Automatic measurements on communications signals
NOTE. If a standard listed in this manual is not available on your instrument, it is because the configuration or bandwidth of
your instrument cannot test that standard.
Although this oscilloscope is not a calibrated optical reference receiver, you can use it with mask testing to evaluate general
optical signal characteristics and wave shape, using an external O/ E converter.
What do you want to do next?
Learn more about mask testing.
Learn more about eye measurements.
Learn about communication triggering.
Levels used in taking eye measurements
All eye-diagram measurements are based on the power level, the voltage level, or the time locations of edges within each
acquisition.
The next figure shows the areas of an eye-diagram used to calculate measurements.
P values. The P values include the mean and standard deviation of the vertical location of PTop and PBase. These areas are
used with a specified sample size to statistically measure the following values
1
:
■
PTop Mean. This is the mean value of PTop.
■
PTop Sigma. This is the standard deviation of PTop.
■
PBase Mean. This is the mean value of PBase within the Eye Aperture.
■
PBase Sigma. This is the standard deviation of PBase within the Eye Aperture.
1
The Eye Aperture defaults to the center 20% of the interval from TCross
1
to TCross
2
.
Oscilloscope reference
740
DPO70000SX, MSO/DPO70000DX, MSO/DPO70000C, DPO7000C, and MSO/DPO5000B Series
Summary of Contents for DPO2304SX
Page 1: ...Tektronix Digital Phosphor Oscilloscopes Printable Help P077006219 077 0062 19...
Page 2: ......
Page 3: ...Tektronix Digital Phosphor Oscilloscopes Printable Help www tek com 077 0062 19...
Page 32: ...Introduction xxviii DPO70000SX MSO DPO70000DX MSO DPO70000C DPO7000C and MSO DPO5000B Series...
Page 130: ...Cursor setups 98 DPO70000SX MSO DPO70000DX MSO DPO70000C DPO7000C and MSO DPO5000B Series...
Page 198: ...MagniVu setup 166 DPO70000SX MSO DPO70000DX MSO DPO70000C DPO7000C and MSO DPO5000B Series...
Page 212: ...Display setups 180 DPO70000SX MSO DPO70000DX MSO DPO70000C DPO7000C and MSO DPO5000B Series...
Page 300: ...Math setups 268 DPO70000SX MSO DPO70000DX MSO DPO70000C DPO7000C and MSO DPO5000B Series...
Page 408: ...MyScope setups 376 DPO70000SX MSO DPO70000DX MSO DPO70000C DPO7000C and MSO DPO5000B Series...
Page 510: ...Trigger setups 478 DPO70000SX MSO DPO70000DX MSO DPO70000C DPO7000C and MSO DPO5000B Series...
Page 548: ...Vertical setups 516 DPO70000SX MSO DPO70000DX MSO DPO70000C DPO7000C and MSO DPO5000B Series...
Page 605: ...Utilities DPO70000SX MSO DPO70000DX MSO DPO70000C DPO7000C and MSO DPO5000B Series 573...
Page 606: ...Utilities 574 DPO70000SX MSO DPO70000DX MSO DPO70000C DPO7000C and MSO DPO5000B Series...
Page 920: ...Index 888 DPO70000SX MSO DPO70000DX MSO DPO70000C DPO7000C and MSO DPO5000B Series...