reference power in --40 dBm:
a.
Set the signal generator amplitude to --40 dBm to measure the power
as Pr5.
Signal generator amplitude Reference power
--40 dBm
Pr5
b.
Disconnect the cable from the spectrum analyzer.
7.
Modify the hookup:
a.
Connect the signal generator output to the RSA3408A analyzer input.
See the following figure.
50
Ω
N-N coaxial cable
Output
Signal generator
RSA3408A
b.
Set the signal generator control:
Amplitude
+10 dBm
. . . . . . . . . . . . . . . .