Section 7: TSP command reference
Models 707B and 708B Switching Matrix Reference Manual
7-80
707B-901-01 Rev. B / January 2015
Also see
(on page 7-63)
(on page 7-65)
(on page 7-75)
(on page 7-77)
display.trigger.clear()
This function clears the front-panel trigger event detector.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Function
Yes
Usage
display.trigger.clear()
Details
The trigger event detector remembers if an event has been detected since the last
display.trigger.wait()
call. This function clears the trigger event detector and discards the
previous history of TRIG key presses.
This attribute also clears the
display.trigger.overrun
attribute.
Also see
display.trigger.overrun
display.trigger.wait()
display.trigger.EVENT_ID
This constant is the event ID of the event generated when the front-panel TRIG key is pressed.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Constant
Yes
Usage
eventID
= display.trigger.EVENT_ID
eventID
The trigger event number
Details
Set the stimulus of any trigger event detector to the value of this constant to have it respond to front-
panel trigger key events.
Example
scan.trigger.channel.stimulus = display.trigger.EVENT_ID
Have the channel action of the trigger model be paced by a user pressing the front-panel TRIG key.
Also see
(on page 7-153)