Models 707B and 708B Switching Matrix Reference Manual
Section 7: TSP command reference
707B-901-01 Rev. B / January 2015
7-109
Example
lan.trigger[5].clear()
Clears the event detector with LAN packet 5.
Also see
(on page 7-108)
(on page 7-113)
(on page 7-115)
lan.trigger[N].connect()
This functionprepares the event generator for outgoing trigger events.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Function
Yes
Usage
lan.trigger[
N
].connect()
N
The LAN event number (1 to 8)
Details
This command prepares the event generator to send event messages. For TCP connections, this
opens the TCP connection.
The event generator automatically disconnects when either the protocol or IP address for this event is
changed.
Example
lan.trigger[1].protocol = lan.MULTICAST
lan.trigger[1].connect()
lan.trigger[1].assert()
Set the protocol for LAN trigger 1 to be
multicast when sending LAN triggers.
Then, after connecting the LAN trigger,
send a message on LAN trigger 1 by
asserting it.
Also see
(on page 7-108)
(on page 7-111)
(on page 7-113)
(on page 7-114)
(on page 7-115)
lan.trigger[N].connected
This attribute stores the LAN event connection state.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Attribute (R)
Yes
Not applicable
Not applicable
Not applicable