Models 707B and 708B Switching Matrix Reference Manual
Section 7: TSP command reference
707B-901-01 Rev. B / January 2015
7-235
Details
This function sets and gets the termination character sequence that is used to indicate the end of a
line for a TSP-Net connection.
Using the
termSequence
parameter sets the termination sequence. The present termination
sequence is always returned.
For the
termSequence
parameter, use the same values listed in the table above for type. There are
four possible combinations, all of which are made up of line feeds (LF or 0x10) and carriage returns
(CR or 0x13). For TSP-enabled devices, the default is
tspnet.TERM_LF
. For devices that are not
TSP-enabled, the default is
tspnet.TERM_CRLF
.
Example
deviceID = tspnet.connect("192.0.2.1")
if deviceID then
tspnet.termination(deviceID, tspnet.TERM_LF)
end
Sets termination type for IP address
192.0.2.1 to
TERM_LF
.
Also see
(on page 7-229)
(on page 7-230)
tspnet.timeout
This attribute sets the timeout value for the
tspnet.connect()
,
tspnet.execute()
, and
tspnet.read()
commands.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Attribute (RW)
No
Instrument reset
Recall setup
Create configuration script
Save setup
20.0 (20 s)
Usage
value
= tspnet.timeout
tspnet.timeout =
value
value
The timeout duration in seconds (0.001 to 30.0 s)
Details
This attribute sets the amount of time the
tspnet.connect()
,
tspnet.execute()
, and
tspnet.read()
commands will wait for a response.
The time is specified in seconds. The timeout may be specified to millisecond resolution, but is only
accurate to the nearest 10 ms.
Example
tspnet.timeout = 2.0
Sets the timeout duration to 2 s.
Also see
(on page 7-229)
(on page 7-231)
(on page 7-232)