Brief Procedures
1- 4
RSA3408A Technical Reference
Functional Tests
The analyzer has a built-in calibration signal source with an amplitude of
approximately --20 dBm and frequency of 100 MHz. Using this source, perform
this quick functional check to verify that your instrument is operating correctly.
Equipment
Required
None
Prerequisites
Power on the analyzer and allow a 20 minute warm-up before doing
this procedure.
1.
Initialize the analyzer:
a.
Press the
SYSTEM
key on the front panel.
b.
Press the
Reset All to Factory Defaults
side key.
2.
Check the system version and options:
a.
Press the
SYSTEM
key on the front panel.
b.
Press the
Versions and Installed Options...
side key.
c.
Check the version in the “Main System” field and the options in the
option table.
3.
Check the display brightness:
a.
Press the
Cancel-Back
(top) side key.
b.
Press the
Display Brightness
side key.
c.
Change the value from 0 to 100% using the general purpose knob to
check that the brightness changes normally.
4.
Display spectrum of the calibration signal:
a.
Press
S/A
>
Spectrum Analyzer
.
b.
Press the
PRESET
key on the front panel to reset the analyzer.
c.
Press
INPUT
>
Signal Input Port...
>
Cal100M
.
The spectrum of the calibration signal appears.
d.
Check that “INPUT: CAL” and “FREE RUN” are displayed in the status
indicator at the upper right of the screen (see Figure 1--1).