Models 707B and 708B Switching Matrix Reference Manual
Section 7: TSP command reference
707B-901-01 Rev. B / January 2015
7-243
Details
This function will wait for all previously started overlapped commands to complete.
Currently, the Models 707B and 708B have no overlapped commands implemented. However, other
TSP-enabled products, such as the Series 2600A System SourceMeter
®
Instruments, have
overlapped commands. Therefore, when the Model 707B or 708B is a TSP master to a subordinate
device with overlapped commands, use this function to wait until all overlapped operations are
completed.
A group number may only be specified when this node is the master node.
If no
group
is specified, the local group is used.
If zero (0) is specified for the
group
, this function waits for all nodes in the system.
Any nodes that are not assigned to a group (group number is 0) are part of the master node's group.
Example 1
waitcomplete()
Waits for all nodes in the local group.
Example 2
waitcomplete(G)
Waits for all nodes in group G.
Example 3
waitcomplete(0)
Waits for all nodes on the TSP-Link network.
Also see
None