Taking cursor measurements
Use the following procedure to take measurements with cursors:
1. To open the Cursor Controls window, press the Cursors button on the front panel, click Cursors on the toolbar, or select
Cursor Setup from the Cursors menu.
NOTE. If cursors are on but the readouts are not attached to the cursor controls, press the Cursors button to attach the
readouts to the cursor controls. Press the Cursors button again to turn the cursors off.
2. Select the waveform you want to measure and a
from the control window.
3. Position the cursors with the multipurpose knobs.
4. Read the cursor information from the on-screen Position controls.
TIP. Select the cursor mode (Tracking or Independent) in the Cursor Setup control window.
NOTE. Cursor measurements are not available on digital channels.
Taking automatic measurements
Use the following procedures to quickly take a measurement based on the default settings for High and Low reference levels.
The instrument must be installed and operating, with a signal connected to an input channel, and the horizontal and vertical
controls and triggering set up.
1. Obtain a stable display of the waveform to be measured.
2. From the toolbar, click Measure to open the Measurement Setup control window.
3. Click the appropriate waveform button for the Channels <1– 4> and Ch, Math, or Ref Source tab.
4. Select the tab on the Measurement Setup control window that contains the measurement you want to take:
■
Ampl
■
Time
■
More
■
Histog
■
Comm
5. Click a button for the Measurements type.
The readout for the measurement and associated statistics is automatically displayed, and the measurement is added to the
Measurements list in the control window.
NOTE. In roll mode, measurements are not available until after you stop acquisitions.
NOTE. Automatic measurements are not available on digital channels.
How to ?
DPO70000SX, MSO/DPO70000DX, MSO/DPO70000C, DPO7000C, and MSO/DPO5000B Series
595
Summary of Contents for DPO2304SX
Page 1: ...Tektronix Digital Phosphor Oscilloscopes Printable Help P077006219 077 0062 19...
Page 2: ......
Page 3: ...Tektronix Digital Phosphor Oscilloscopes Printable Help www tek com 077 0062 19...
Page 32: ...Introduction xxviii DPO70000SX MSO DPO70000DX MSO DPO70000C DPO7000C and MSO DPO5000B Series...
Page 130: ...Cursor setups 98 DPO70000SX MSO DPO70000DX MSO DPO70000C DPO7000C and MSO DPO5000B Series...
Page 198: ...MagniVu setup 166 DPO70000SX MSO DPO70000DX MSO DPO70000C DPO7000C and MSO DPO5000B Series...
Page 212: ...Display setups 180 DPO70000SX MSO DPO70000DX MSO DPO70000C DPO7000C and MSO DPO5000B Series...
Page 300: ...Math setups 268 DPO70000SX MSO DPO70000DX MSO DPO70000C DPO7000C and MSO DPO5000B Series...
Page 408: ...MyScope setups 376 DPO70000SX MSO DPO70000DX MSO DPO70000C DPO7000C and MSO DPO5000B Series...
Page 510: ...Trigger setups 478 DPO70000SX MSO DPO70000DX MSO DPO70000C DPO7000C and MSO DPO5000B Series...
Page 548: ...Vertical setups 516 DPO70000SX MSO DPO70000DX MSO DPO70000C DPO7000C and MSO DPO5000B Series...
Page 605: ...Utilities DPO70000SX MSO DPO70000DX MSO DPO70000C DPO7000C and MSO DPO5000B Series 573...
Page 606: ...Utilities 574 DPO70000SX MSO DPO70000DX MSO DPO70000C DPO7000C and MSO DPO5000B Series...
Page 920: ...Index 888 DPO70000SX MSO DPO70000DX MSO DPO70000C DPO7000C and MSO DPO5000B Series...