Horizontal trigger position
Horizontal position is an adjustable feature that defines where the trigger occurs on the waveform record. It lets you choose how
much the instrument acquires before and after the trigger event. The part of the record that occurs before the trigger is the
pretrigger portion. The part that occurs after the trigger is the posttrigger portion. A longer posttrigger period may be useful when
you want to see the effects an event has on your system under test.
Pretrigger data can be valuable when troubleshooting. For example, if you are trying to find the cause of an unwanted glitch in
your test circuit, you can trigger on the glitch and make the pretrigger period large enough to capture data before the glitch. By
analyzing what happens before the glitch, you may uncover information that helps you find the source of the glitch.
Delayed trigger system
You can trigger with the A (Main) trigger system alone or you can combine the A (Main) trigger with the B (Delayed) trigger to
trigger on sequential events. When using sequential triggering, the A trigger event arms the trigger system, and the B trigger
event triggers the instrument when the B trigger conditions are met.
A and B triggers can (and typically do) have separate sources. The B trigger condition can be based on a time delay or a
specified number of counted events.
For more details on delayed triggering, click
.
Learn about advanced triggering.
Advanced triggering
You can check the advanced trigger status in the readout. The readout indicates the trigger type and then shows sources, levels,
or any other parameters that are important for the particular trigger type.
Oscilloscope reference
668
DPO70000SX, MSO/DPO70000DX, MSO/DPO70000C, DPO7000C, and MSO/DPO5000B Series
Summary of Contents for DPO2304SX
Page 1: ...Tektronix Digital Phosphor Oscilloscopes Printable Help P077006219 077 0062 19...
Page 2: ......
Page 3: ...Tektronix Digital Phosphor Oscilloscopes Printable Help www tek com 077 0062 19...
Page 32: ...Introduction xxviii DPO70000SX MSO DPO70000DX MSO DPO70000C DPO7000C and MSO DPO5000B Series...
Page 130: ...Cursor setups 98 DPO70000SX MSO DPO70000DX MSO DPO70000C DPO7000C and MSO DPO5000B Series...
Page 198: ...MagniVu setup 166 DPO70000SX MSO DPO70000DX MSO DPO70000C DPO7000C and MSO DPO5000B Series...
Page 212: ...Display setups 180 DPO70000SX MSO DPO70000DX MSO DPO70000C DPO7000C and MSO DPO5000B Series...
Page 300: ...Math setups 268 DPO70000SX MSO DPO70000DX MSO DPO70000C DPO7000C and MSO DPO5000B Series...
Page 408: ...MyScope setups 376 DPO70000SX MSO DPO70000DX MSO DPO70000C DPO7000C and MSO DPO5000B Series...
Page 510: ...Trigger setups 478 DPO70000SX MSO DPO70000DX MSO DPO70000C DPO7000C and MSO DPO5000B Series...
Page 548: ...Vertical setups 516 DPO70000SX MSO DPO70000DX MSO DPO70000C DPO7000C and MSO DPO5000B Series...
Page 605: ...Utilities DPO70000SX MSO DPO70000DX MSO DPO70000C DPO7000C and MSO DPO5000B Series 573...
Page 606: ...Utilities 574 DPO70000SX MSO DPO70000DX MSO DPO70000C DPO7000C and MSO DPO5000B Series...
Page 920: ...Index 888 DPO70000SX MSO DPO70000DX MSO DPO70000C DPO7000C and MSO DPO5000B Series...