Performance Tests
1- 40
RSA3408A Technical Reference
Equipment
Required
One signal generator (Item 2)
One 50
Ω
N-N coaxial cable (Item 10)
Prerequisites
Refer to page 1--10.
1.
Prepare for the test:
a.
Initialize the analyzer:
H
Press the
SYSTEM
key on the front panel.
H
Press the
Reset All to Factory Defaults
side key.
b.
Modify the default controls:
H
Press the
SPAN
key on the front panel.
H
With the
Span
side key, set the span to 10 MHz using the general
purpose knob.
H
Press the
RBW/FFT
key on the front panel.
H
Press the
RBW/FFT
side key to select
Man
.
H
With the
RBW
side key, set the value to 50 kHz using the general
purpose knob.
H
Press
TRACE/AVG
>
Trace 1 Type...
>
Average
.
H
Press the
Number Of Averages
side key and set the value to 50.
H
Press the
MARKER SETUP
key.
H
Press the
Markers
side key to select
Delta
.
c.
Set the generator controls:
Level
--5 dBm
. . . . . . . . . . . . . . . . . . . .
d.
Hook up the signal generator:
Connect the generator output through a 50
Ω
N-N coaxial cable to the
analyzer
INPUT
. See the following figure.
Check Spurious Response