Models 707B and 708B Switching Matrix Reference Manual
Section 2: General operation
707B-901-01 Rev. B / January 2015
2-55
Change the IP configuration through the web interface
The LAN settings, such as IP address, subnet mask, gateway, and DNS address, can be changed
through the web page of the instrument.
If you change the IP address through the web page, the web page will try to redirect to the IP address
that gets configured in the instrument. In some cases, this may fail. This generally happens if you
switch from static IP address assignment to IP address assignment using a DHCP server. If this
happens, you need to revert to either using the front panel to set the IP address or use an automatic
discovery tool to determine the new IP address.
You can also change the IP configuration using TSP commands. See
(on page 5-10)
for more information.
To change the IP configuration using the instrument web page:
1. Access the instrument's internal web page (see
(on page 2-66)).
2. From the navigation bar on the left, in the LXI Home menu, select
IP Config
.
3. Click
Modify
.
4. You are prompted for a password. The default is admin.
Figure 42: Modify IP Configuration page
5. Change the values.
6. Click
Submit
. The instrument reconfigures its settings, which may take a few moments.