Performance Tests
RSA3408A Technical Reference
1- 15
Equipment
Required
One frequency counter (Item 5)
One 50
Ω
BNC coaxial cable (Item 9)
One 50
Ω
terminator (Item 12)
Prerequisites
Refer to page 1--10.
The frequency counter must have been operating for a warm-up period
of at least 24 hours.
1.
Prepare for the test:
a.
Hook up the frequency counter:
Connect
REF OUT
on the rear panel of the analyzer through a 50
Ω
precision coaxial cable and a 50
Ω
precision terminator to Channel 1 of
the counter. See the following figure.
Frequency counter
50
Ω
BNC coaxial cable
50
Ω
terminator
RSA3408A (rear)
CH1 input
REF OUT
b.
Set the counter controls:
Function
Frequency 1
. . . . . . . . . . . . . . . . .
Gate time
2 s
. . . . . . . . . . . . . . . .
2.
Check the frequency:
Check that the frequency counter reads 10 MHz
2.0 Hz.
3.
Disconnect the hookup:
Disconnect the cable at
REF OUT
.
Check 10 MHz Reference