Performance Tests
RSA3408A Technical Reference
1- 21
3.
Prepare for the test:
a.
Hook up the instruments:
Connect the signal generator output through a 50
Ω
N-N coaxial cable to
the RSA3408A
INPUT
. See Figure 1--8.
50
Ω
N-N coaxial cable
Output
Signal generator
RSA3408A
Figure 1- 8: Hookup for checking the absolute amplitude accuracy
b.
Initialize the analyzer:
H
Press the
SYSTEM
key on the front panel.
H
Press the
Reset All to Factory Defaults
side key.
c.
Modify the default settings:
H
Press the
FREQUENCY/CHANNEL
key on the front panel.
H
With the
Center Freq
side key, set the frequency to 25 MHz using
the numeric keypad.
H
Press the
SPAN
key on the front panel.
H
With the
Span
side key, set the span to 10 MHz using the general
purpose knob.
d.
Calibrate the analyzer gain:
H
Press
CAL
>
Calibrate Gain
.
H
Press
Amplitude
>
Auto Level
.