Performance Tests
1- 16
RSA3408A Technical Reference
Noise Sideband Checks
These procedures check the noise sideband-related characteristics and are listed
as checked in Chapter 2,
Specifications
, Table 2--2 on page 2--3.
Equipment
Required
One signal generator (Item 3)
One 50
Ω
N-N coaxial cable (Item 10)
Prerequisites
Refer to page 1--10.
1.
Prepare for the test:
a.
Initialize the analyzer:
H
Press the
SYSTEM
key on the front panel.
H
Press the
Reset All to Factory Defaults
side key.
b.
Modify the default settings:
H
Press the
FREQUENCY/CHANNEL
key on the front panel.
H
The
Center Freq
menu item is selected by default.
Set the frequency to 1 GHz using the numeric key pad.
H
Press the
AMPLITUDE
key on the front panel.
H
Press the
RF Atten/Mixer
side key to select
Mixer
.
H
With the
Mixer Level
side key, set the level to --10 dBm.
H
Press
TRACE/AVG
>
Trace 1 Type...
→
Average
.
H
Press the
Number Of Averages
side key and set the value to 50
using the general purpose knob.
c.
Set the generator controls:
Frequency
1000 MHz
. . . . . . . . . . . . . . . .
Amplitude
0 dBm
. . . . . . . . . . . . . . . .
RF output
On
. . . . . . . . . . . . . . . .
d.
Hook up the signal generator:
Connect the generator output through a 50
Ω
N-N coaxial cable to the
analyzer
INPUT
. See the following figure.
Check Noise Sideband
Summary of Contents for RSA3408A
Page 4: ......
Page 8: ...Table of Contents iv RSA3408A Technical Reference...
Page 12: ...Preface viii RSA3408A Technical Reference...
Page 13: ...Performance Verification...
Page 14: ......
Page 61: ...Specifications...
Page 62: ......
Page 85: ......
Page 86: ......