Models 707B and 708B Switching Matrix Reference Manual
Section 7: TSP command reference
707B-901-01 Rev. B / January 2015
7-151
•
tsplink.trigger[
N
].EVENT_ID
: An edge (either rising, falling, or either based on the configuration
of the line) on the tsplink trigger line.
•
lan.trigger[
N
].EVENT_ID:
Event identifier use to route the LAN trigger to other subsystems
(using stimulus properties).
•
scan.trigger.EVENT_SCAN_READY
: Scan Ready Event.
•
scan.trigger.EVENT_SCAN_START
: Scan Start Event.
•
scan.trigger.EVENT_CHANNEL_READY
: Channel Ready Event.
•
scan.trigger.EVENT_SCAN_COMP
: Scan Complete Event.
scan.trigger.EVENT_IDLE
: Idle Event.
Use one of the text trigger event IDs (for example,
digio.trigger[
N
].EVENT_ID)
to set the
stimulus value rather than the numeric value. Doing this will make the code compatible for future
upgrades.
Example 1
scan.trigger.arm.stimulus =
scan.trigger.EVENT_SCAN_READY
Sets trigger stimulus of the arm
event detector to scan ready
event.
Example 2
scan.trigger.arm.stimulus = 0
The scan begins immediately.
Example 3
scan.trigger.arm.stimulus = digio.trigger[3].EVENT_ID
The scan begins when the
instrument receives a signal
from digital I/O line 3.
Also see
(on page 7-149)
(on page 7-150)
(on page 3-1)