Models 707B and 708B Switching Matrix Reference Manual
Section 7: TSP command reference
707B-901-01 Rev. B / January 2015
7-137
Example 2
scan.create("1A01:1A08")
scan.add("1A10")
scan.add("1A09")
Replaces the active scan list with an empty
scan list, and then adds row A, columns 1
through 8 on slot 1 to the new scan list.
Adds row A, column 10 on slot 1 to the end of
the scan list.
Adds row A, column 9 on slot 1 to the end of
the scan list.
Scan list now includes channels 1A01
through 1A10, with channels 1A01 through
1A08 in order, followed by channel 1A10, and
then channel 1A09.
Example 3
scan.create("")
Clears the old scan list and creates a new
empty scan list.
Also see
(on page 7-141)
(on page 3-1)