Learn more about eye measurements.
Go to the mask testing setup overview.
Glitch trigger
A glitch trigger occurs when the instrument detects a pulse narrower (or wider) than some specified time. You can set the
instrument to trigger on glitches of either polarity or to reject glitches of either polarity.
Pattern trigger
A pattern trigger occurs when the inputs to the selected logic function cause the function to become True or False. When you use
a pattern trigger, you define:
■
The precondition for each logic input: logic high, low, or "don't care"; the logic inputs are the instrument channels
■
The Boolean logic function: AND, NAND, OR, or NOR
■
The condition for triggering: the Boolean function becomes True (logic high) or False (logic low), and whether the True
condition is time qualified
Pattern trigger logic choices are summarized in the next table.
Pattern
State
Definition
1
,
2
AND
Clocked AND
If all the preconditions selected for the
logic inputs
3
are TRUE, then the
instrument triggers.
NAND
Clocked NAND
If not all of the preconditions selected for
the logic inputs
4
are TRUE, then the
instrument triggers.
OR
Clocked OR
If any of the preconditions selected for the
logic inputs
5
are TRUE, then the
instrument triggers.
NOR
Clocked NOR
If none of the preconditions selected for
the logic inputs
6
are TRUE, then the
instrument triggers.
1
For state triggers, the definition must be met at the time the clock input changes state
2
The definitions given here are correct for the Goes TRUE setting in the Trigger When menu. If that menu is set to Goes False, swap the definition for AND with that for
NAND and for OR with NOR for both pattern and state types.
3
The logic inputs are channels 1, 2, 3, and 4 when using Pattern triggers. For State triggers, channel 4 becomes the clock input, leaving the remaining channels as logic
inputs.
4
The logic inputs are channels 1, 2, 3, and 4 when using Pattern triggers. For State triggers, channel 4 becomes the clock input, leaving the remaining channels as logic
inputs.
5
The logic inputs are channels 1, 2, 3, and 4 when using Pattern triggers. For State triggers, channel 4 becomes the clock input, leaving the remaining channels as logic
inputs.
6
The logic inputs are channels 1, 2, 3, and 4 when using Pattern triggers. For State triggers, channel 4 becomes the clock input, leaving the remaining channels as logic
inputs.
Oscilloscope reference
670
DPO70000SX, MSO/DPO70000DX, MSO/DPO70000C, DPO7000C, and MSO/DPO5000B Series
Summary of Contents for DPO2304SX
Page 1: ...Tektronix Digital Phosphor Oscilloscopes Printable Help P077006219 077 0062 19...
Page 2: ......
Page 3: ...Tektronix Digital Phosphor Oscilloscopes Printable Help www tek com 077 0062 19...
Page 32: ...Introduction xxviii DPO70000SX MSO DPO70000DX MSO DPO70000C DPO7000C and MSO DPO5000B Series...
Page 130: ...Cursor setups 98 DPO70000SX MSO DPO70000DX MSO DPO70000C DPO7000C and MSO DPO5000B Series...
Page 198: ...MagniVu setup 166 DPO70000SX MSO DPO70000DX MSO DPO70000C DPO7000C and MSO DPO5000B Series...
Page 212: ...Display setups 180 DPO70000SX MSO DPO70000DX MSO DPO70000C DPO7000C and MSO DPO5000B Series...
Page 300: ...Math setups 268 DPO70000SX MSO DPO70000DX MSO DPO70000C DPO7000C and MSO DPO5000B Series...
Page 408: ...MyScope setups 376 DPO70000SX MSO DPO70000DX MSO DPO70000C DPO7000C and MSO DPO5000B Series...
Page 510: ...Trigger setups 478 DPO70000SX MSO DPO70000DX MSO DPO70000C DPO7000C and MSO DPO5000B Series...
Page 548: ...Vertical setups 516 DPO70000SX MSO DPO70000DX MSO DPO70000C DPO7000C and MSO DPO5000B Series...
Page 605: ...Utilities DPO70000SX MSO DPO70000DX MSO DPO70000C DPO7000C and MSO DPO5000B Series 573...
Page 606: ...Utilities 574 DPO70000SX MSO DPO70000DX MSO DPO70000C DPO7000C and MSO DPO5000B Series...
Page 920: ...Index 888 DPO70000SX MSO DPO70000DX MSO DPO70000C DPO7000C and MSO DPO5000B Series...