Section 7: TSP command reference
Models 707B and 708B Switching Matrix Reference Manual
7-60
707B-901-01 Rev. B / January 2015
Details
Set this attribute to zero (0) to disable the automatic trigger output.
Do not use the stimulus attribute for generating output triggers under script control. Use
digio.trigger[
N
].assert()
instead.
The trigger stimulus for a digital I/O line
may be set to one of the existing trigger event IDs, described
in the following table.
Trigger event IDs
Trigger event ID
Description
digio.trigger[N].EVENT_ID
An edge (either rising, falling, or either based on the
configuration of the line) on the digital input line.
display.trigger.EVENT_ID
The trigger key on the front panel is pressed.
trigger.EVENT_ID
A *trg message on the active command interface. If
GPIB is the active command interface, a GET
message also generates this event.
trigger.blender[N].EVENT_ID
A combination of events has occurred.
trigger.timer[N].EVENT_ID
A delay expired.
tsplink.trigger[N].EVENT_ID
An edge (either rising, falling, or either based on the
configuration of the line) on the TSP-Link trigger line.
lan.trigger[N].EVENT_ID
A LAN trigger event has occurred.
scan.trigger.EVENT_SCAN_READY
Scan ready event.
scan.trigger.EVENT_SCAN_START
Scan start event.
scan.trigger.EVENT_CHANNEL_READY
Channel ready event.
scan.trigger.EVENT_SCAN_COMP
Scan complete event.
scan.trigger.EVENT_IDLE
Idle event.
Example 1
digio.trigger[3].stimulus = 0
Clear the trigger
stimulus of digital
I/O line 3.
Example 2
digio.trigger[3].stimulus = scan.trigger.EVENT_CHANNEL_READY
Set the trigger
stimulus of digital
I/O line 3 to be the
channel ready
event during a
scan.
Also see
(on page 7-53)