Brief Procedures
1- 8
RSA3408A Technical Reference
Diagnostics
You can run the internal diagnostic program to check hardware states.
Table 1--2 shows the test items and their descriptions.
Table 1- 2: Internal diagnostic program
Menu item
Description
Possible failure
Temp Sensor
Checks that the temperature sensor works correctly. RF5 module
Data Memory
Checks that the acquisition memory works correctly. A40 board
Equipment
Required
None
Prerequisites
Power on the analyzer and allow a 20 minute warm-up before doing
this procedure.
1.
Press
CAL
>
Service...
>
Password
.
2.
Enter
270833
and then press the
ENTER
key using the numeric keypad.
3.
Press the
DIAG
side key.
4.
Select the diagnostic menu item to be executed.
You can press the
All
side key to run all the tests.
5.
Check the result shown in the lower left view (“Pass” or “Fail”).
The process is
graphically displayed
on the upper views.
Figure 1- 5: Diagnostic screen