background image

To use.

Click in an entry box, and then use the multipurpose knobs to set the Center Freq , Freq Span, and Res BW.

Click Full to set the frequency span to full.

 

 

Math setups

 

252

 DPO70000SX, MSO/DPO70000DX, MSO/DPO70000C, DPO7000C, and MSO/DPO5000B Series

Summary of Contents for DPO2304SX

Page 1: ...Tektronix Digital Phosphor Oscilloscopes Printable Help P077006219 077 0062 19...

Page 2: ......

Page 3: ...Tektronix Digital Phosphor Oscilloscopes Printable Help www tek com 077 0062 19...

Page 4: ...s of Tektronix Inc FastFrame OpenChoice iCapture MyScope MultiView Zoom SignalVu TekExpress TriMode TekSecure TekProbe TekVPI TekVISA MagniVu DPX and PinPoint are trademarks of Tektronix Inc Microsoft...

Page 5: ...x oscilloscope 22 Windows environment 22 Oscilloscope interface 27 Basic operations 33 Multi instrument configuration 46 DPO7AFP Auxiliary Front Panel optional 61 Waveform operations 62 Controls and c...

Page 6: ...on 101 Turn on or off the digital channels D15 D8 and D7 D0 103 Set the display size of logic channels 104 Set the global threshold for digital channels 104 Set up digital channels 105 Bus setups Bus...

Page 7: ...Select the RS 232 bus data bits 142 Select the RS 232 bus parity 143 Select the RS 232 bus bit rate 143 Select the SPI bus framing 144 Set the SPI bus idle time 145 Set the SPI bus word size 145 Sele...

Page 8: ...al and acquisition setups Horizontal Acquisition control window Horizontal tab 181 Set record length limit 182 Horizontal digital control window 183 Horizontal Position Scale control window 184 Horizo...

Page 9: ...om the programming interface 215 Mask testing setups Mask setup 217 Mask testing control window Masks tab 217 Serial mask testing option 218 Select a mask type 218 Mask types and standards 219 Use mas...

Page 10: ...th averaging 246 Set math variables 247 Custom analysis functions 247 Use custom analysis functions 248 MATLAB custom analysis interface 248 Basic spectral setups 249 Set up basic spectral analysis 24...

Page 11: ...control window Communications tab 278 Select a communications measurement 280 Measurement setup control window More tab 282 Select a miscellaneous measurement 284 Counter measurement setup 285 Set up...

Page 12: ...43 Set up FLEXRAY search parameters 346 Set up MIL 1553 search parameters 350 Set up ethernet search parameters 353 Set up DDR search parameters 359 Set up custom search parameters 361 Set up parallel...

Page 13: ...er 414 Set up a Freq Period trigger 415 Set up a glitch trigger 417 Set up a logic pattern trigger 420 Set up a logic pattern trigger 421 Set up a runt trigger 421 Set up a serial pattern trigger 423...

Page 14: ...gger position correction 476 Vertical setups Set up vertical controls 479 Set up input channels 481 Zoom control window 483 Waveform display control window 484 Set the Position Offset 485 Waveform lab...

Page 15: ...her applications 522 Copying measurements for use in other applications 522 Printing 523 Page setup dialog box 523 Print preview dialog box 523 Print dialog box 524 Printing the screen to a printer 52...

Page 16: ...550 Touch screen 552 Instrument calibration control window 552 Instrument diagnostics control window 554 E mail on event control window 555 Measurement save options 556 Multipurpose knobs 558 User pr...

Page 17: ...2 Set up an RS 232 serial bus 582 Set up a USB serial bus 582 Set up a MIPI DSI 1 serial bus 583 Set up a MIPI CSI 2 serial bus 583 Set up an 8B10B serial bus 583 Set up a custom serial bus 584 Set up...

Page 18: ...fined mask 607 Save a user mask to disk 608 Recall a user mask from disk 608 Create a user mask with mask testing 608 Use mask testing 608 Using e mail on event 609 Send an e mail on events 609 Creati...

Page 19: ...e 631 Trigger on setup hold time violations 632 Trigger on window threshold violations 634 Trigger on a video signal 635 Trigger on a communication signal 635 Trigger on a bus 636 Mark all trigger eve...

Page 20: ...e display contrast 648 Oscilloscope reference Functional model 649 Process overview 650 Waveform acquisition 651 Signal connection 651 Probes and signal connection 651 Input conditioning 652 Bandwidth...

Page 21: ...ransition time trigger 673 Video trigger 673 Width trigger 674 Window trigger 674 Sequential triggering 674 Triggering with horizontal delay off 674 Triggering with horizontal delay on 675 Triggering...

Page 22: ...tom functions 714 Spectral math waveforms 716 Spectral math controls 716 Spectral math time controls 717 Spectral math gating controls 719 Spectral math frequency domain controls 720 Spectral math mag...

Page 23: ...al type 745 Center frequency 745 Channels selector 746 Clear button 746 Clear data control 746 Clock edge 746 Level 746 Communication trigger codes and standards 746 Set up communication trigger codin...

Page 24: ...757 Enter math equation editor Operands 757 Set up e mail on trigger 758 Extinction ratio 758 Extinction ratio decibel 758 Eye base 758 Eye diagram 758 Eye height 758 Eye top 759 Eye width 759 Set up...

Page 25: ...izontal Position Scale control window Delay mode 768 Set the horizontal position 768 Set the horizontal scale 768 Horizontal reference marker 769 Horizontal delay mode 769 Set the horizontal delay 769...

Page 26: ...trigger level 784 Set mask autoset vertical controls 784 Mask hit count 785 Math arbitrary filters 785 Enter a math expression 793 Display a math equation 793 Set the math display and vertical control...

Page 27: ...Select trigger polarity 814 Positive duty cycle 814 Positive overshoot 815 Positive width 815 Predefined functions 815 Probe calibration output 815 Probe calibration restrictions 815 Probe compensati...

Page 28: ...data source channel 828 Set the data threshold level 828 Set the setup and hold times 828 Enter a spectral math equation 828 Enter a spectral math expression 828 Spectral tracking 830 Set up the spec...

Page 29: ...gger width limits 839 Set the trigger polarity 839 Using FastFrame acquisitions 840 Use print screen 840 Analyze menu 840 Cursors menu 840 Digital menu 841 Display menu 841 Edit menu 841 File menu 841...

Page 30: ...et display characteristics 857 Set the offset 858 Set the vertical position scale 859 Select the probe controls 861 Select the termination 862 Set the termination voltage 863 Video trigger autoset 864...

Page 31: ...and a touch screen interface on a Microsoft Windows platform Choose the methods of controlling oscilloscope operations that are most convenient for you Customary front panel controls Menus and toolbar...

Page 32: ...Introduction xxviii DPO70000SX MSO DPO70000DX MSO DPO70000C DPO7000C and MSO DPO5000B Series...

Page 33: ...include Bandwidth DPO77002SX 70 GHz bandwidth and 200 GS s on 1 analog channel or 33 GHz bandwidth and 100 GS s on 2 analog channels DPO75902SX 59 GHz bandwidth and 200 GS s on 1 analog channel or 33...

Page 34: ...tomatic measurements eye pattern and measurement statistics Basic math as well as advanced equation editor and spectral analysis Customizable MyScope control window An intuitive graphical user interfa...

Page 35: ...time histogram and communication measurements You can customize the measurements by changing reference levels or by adding measurement gating Statistics mean maximum minimum standard deviation and pop...

Page 36: ...it is recommended that you exit the instrument application only when installing new software Support Software The instrument contains additional software and files that may be useful to you Readme fi...

Page 37: ...016 1977 XX Service manual as a PDF file 077 0076 XX DPO70000SX DPO7AFP Auxiliary Front panel 016 2095 xx Rackmount Kit 016 2102 xx SSD mounting kit front of instrument rackmount tray 016 2104 00 Tra...

Page 38: ...ial voltage probe with 2 V differential input voltage TPP0502 500 MHz 2X TekVPI passive voltage probe TPP0850 2 5 kV 800 MHz TekVPI high voltage passive probe TPP500 TPP1000 10X Passive Probes are com...

Page 39: ...TekVPI differential voltage probe with 42 V differential input voltage TDP1000 1 GHz TekVPI differential voltage probe with 42 V differential input voltage TDP1500 1 5 GHz TekVPI differential voltage...

Page 40: ...z with optional tip P7630 TriMode probe 30 GHz TCA BNC Adapter TekConnect to BNC TCA N Adapter TekConnect to N TCA SMA Adapter TekConnect to SMA TCA 1MEG High impedance buffer amplifier Includes P6139...

Page 41: ...tial probe 12 5 GHz P7380 P7380SMA Differential probe 8 GHz P7504 TriMode probe 4 GHz P7506 TriMode probe 6 GHz P7508 TriMode probe 8 GHz P7513A TriMode probe 13 GHz P7516 TriMode probe 15 GHz P7520A...

Page 42: ...on for MOST50 and MOST150 Opt MSOE 16 channel MSO 1 Opt MTM Mask Testing for Serial Communication Standards up to 1 5 Gb s Opt PS2 Power Solution Bundle DPOPWR THDP0200 TCP0030A 067 1686 XX Deskew Fix...

Page 43: ...Ethernet Compliance Test Opt D PHY MIPI D PHY Characterization and Compliance Testing Opt HSIC USB HSIC protocol decode and electrical validation For models of bandwidth 2 5 GHz only requires DJA Opt...

Page 44: ...O Flexible OFDM Analysis requires SVE Opt SVP SignalVu Pulse Advance Signal Analysis Software requires SVE Opt SVT Settling Time Measurements such as Frequency and Phase requires SVE Opt USB Universal...

Page 45: ...l standards 4 25 GB s standard on MSO instruments Opt MOST Electrical Compliance and Debug Test Solution for MOST50 and MOST150 Requires Opt DJA Opt PCE3 PCI Express Transmitter Compliance and Debug R...

Page 46: ...Bus USB 2 0 Compliance Test Opt USBHSIC USB2 0 and HSIC Bundle includes options Opt DJA HSIC SR USB and USB Opt USBPWR USB Power Adapter EPS Compliance Automated Test Solution Not supported on 70000DX...

Page 47: ...re requires SVE Opt SVT Settling Time Measurements such as Frequency and Phase requires SVE Opt SV23 WLAN802 11a b g j p measurement application requires Opt SVE Opt SV24 WLAN 802 11n measurement appl...

Page 48: ...drive For quick access to PDF files of the user manual probe manuals programmer guide primers and applications specifications and performance verification manual service and installation manuals and...

Page 49: ...e user that runs the utility This allows TekScope to generate fresh files the next time it starts Finding the utility A link to the utility can be found under the Start All Programs Tektronix TekScope...

Page 50: ...d report file will also contain copies of all files found in the following directories C programdata Tektronix TekScope Internal C programdata Tektronix TekScope Calibration CalDramDump C programdata...

Page 51: ...as Lab Windows and LabVIEW Click the following tools to find out how they can help you connect to your instrument and analyze the data that it collects VXI Plug and Play Instrument Driver TekVISA TekV...

Page 52: ...he Visual BASIC for Applications macro editor built into Excel Source code for the toolbar is accessible through Excel s built in VBA editor for use as a learning tool for using the TekVISA ActiveX Co...

Page 53: ...key details of the applications that are installed in the instrument are also listed On MSO DPO70000C models About TekScope also indicates if TekProtect is Installed or Not Installed on the instrument...

Page 54: ...al monitor attached to the instrument avoid changing the taskbar properties to Always on top and Auto hide The Windows taskbar may obscure buttons or readouts on the screen However if you do have an e...

Page 55: ...h the instrument powered off TIP Consult your network administrator for further assistance You may want to print these instructions before you proceed 1 Attach a keyboard and mouse to the instrument 2...

Page 56: ...g the LAN Reset button does a LAN configuration initialization enables DHCP enables Auto IP enables ICMP ping responder sets the Web password to default No password and updates the LXI XML identificat...

Page 57: ...changes to some of the instrument configuration Click on the Help SUPPORT link to see help on the configuration information Click on Datasheets Manuals or Driver Download to go to the Tektronix Oscill...

Page 58: ...int the full screen or just the graticule area The instrument uses Windows print drivers so the process to install printers is the same as it is on any other PC The Page Setup dialog box contains some...

Page 59: ...tput Aux Out to obtain a TTL compatible negative polarity pulse when the oscilloscope triggers or when certain other events occur such as a mask test failure or completion Use the rear panel external...

Page 60: ...o do next Display help on the File menu Display help on the Edit menu Display help on the Vertical menu Display help on the Digital menu Display help on the Horizontal Acquisition Horiz Acq menu Displ...

Page 61: ...f the screen You can also customize the toolbar to add menus you use more frequently Click the Menu button with the arrow to see the list of menus you can add and to access the Customize Toolbar dialo...

Page 62: ...e menu buttons that you use more frequently To access some of the less common controls click the Menu Bar Mode To do this click the Menu button with the arrow in the upper right corner of the screen a...

Page 63: ...Waveform handle Click and drag to change the waveform vertical position 6 Controls status Quick reference to vertical selections scale and parameters Probe status is provided for some probe tips TekCo...

Page 64: ...Learn how to contact Tektronix Support Provide feedback on this Tektronix product Provide information about Tekscope Navigating the online help To open the online help select Help and then select Con...

Page 65: ...discharges of static electricity directly into the signal input may damage the instrument To avoid damage to the instrument use the following techniques to prevent electrostatic discharge to the inst...

Page 66: ...from all cables before connecting them to the instrument or device under test by momentarily grounding the center conductor of the cable or by connecting a 50 termination to one end before attaching...

Page 67: ...unding the oscilloscope is necessary to ensure safety and to take accurate measurements The oscilloscope must share the same ground as any circuits that you plan to test Observe maximum input voltage...

Page 68: ...o keep dust out Before each use do the following 1 Verify that the connectors are free of dirt metal particles scratches and deformations 2 Verify the connectors are the correct size WARNING To avoid...

Page 69: ...is purpose Use the ground connection with the black plastic grommet as this has a 1 M series isolation resistance so that there is no chance of shock but the static voltage can be discharged Figure 1...

Page 70: ...stalling a cable CAUTION Failure to prevent rotation of the center conductors when making connections to the ATI input will damage your instrument If necessary use a second backing wrench on the cable...

Page 71: ...Figure 4 Use a second backing wrench About Tektronix oscilloscopes DPO70000SX MSO DPO70000DX MSO DPO70000C DPO7000C and MSO DPO5000B Series 39...

Page 72: ...nother decimal place for greater accuracy This online help usually refers to using the multipurpose knobs to change control settings although the following methods are generally available Some paramet...

Page 73: ...Pop up keypad 1 Double click in a data entry box or Click in a data entry box Click on the keypad icon 2 Use the pop up keypad to enter data About Tektronix oscilloscopes DPO70000SX MSO DPO70000DX MSO...

Page 74: ...Sample Rate Keypad 3 Click the Enter Button Other keypads 1 8b10b Character Keypad About Tektronix oscilloscopes 42 DPO70000SX MSO DPO70000DX MSO DPO70000C DPO7000C and MSO DPO5000B Series...

Page 75: ...button to grab the handle Hold the button and drag with the mouse to another position on the screen You can move freely to any position on the screen or to a second monitor if you have set up a dual d...

Page 76: ...arker to set the trigger level You can also click and drag or touch and drag across part of a waveform to select it and investigate it further Once the selection is made choose one of the following ac...

Page 77: ...instrument with your next task This may be especially important if you share the instrument with others previous settings may make it difficult for you to perform the required task To quickly restore...

Page 78: ...ulti instrument configuration to create an up to 4 channel highest bandwidth and highest sample rate system with synchronized trigger and sample clock About Tektronix oscilloscopes 46 DPO70000SX MSO D...

Page 79: ...ave space and allow shorter cables and more convenient connections WARNING When using an instrument upside down be careful to avoid pinching your fingers About Tektronix oscilloscopes DPO70000SX MSO D...

Page 80: ...o use when stacking an OM4000 on the instrument Use the rotated orientation of the top feet to stack two instruments on top of each other When stacking two instruments the bottom feet sit in the holes...

Page 81: ...instrument configuration the instrument powers on in its stand alone mode not multi instrument mode UltraSync bus cable An UltraSync bus cable will connect any multi instrument configuration capable...

Page 82: ...2 Connect the clock cable to the 12 5 GHz Clock Out connector 8 in lbs torque 3 Connect the Trigger cable to the Trigger connector 4 Connect the PCIe cable to the PCIe connector Connect UltraSync bus...

Page 83: ...About Tektronix oscilloscopes DPO70000SX MSO DPO70000DX MSO DPO70000C DPO7000C and MSO DPO5000B Series 51...

Page 84: ...om left to right Extension B C and then D NOTE Connect the extensions to the master in a star pattern do not daisy chain the instruments If powering off a multi instrument system power off the master...

Page 85: ...onfiguration NOTE After successful power on of your multi instrument system wait for the warm up time to complete and then run Signal Path Compensation The following image shows an extension with a co...

Page 86: ...The following images show the master and extension with valid connections About Tektronix oscilloscopes 54 DPO70000SX MSO DPO70000DX MSO DPO70000C DPO7000C and MSO DPO5000B Series...

Page 87: ...its status and timer display Pause the timer if you have additional changes to make before the oscilloscope application starts Press Start Scope to bypass the timer When the timer runs down the oscill...

Page 88: ...e Master If you plan to use Time Synchronized mode contact your Tektronix representative for additional information After the multi instrument connections are verified the instruments display their op...

Page 89: ...About Tektronix oscilloscopes DPO70000SX MSO DPO70000DX MSO DPO70000C DPO7000C and MSO DPO5000B Series 57...

Page 90: ...ertical Configuration menu select either ATI TekConnect or TimeSync 2 Click Yes to continue Restart of a multi instrument system To restart your multi instrument system without powering down the oscil...

Page 91: ...hen using the ATI channel the TekConnect channels are not available Multi instrument status displays In a multi instrument configuration the display shows status information such as channel numbers on...

Page 92: ...nels Mapped channels Mapped channels Pattern Boolean Not available Not available Not available Not available Not available Setup Hold Clock Data Not available Not available Not available Not available...

Page 93: ...til the Windows OS recognizes and loads the required drivers 3 Start the TekScope application The panel LEDs will light when TekScope connects to the panel 4 To verify that the DPO7AFP is working push...

Page 94: ...channel In either case simple or complex there is just one trigger event The time base can be either delayed or not delayed If it is not delayed acquisition occurs surrounding the trigger event depend...

Page 95: ...detail you can also use horizontal delay as described in the previous topic to focus the acquisition on a segment of a waveform separated from the trigger event by a significant interval of time This...

Page 96: ...al delay you can delay the acquisition to a point well beyond the trigger event similar in effect to a delayed time base By alternately turning horizontal delay on and off you can make quick detailed...

Page 97: ...econd The digital phosphor oscilloscope shows you not only the displayed waveform but also the sample density You can choose to see sample density represented in a color palette or as shades of gray o...

Page 98: ...earn about bus setups Learn about digital setups Using waveform math Waveform math in the instrument can be as simple as basic arithmetic or as powerful as building a complicated math expression Equat...

Page 99: ...as numerical statistics or graphically in a histogram cycle trend time trend or spectral plot Ethernet Compliance Testing Use the ET3 software to perform 10 100 1000 Base T Ethernet compliance testing...

Page 100: ...About Tektronix oscilloscopes 68 DPO70000SX MSO DPO70000DX MSO DPO70000C DPO7000C and MSO DPO5000B Series...

Page 101: ...el overview The instrument features the following front panel controls and connectors DPO70000SX Series instruments only have front panel connectors and the On Standby switch on the front panel DPO700...

Page 102: ...Cal output 13 Logic probe input inactive on DPO models MSO capabilities are available as a user upgradeable option on DPO70000DX instruments On Standby switch 14 Clear button 15 Save Recall buttons 1...

Page 103: ...probe input inactive on DPO models MSO capabilities are available as a user upgradeable option on DPO70000DX instruments On Standby switch 14 Clear button 15 Save Recall buttons 16 Digital Channel set...

Page 104: ...tput Probe Calibration output Use the Probe Calibration output to calibrate or deskew passive or active voltage probes Channel inputs You can connect up to four probes to the inputs of the instrument...

Page 105: ...rument application and shut down in one step select Shutdown in the File menu Touch screen You can use touch to control the instrument in addition to the front panel controls toolbar buttons and menu...

Page 106: ...orm Panning a waveform Rotate the pan outer knob of the pan zoom controls to pan the waveform Turn the knob clockwise to pan forward Turn it counterclockwise to pan backwards The further you turn the...

Page 107: ...ay scale The range and resolution of the scale value depends on the attached probes and on any other external factors you have specified For reference or math waveforms the scale setting controls the...

Page 108: ...uire waveform details that are separated from the trigger event by a significant interval of time NOTE The Delay button is not available on MSO DPO5000B Series instruments Trigger controls Use the Tri...

Page 109: ...the Level knob to set the trigger level to 50 To set a different trigger level rotate the Level knob The trigger level appears on the screen The trigger level knob affects Edge Width Glitch Timeout Se...

Page 110: ...tes signals below 80 kHz before passing the signal to the trigger circuitry Noise Rej This coupling provides stable triggering by increasing the trigger hysteresis Increased hysteresis reduces the tri...

Page 111: ...General purpose controls Use the following front panel controls to perform general purpose tasks such as to Autoset the instrument to send items to the printer or to enable cursors Controls Descripti...

Page 112: ...eform High intensity settings show less frequently acquired points more clearly MultiView zoom controls Push the front panel MultiView Zoom button to do the following Controls Description Open a magni...

Page 113: ...re normal operation If the instrument front panel and or touch screen do not respond press the On Standby switch for 5 seconds to cycle power Use this VGA port to connect a monitor for extended deskto...

Page 114: ...n a TTL compatible negative polarity pulse when the instrument triggers Use the CH 3 OUT connector to connect to a buffered version of the signal that is attached to the Channel 3 input The amplitude...

Page 115: ...ronize the instrument time base to an external 9 8 10 2 MHz reference frequency source When not using the external reference disconnect signals from the Ext Ref In BNC connector For MSO DPO5000B instr...

Page 116: ...O70000SX instruments Use the Display Port to connect to a monitor Use the PCIe ports to connect to a PC or to set up a multi instrument configuration Use the Trigger ports to set up a multi instrument...

Page 117: ...re that at least one analog channel be displayed Alternatively to take a measurement of a digital channel you can first send the waveform acquired on the digital channel to an analog channel using iCa...

Page 118: ...play format you can also select the type of readout from the XY Readout Type drop down list Only Waveform and Screen cursors are available in the XY or XYZ display format TIP The Cursors menu Cursors...

Page 119: ...one of the Cursor Type buttons to select a cursor type TIP Select a cursor type directly from the Cursors menu Cursor Type submenu Behavior If you are using H Bar or V Bar Cursors and want to measure...

Page 120: ...Cursor 1 source sets the Cursor 2 source to the same value To have different sources for Cursor 1 and Cursor 2 set the Cursor 1 source first and then set the Cursor 2 source NOTE Only Screen Cursors h...

Page 121: ...only the X when used with waveform cursors in XY display mode The vertical line is not drawn Symbol style Cursor type Display format Cursor position Lines Waveform and Screen YT Intersection of horizo...

Page 122: ...ical value There are six readouts for Waveform cursors and three for Screen cursors Ratio Waveform and Screen XY Shown as rectangular coordinates and the Y vertical value is divided by X vertical valu...

Page 123: ...nu highlight Cursor Type and then select H bars from the submenu 2 Use the multipurpose knobs to move the cursors 3 Use the cursor Source drop down list boxes to select the waveform you want to measur...

Page 124: ...the boundaries of the vertical bar cursors rather than use the entire display width What do you want to do next Learn more about the Cursor controls Go to a step by step procedure for taking cursor m...

Page 125: ...e or frequency Readouts display measurement parameters for both cursors Vertical readouts display voltage values for each cursor with respect to the zero reference point and with respect to each other...

Page 126: ...you can first send the waveform acquired on the digital channel to an analog channel using iCapture This online help supports many oscilloscope models from Tektronix This feature is only available on...

Page 127: ...ontal readout both in the control window and in the readout at the bottom of the screen The following relationships hold Time from First Point Horiz Position when Horiz Delay and Ref Position are zero...

Page 128: ...tep by step procedure for taking automatic measurements Cursor setup control window XY display format From the Cursors menu select Cursor Setup Overview Use these controls to set the cursor parameters...

Page 129: ...tions Waveform cursors display nine rectangular readouts X Y t X1 X2 Y1 Y2 t1 and t2 Screen cursors display six readouts X Y X1 X2 Y1 and Y2 Polar Shows the cursor position as polar coordinate positio...

Page 130: ...Cursor setups 98 DPO70000SX MSO DPO70000DX MSO DPO70000C DPO7000C and MSO DPO5000B Series...

Page 131: ...ls Use the D15 D8 tab to set up the upper eight bits and the D7 D0 tab to set up the lower eight bits For information on the controls click the buttons Threshold Position Size Global Threshold D15 D0...

Page 132: ...log characteristics of a digital waveform Learn how to trigger on a bus Learn how to track down serial bus anomalies Learn how to troubleshoot circuits using a parallel bus Learn how to troubleshoot a...

Page 133: ...menu What do you want to do next Learn about setting up the digital channels Learn about bus setups Learn about bus configuration Digital channel threshold and position NOTE The Digital Channel Thresh...

Page 134: ...ultipurpose knobs to select a voltage level Behavior The voltage threshold level determines when a voltage level on a digital channel is considered to be a logic high and a logic low Set digital chann...

Page 135: ...D15 D8 On button to turn on the digital channels D0 to D7 or D8 to D15 respectively Click the Threshold entry box and use the keypad to enter a value for the voltage level or use the multipurpose knob...

Page 136: ...mall displays the channel as a quarter of a division Extra Small displays the channel as a fifth of a division What do you want to do next Learn about setting up digital signals Learn more about digit...

Page 137: ...ion is available only on MSO5000B and MSO70000C DX Series instruments and DPO5000B Series instruments with option MSOE installed and DPO70000DX instruments with option MSOU installed From the Digital...

Page 138: ...es to match the color on the digital channel button What do you want to do next Learn about bus setups Learn about bus configuration Learn more about digital setups Learn how to turn on or turn off th...

Page 139: ...re decode and display data You can set up serial buses for the SPI I2C USB RS232 8b10b and MIPI serial data standards You can also set up clocked or unclocked parallel buses NOTE The instrument displa...

Page 140: ...of the available options from the drop down list the instrument provides additional parameters to set up for each standard Behavior Click the Clear Bus button to clear all the channels from the selec...

Page 141: ...til the desired Bus number or name is visible Click the Bus number B1 B16 or name to be set up Behavior Click the Bus button to display the bus On or to remove it from the display Off What do you want...

Page 142: ...he scroll wheel or click the up down arrows until the desired Bus number or name is visible Click the Bus number B1 B16 or name to be set up Click OK Behavior Click the OK to return to the Search Conf...

Page 143: ...the Clear Bus button the instrument removes the bus with the channels used to build the bus and the Bus button changes to Off What do you want to do next Learn more about bus setups Learn about bus c...

Page 144: ...rument also displays the label adjacent to the bus number on the screen for easy identification What do you want to do next Learn more about bus setups Learn about bus configuration Learn about digita...

Page 145: ...t to do next Learn more about bus setups Learn about bus configuration Learn about digital setups Set up an I2C serial bus NOTE I2C serial bus setup is available only on instruments with the SR EMBD o...

Page 146: ...l setups Learn how to set up an I2C serial bus Learn how to set up an SPI serial bus Learn how to set up a USB serial bus Learn how to set up an RS 232 serial bus Learn how to set up a MIPI DSI1 Seria...

Page 147: ...o use To change the input for a component click the Input and select a channel from the cascading lists To change the voltage threshold level click the Threshold entry box and use the keypad to enter...

Page 148: ...n how to set up a FLEXRAY Serial Bus Learn how to set up a MIL 1553 Bus Learn how to set up an Ethernet Bus Learn how to set up an Custom serial bus Learn how to troubleshoot a serial bus Learn how to...

Page 149: ...om the drop down list To set the bit rate click the Bit Rate entry box and select from the drop down list Behavior The instrument changes the components voltage threshold levels polarity data bits par...

Page 150: ...NOTE USB serial bus setup is available only on instruments with the SR USB option From the Bus Setup window in the Config tab select a bus from the Bus list and then select the USB Serial bus type NOT...

Page 151: ...ype and uses those to decode the selected USB bus What do you want to do next Learn more about bus setups Learn about bus configuration Learn about digital setups Learn how to set up an I2C serial bus...

Page 152: ...channel type click the Channel Type entry box and select from the drop down list To change the input for a component click the Input and select a channel from the cascading lists To change the voltage...

Page 153: ...s Learn how to track down serial bus anomalies Learn how to view analog characteristics of a digital waveform Learn how to trigger on a bus Learn to view the bus packets Set up a MIPI CSI 2 serial bus...

Page 154: ...e result Behavior The instrument changes the components voltage threshold levels polarity data bits parity and bit rate and uses those to decode the selected MIPI CSI 2 bus What do you want to do next...

Page 155: ...t Data Rate Input Serial Threshold To use To change the input for a component click the Input and select a channel from the cascading lists To change the voltage threshold level click the Threshold en...

Page 156: ...p a MIL 1553 Bus Learn how to set up an Ethernet Bus Learn how to set up an Custom serial bus Learn how to troubleshoot a serial bus Learn how to track down serial bus anomalies Learn how to view anal...

Page 157: ...setups Learn about bus configuration Learn about digital setups Learn how to set up an I2C serial bus Learn how to set up an SPI serial bus Learn how to set up a USB serial bus Learn how to set up an...

Page 158: ...op down list To change the voltage threshold level click the Threshold entry box and use the keypad to enter a value or use the multipurpose knobs to select a voltage level To change the voltage hyste...

Page 159: ...how to set up a PCIe Serial Bus Learn how to set up an Custom serial bus Learn how to troubleshoot a serial bus Learn how to track down serial bus anomalies Learn how to view analog characteristics of...

Page 160: ...Learn about bus configuration Learn about digital setups Learn how to set up an I2C serial bus Learn how to set up an SPI serial bus Learn how to set up a USB serial bus Learn how to set up an RS 232...

Page 161: ...hreshold level click the Threshold entry box and use the keypad to enter a value or use the multipurpose knobs to select a voltage level To set the polarity click the Polarity entry box and select fro...

Page 162: ...al Bus Learn how to set up an Custom serial bus Learn how to troubleshoot a serial bus Learn how to track down serial bus anomalies Learn how to view analog characteristics of a digital waveform Learn...

Page 163: ...more about bus setups Learn about bus configuration Learn about digital setups Learn how to set up an I2C serial bus Learn how to set up an SPI serial bus Learn how to set up a USB serial bus Learn ho...

Page 164: ...d level click the Threshold entry box and use the keypad to enter a value or use the multipurpose knobs to select a voltage level To select the polarity click the Polarity and select the polarity from...

Page 165: ...et up an Custom serial bus Learn how to troubleshoot a serial bus Learn how to track down serial bus anomalies Learn how to view analog characteristics of a digital waveform Learn how to trigger on a...

Page 166: ...Learn more about bus setups Learn about bus configuration Learn about digital setups Learn how to set up an I2C serial bus Learn how to set up an SPI serial bus Learn how to set up a USB serial bus Le...

Page 167: ...instruments only allow 1 lane What do you want to do next Learn more about bus setups Learn about bus configuration Learn about digital setups Select the serial bus channel type NOTE Serial bus Channe...

Page 168: ...rn more about bus setups Learn about bus configuration Learn about digital setups Select the serial bus channel input From the Bus Setup window select Serial from the Bus Type list and then select one...

Page 169: ...elect Serial from the Bus Type list and then select the USB Serial bus type To use To change the speed for the displayed component of the bus click Speed and select from the cascading list Behavior Th...

Page 170: ...m the cascading list Behavior The instrument acquires data from the channel you select for each component What do you want to do next Learn more about bus setups Learn about bus configuration Learn ab...

Page 171: ...s click the Input and select from the cascading list Select the USB bus threshold From the Bus Setup window select Serial from the Bus Type list and then select USB Serial bus type NOTE USB bus thresh...

Page 172: ...el Threshold option is available only on instruments with one or more of the SR options To use To change the voltage threshold level of the component Input click the Threshold entry box and use the ke...

Page 173: ...nal is at a low level Active High causes the instrument to decode the bus when the Slave Select signal is at a high level For the Data component Active Low causes the instrument to decode signals at a...

Page 174: ...able only on instruments with option SR COMP From the Bus Setup window select the Select tab and the RS 232 Serial bus type To use Scroll through the Bus list and select the RS 232 bus to configure Cl...

Page 175: ...There is no parity selection when the number of data bits is 9 What do you want to do next Learn more about bus setups Learn about bus configuration Learn about digital setups Select the RS 232 bus bi...

Page 176: ...Framing option is available only on instruments with option SR EMBD From the Bus Setup window select the Select tab and the SPI Serial bus type To use Select the framing from the drop down list Select...

Page 177: ...e selected SPI bus What do you want to do next Learn more about bus setups Learn about bus configuration Learn about digital setups Set the SPI bus word size NOTE The SPI Bus Word Size option is avail...

Page 178: ...and the SPI Serial bus type To use Scroll through the Bus list and select the SPI bus to configure Click the Bit Order box and select the direction from the drop down list MS First or LS First Behavio...

Page 179: ...der Remove Behavior If the bus is not clocked the instrument acquires all data from the parallel bus at the sample rate of the instrument Digital channels are not available on SX models What do you wa...

Page 180: ...ock Source select a channel from the cascading list Digital channels are not available on SX models To change the Clock Polarity select from Rising Falling or Either Behavior Rising causes the instrum...

Page 181: ...ources to a parallel bus From the Bus Setup window click the Add Sources Select button To use Click the button for each channel to add to the bus Click the OK button NOTE You have the option to add us...

Page 182: ...Channel Type is only available on instruments with an SR xxxx option From the Bus Setup window select a bus number from the Bus list To use Click the Thresholds Setup button Double click the Threshol...

Page 183: ...Up or Down arrows to change the bit order of the selected channel Behavior Moves the selected channel towards the MSB most significant bit position in the bit order when you click the arrow that point...

Page 184: ...r accordingly What do you want to do next Learn more about bus setups Learn about bus configuration Learn about digital setups Select the bus clock source channel From the Bus Setup window select a bu...

Page 185: ...he Parallel bus type and click Clocked To use Connect a probe between the instrument and the clock signal in your system under test Click the Clock Polarity selection box and select from Rising Fallin...

Page 186: ...et View Results Table Behavior The instrument displays data from the bus based on the selected Bus Components Busform Decode option and the symbol file For the LIN FLEXRAY MIL1553 MIPI DSI 1 MIPI CSI...

Page 187: ...isplay the decode of the bus in Symbol Character Symbol Character KCode Data Hex Data Binary Descrambled Hex and Descrambled Binary Select the decode from the drop down list For the RS232 serial bus y...

Page 188: ...Binary Select the decode form the drop down list What do you want to do next Learn about bus setups Learn about digital setups Learn how to set up digital waveforms Learn how to view analog characteri...

Page 189: ...by clicking on the box or box in the display Available views may include Transaction transport link framing packets physical bytes symbols and bit level Behavior Clicking on a box displays additional...

Page 190: ...he virtual channel and word count in decimal All other fields are displayed in hexadecimal Mixed decoding methods like symbol for DT Data Type MIPI element decimal for VC Virtual Channel MIPI element...

Page 191: ...tab To use Scroll through the Bus list B1 B16 and select the bus to configure Select the Display tab Click the Use Symbol File check box Click the Browse button to locate the file NOTE The instrument...

Page 192: ...nes are comments that describe the file and are not required as part of the file header After the comment lines if any the file must contain the TSF header line that starts with the characters and the...

Page 193: ...e in the symbol table file and displays the corresponding symbol If the instrument does not find the numeric value there is no symbol and the numeric value displays What do you want to do next See an...

Page 194: ...directory where the instrument expects to find it When you open a symbol table file the instrument uses the values to decode and display bus values to be easier to recognize NOTE The only available f...

Page 195: ...her resolution to accurately determine edge placement This helps you make precise timing measurements on digital edges You can see up to 32 times more detail than with normal digital channel sampling...

Page 196: ...With MagniVu option enabled the Sample Rate and Resolution values set MagniVu setup 164 DPO70000SX MSO DPO70000DX MSO DPO70000C DPO7000C and MSO DPO5000B Series...

Page 197: ...hese cases you can find the digital record by looking for the digital signal in the upper overview and panning accordingly NOTE You should turn on MagniVu when light gray shading is displayed to indic...

Page 198: ...MagniVu setup 166 DPO70000SX MSO DPO70000DX MSO DPO70000C DPO7000C and MSO DPO5000B Series...

Page 199: ...b controls to change the display style persistence intensity format and interpolation For information on the controls click the buttons Style Persistence Intensity Format Interpolation NOTE The follow...

Page 200: ...xt to provide notes for screen shots printouts or for other instrument users Click in a text field 1 through 8 Click the Text 1 8 button to turn the message text On or Off on the display Click the Fon...

Page 201: ...control window to position text to specific locations on the graticule To use Select the line of text from the Screen Text drop down list then use the X Pos and Y Pos controls to define the location...

Page 202: ...on on the controls click the buttons Graticule Style Trigger Level Marker To use Click the Date Time button to show the current date and time on the graticule Click the Trigger T button to turn this d...

Page 203: ...aveforms If you are using Waveform Database acquisition mode select the FastAcq WfmDB Palette For all other acquisition modes use the Record View Palette For information on the controls click the butt...

Page 204: ...ence for displaying record points that may occur outside the normal acquisition envelope Use Variable to accumulate record points on the waveform for a specified time interval Each record point decays...

Page 205: ...le selection has no affect when you using XY format When using this format the horizontal scale continues to control the time base and the horizontal position continues to control the portion of the w...

Page 206: ...nel channel waveformas ar assigned to the axis indicated in the following table XY Pair X Axis Y Axis Ch 1 and Ch 3 Ch 1 Ch 3 If you only display one source when you enter the XY format the instrument...

Page 207: ...of the backlight to the required option What do you want to do next Learn about using text with screen displays Learn about changing the appearance of objects in the display Learn about changing the d...

Page 208: ...t rise times such as pulse trains TIP Use the Display menu Waveform Interpolation submenu to directly access the interpolation options You may want to set the display style so that the real samples ar...

Page 209: ...are not needed Use IRE if you are working with NTSC video signals The instrument sets the graticule to accommodate the vertical position and scale of the signal Use mV if you are working with video s...

Page 210: ...of the waveform with the highest sample density appear in warmer colors red shades while the areas of lowest sample density appear in cooler colors blue shades Click Gray Monochrome Gray to define wa...

Page 211: ...setup program in the Windows Start menu Either select All Programs Accessories Remote Desktop Connection or select Programs Accessories Communications Remote Desktop Connection Display setups DPO70000...

Page 212: ...Display setups 180 DPO70000SX MSO DPO70000DX MSO DPO70000C DPO7000C and MSO DPO5000B Series...

Page 213: ...e horizontal scale Normally the horizontal axis displays time information However in some cases the horizontal axis can represent frequency rather than time NOTE This document is for more than one ins...

Page 214: ...ton to set the record length limit to its maximum value Use the Min button to set the record length limit to its minimum value To increase the update rate lower the record length limit To enable highe...

Page 215: ...rms The values track the sample rate and resolution of the analog system To use Use the Sample Rate and Scale entry boxes to set the Digital Sample Rate and the Digital Resolution The following screen...

Page 216: ...this control window to turn on horizontal delay so the acquisition window can be focused on events that occur at a fixed delay after the trigger point NOTE The dedicated Horizontal controls on the fro...

Page 217: ...to toggle the horizontal delay on or off Behavior You can use this control window to turn on horizontal delay so the acquisition window can be focused on events that occur at a fixed delay after the...

Page 218: ...annels This means that you cannot acquire in Sample mode on channel 1 while channel 2 acquires in Envelope mode NOTE This document is for more than one instrument Your instrument or instrument setup m...

Page 219: ...on this tab to define the sampling method the acquisition mode turn Fast Acquisitions on or off and activate an external reference On MSO5000B models and DPO5000B models with option MSOE you can turn...

Page 220: ...ction The following table lists the Autoset default settings Control Changed by Autoset to Selected Channel Last channel selected Acquisition Mode Sample Run Stop Run Deskew Channel Probe Unchanged Di...

Page 221: ...isplay the Prompt Before Action window Click Autoset to toggle between Off and On Off immediately performs an Autoset when the Autoset button is pushed On displays a prompt when the Autoset button is...

Page 222: ...the Autoset Undo button to Off What do you want to do next Learn about Autoset Enable enhanced effective number of bits From the Horiz Acq menu select Horizontal Acquisition Setup and then select the...

Page 223: ...tions are not available on MSO DPO5000B Series instruments Behavior Fast Acquisitions mode keeps the dead time between acquisitions very short This improves the chances of acquiring infrequent events...

Page 224: ...d then stops In Average or Envelope acquisition modes the instrument makes the specified number of acquisitions to complete the averaging or enveloping task In Waveform Database mode the instrument ac...

Page 225: ...When using Roll mode you do not have to wait for slow acquisitions to fill the display When Roll mode is enabled the instrument displays newly acquired data points at the right edge of the display wh...

Page 226: ...all the points in a waveform in one trigger event NOTE In real time sampling the instrument digitizes all of the points it acquires after one trigger event Always use real time or interpolated samplin...

Page 227: ...DPO7254C and DPO7104C with Option 2SR Channels on 1 2 3 or 4 Time base 100 ps pt RT RT RT 50 ps pt RT RT ET 25 ps pt RT ET ET 25 ps pt ET ET ET Channels on 1 2 3 or 4 Time base 200 ps pt RT RT RT 100...

Page 228: ...tion Sample Rate and Resolution controls are readouts only Their settings are impacted by other controls Duration specifies the time across the 10 division screen Sample Rate specifies the number of s...

Page 229: ...t shows the extremes in variations over several acquisitions The instrument saves the highest and lowest values in two adjacent intervals similar to the Peak Detect mode Unlike Peak Detect mode the pe...

Page 230: ...defining the trigger conditions you can use FastFrame to capture only multiple events of the waveform that you are interested in Each captured event is stored sequentially in its own frame The frames...

Page 231: ...Dead time Dead time is the time that the instrument requires to process an acquisition The instrument cannot acquire any new data until the previous data has been processed If dead time is long the in...

Page 232: ...e the controls to change the position and factor or scale in the Zoom area They can be adjusted with the multipurpose knobs To reset the Position and Factor controls click Reset Click the Graticule Si...

Page 233: ...tal Position and Factor will affect all zoom areas that are locked together such that a change to one zoom area will change the horizontal position and factor of all the locked zoom areas On the Verti...

Page 234: ...Horizontal and acquisition setups 202 DPO70000SX MSO DPO70000DX MSO DPO70000C DPO7000C and MSO DPO5000B Series...

Page 235: ...ctor and may be used to test a device under test DUT To use Select Analyze Serial Error Detector to display the Error Detector control window Select a setup from the Preset Setting drop down list and...

Page 236: ...al There will be massive errors when the signal is disconnected but after the signal is reconnected the Error Detector will resynchronize to the signal clear the error counts and rates and resume test...

Page 237: ...or What do you want to do next Use the Error Detector Set error detector advanced settings Specify the error detector test pattern Set the error detector stop condition Data Rates Supported by the Err...

Page 238: ...n and specify the primitives using the keypad or keyboard The Align Character informs the SerDes to align to 10 bit character boundaries when this character is found in the signal Alignment Primitives...

Page 239: ...30 or 40 bit unique pattern in the signal that the error detector uses to align the signal to the acquired comparison pattern The Advanced check box just allows you to set the Disparity of each of the...

Page 240: ...t setup the email address and the SMTP server for Send E mail to work What do you want to do next Use the Error Detector Set up the serial error detector Set error detector advanced settings Specify t...

Page 241: ...press the Sync and Run buttons because the Learn operation was completed in the previous step If you do another Learn operation you risk learning errors created by the DUT To avoid having to rearrang...

Page 242: ...r Set up the serial error detector Set error detector advanced settings Specify the error detector test pattern Set the error detector stop condition Error detector setups 210 DPO70000SX MSO DPO70000D...

Page 243: ...gnaling activity is captured along with Control Channel protocol information to provide a complete view of the link training process Each left right side of the lane transmits Coefficient Update Reque...

Page 244: ...onfiguration The signals are accessed using breakout fixtures 80 20 power splitters and P7633 differential probes Lane1 uses oscilloscope Ch1 and Ch3 while Lane2 uses oscilloscope Ch2 and Ch4 to maxim...

Page 245: ...Analyze Link Training 2 Press the Arm button 3 Start Link Training on the device under test DUT Starting Link Training on the DUT is DUT specific but is typically done from a command line console prog...

Page 246: ...will occur closer in time than the oscilloscope FastFrame rearm time For example a Control Channel event triggers storage of a Frame analog waveform data but a closely following Control Channel event...

Page 247: ...Table may be exported to a comma separated value CSV file or queried over the instrument programming interface PI for further post processing or formatting Using Link Training from the programming in...

Page 248: ...00G_KR4 LINKTRaining LANE1 OFF ON Default ON LINKTRaining LANE2 OFF ON Default OFF LINKTRaining LANE3 OFF ON Default OFF LINKTRaining LANE4 OFF ON Default OFF LINKTRaining MARK FRAme CONtrol TRAining...

Page 249: ...ing Mask testing automatically uses Communication triggers You must install the Serial Communications Mask Testing Option on some instruments before you can access the Comm triggers DPO70000SX instrum...

Page 250: ...select the mask standard 4 If the Mask Polarity control is displayed click it to select positive or negative Behavior The type of mask you choose will determine which standards are available from the...

Page 251: ...6 Mb s DS4NA Max Output 139 26 Mb s STS 1 Pulse 51 84 Mb s STS 1 Eye 51 84 Mb s STS 3 155 52 Mb s STS 3 Max Output 155 52 Mb s None 100Base TX STP 125 Mb s 100Base TX UTP 125 Mb s Gigabit Ethernet 1 2...

Page 252: ...6 Gb s S1600b Optical 1 966 Gb s None Drv 500 Mb s Drv 750 Mb s Drv 1 0 Gb s Drv 1 5 Gb s Drv 2 0 Gb s Ext Drv 500 Mb s Ext Drv 750 Mb s Ext Drv 1 0 Gb s Ext Drv 1 5 Gb s Ext Drv 2 0 Gb s Rcv 500 Mb s...

Page 253: ...this control window to set whether changes in the scale will also be made in the mask You can also use this control window to return to the Mask Setup control window To use Click the Lock Mask to Wave...

Page 254: ...n on the controls click the buttons Vertical Horizontal Trigger Mode Autofit To use Click Autoset to align the signal within the mask This operates the same as the Autoset button on the front panel Cl...

Page 255: ...ill search for the best fit to minimize hits For example if you set the Vertical Maximum to 10 per division Autofit searches up to and including 10 percent of the vertical divisions for the best fit A...

Page 256: ...fter each acquisition when autofit is used NOTE The interaction between Autofit and Persistence can affect the display of persistence data Autofit moves the waveform vertically and or horizontally to...

Page 257: ...down arrows increase or decrease the vertex by 1 To set a specific number click in the Vertex entry box to activate the pop up calculator Clicking the Add button adds a point to the boundary in the se...

Page 258: ...ick the number 1 4 of the channel math expression or reference you want to use Behavior When you click the Ch tab and select a channel number it is used as the source for mask testing You can select o...

Page 259: ...s to increase or decrease the percentage in 1 increments A tolerance that is 0 makes passing a mask test more difficult A mask test that passes with a tolerance that is 0 demonstrates that the signal...

Page 260: ...mask testing Go to the Mask Setup control window Set up mask test fail notification From the Mask menu select Pass Fail Setup Overview Use the Test Fail Notification controls to set the mode of notif...

Page 261: ...printer when a mask test fails SRQ When SRQ is turned on the instrument sends a GPIB SRQ command when a mask test fails AUX Out When AUX Out is turned on the instrument sends a trigger to the Auxilia...

Page 262: ...rom the Mask menu select Pass Fail Setup Overview Use the Number of Waveforms Samples Fail Threshold and Pre Test Delay controls to set the parameters for the mask test NOTE If you are using Waveform...

Page 263: ...rument must be turned off With the sound card off you will not have access to standard Windows audio features such as the ability to listen to a music CD If you turn the sound card on the internal spe...

Page 264: ...selected the mask test is performed on the negative pulse When the Both control is selected the mask test is performed on both the positive and negative pulses NOTE The mask polarity controls operate...

Page 265: ...rameters from the Mask Test control window and the Pass Fail Results window is displayed What do you want to do next Learn about waveform parameters Learn about test notification Learn about mask pola...

Page 266: ...egment Pass Fail Test Controls What do you want to do next Learn about test summary Learn about hits per segment Learn about the Mask Test Controls window Go to the Test Setup control window View the...

Page 267: ...e Mask menu select Pass Fail Results Overview Use this window to view the hits for each mask segment of the test To use The Hits per segment is a read only field Behavior Each mask test can have up to...

Page 268: ...ss Fail Test Setup control window When the test is reset the Total Hits Failed Wfms Hits per segment and Waveform Tested fields are reset to zero What do you want to do next Learn about test summary L...

Page 269: ...dit User Mask Display Autoset Autofit To use Click the Edit User Mask button to access the Mask Edit control window Click the Copy Current Mask button To User Mask to make the current mask a user mask...

Page 270: ...ior The Segment up and down arrows increase or decrease the segment number by one The Vertex up and down arrows increase or decrease the vertex by 10 To set a specific number click the Vertex entry bo...

Page 271: ...rop down list box navigate to the location of the user mask Click on the name of the mask file you want to recall You can also click the keyboard icon to access the pop up keyboard and enter the file...

Page 272: ...ekScope Masks unless you choose another directory What do you want to do next Learn about saving masks Learn more about mask types Learn more about mask setups Limit test Set up limit testing From the...

Page 273: ...ion tab and set up the mode you prefer for failure notification b Click the Setup tab to return to the Limit Test setup window 3 Select Test Options a Click the Lock Template to Waveform button On to...

Page 274: ...the Limit Test Setup control window The default file type is a waveform WFM The default folder is C Users Username Tektronix TekScope waveforms NOTE Any characters that you enter in the file name box...

Page 275: ...tup button and complete the E mail on Event Setup setup information Save WFM When Save WFM is On failed waveform data is saved to a file Click the Path button and enter a location to save the file Log...

Page 276: ...Mask testing setups 244 DPO70000SX MSO DPO70000DX MSO DPO70000C DPO7000C and MSO DPO5000B Series...

Page 277: ...Predefined Functions Editor Custom Analysis Functions MATLAB Analysis Basic Spectral Advanced Spectral You can define up to four math waveforms The controls for each waveform are similar but independe...

Page 278: ...d be averaged Behavior The selected math waveform is averaged over the number of acquisitions shown in the control Each math waveform has its own set of controls The following buttons allow you to cha...

Page 279: ...ditor button to access the Math Equation Editor Behavior You can set or change the variable values in this control window To add a variable to a math expression go to the math editor and click the VAR...

Page 280: ...talled you can select and use them just like the instrument built in functions Click the question mark to display information on custom analysis function plugins Custom analysis functions What do you...

Page 281: ...bout writing Math Plugins Learn about writing MATLAB custom function plugins Basic spectral setups Set up basic spectral analysis From the Math menu select Basic Spectral or click Basic in the Math Se...

Page 282: ...ndow 5 Use the front panel Resolution knob to directly control Frequency Span 6 Use the Horizontal Scale knob to directly control Resolution Bandwidth Advanced spectral setups Set up advanced spectral...

Page 283: ...al FFT waveforms using the time domain controls gating controls or frequency domain controls Some of these controls are interactive making changes with one set of controls will sometimes change settin...

Page 284: ...try box and then use the multipurpose knobs to set the Center Freq Freq Span and Res BW Click Full to set the frequency span to full Math setups 252 DPO70000SX MSO DPO70000DX MSO DPO70000C DPO7000C an...

Page 285: ...decrease the center frequency beyond the present limits reduce the frequency span Frequency span The frequency span is the range of frequencies over which the spectral analysis is performed The maxim...

Page 286: ...nt In most cases the phase reference point is located at the center of the gated data However if you use a Tek Exponential window the phase reference point is at the 20 point of the gated data Gate Du...

Page 287: ...indow has less spectral leakage than the Hanning Hamming or Rectangular windows The Gaussian window provides optimal localization in both the time and frequency domain The Flattop2 window is good for...

Page 288: ...y The FFT waveform displays the input signal frequency components from DC 0 Hz to the Nyquist frequency Continue Previous The FFT display The following illustration shows a typical time domain wavefor...

Page 289: ...nts to be at different amplitudes The transitions between the start and end points cause discontinuities in the waveform that introduce high frequency transients These transients add false frequency i...

Page 290: ...nantly single frequency signals to look for higher order harmonics You can also determine the best window empirically by first selecting the Rectangular window and then selecting in the following orde...

Page 291: ...uist frequency as you increase the time per division the aliased frequency components should appear at their proper frequency If the increased number of frequency components shown on the screen makes...

Page 292: ...window to define or edit math waveforms The readout shows the math expression as you construct it New functions are added at the beginning of the expression each time you click one of the function bu...

Page 293: ...that is longer than the physical space in the control window readout you can scroll through the expression using the Text Edit controls To edit another math waveform select the waveform from the Math...

Page 294: ...sert a Time function into the math expression Primitive buttons Average Sqrt FABS Sin Cos Tan Integral Log 10 Ceil ASin ACos ATan Derivative Log e Floor Sinh Cosh Tanh Invert Exp Min Max WfmDB 2 Click...

Page 295: ...pression Arc cosine Inserts the text ACOS into the math expression Arc tangent Inserts the text ATAN into the math expression Average Inserts the text AVG into the math expression Enter an argument to...

Page 296: ...t to the function The natural antilog function raises e to the power of the argument Natural logarithm 1 Inserts the text LN into the math expression 2 Enter an argument to the function The LN functio...

Page 297: ...ing the Measurement list with math waveforms Magnitude Inserts the text SpectralMag into the math expression Select one of the waveforms as an argument to the function This function creates an FFT mag...

Page 298: ...tton enters the variable into the math expression Clicking the More button accesses the other four variables The Controls button activates the Math Variables control window where you can quickly set o...

Page 299: ...ion of available Math Arbitrary Filters in C TekScope Math Arbitrary Filters or browse to the location of any custom filters 3 Select the filter you want to use then click Open 4 Click the correspondi...

Page 300: ...Math setups 268 DPO70000SX MSO DPO70000DX MSO DPO70000C DPO7000C and MSO DPO5000B Series...

Page 301: ...th the measurements Measurement Snapshot executes all of the single waveform measurements once the measurements are not continuously updated Any measurements previously taken are retained on the displ...

Page 302: ...he Measurement control window overview Measurement setup control window Amplitude tab From the Measure menu select Measurement Setup then open the Ampl tab Overview Use this control window to set up t...

Page 303: ...isitions mode for example timing measurements Automatic measurements are not available on digital channels What do you want to do next Go to a step by step procedure to set up the instrument for takin...

Page 304: ...is seen during measurements is usually seen when using ET mode when the instrument is still filling the waveform record Warnings appear in the measurement readouts because of ET mode but eventually go...

Page 305: ...nt Setup Control Window to turn the measurement display on and off Behavior Measurements are updated approximately three times a second or less for long record lengths Measurements are calculated afte...

Page 306: ...ed region Low The low level measurement is the value used as 0 whenever high reference mid reference or low reference values are needed such as in fall time or rise time measurements This value can be...

Page 307: ...gated region Measurement setup control window Time tab From the Measure menu select Measurement Setup then open the Time tab Overview Use this control window to set up the instrument for taking autom...

Page 308: ...to enhance your measurements In Fast Acquisitions mode measurements are calculated on an array of acquired waveforms As the waveforms accumulate with persistence on your measurement accuracy may incre...

Page 309: ...control window The delay measurement is added to the list TIP Select the individual measurements directly from the Measure menu Time submenu Rise Time This timing measurement is the time required for...

Page 310: ...required to complete the first cycle in a waveform or gated region Period is the reciprocal of frequency and is measured in seconds Frequency This timing measurement is a measure of the first cycle in...

Page 311: ...lable on digital channels TIP Turn on Fast Acquisitions to enhance your measurements In Fast Acquisitions mode measurements are calculated on an array of acquired waveforms As the waveforms accumulate...

Page 312: ...rn the measurement display on and off Behavior Measurements are updated approximately three times a second or less for long record lengths Measurements are calculated after each acquisition You can di...

Page 313: ...n a reference waveform saved in FastAcq mode ExtdB 10 Log pTopMean pBaseMean Eye Height Eye height is the measurement of the eye height in volts Eye Height pTopMean 3 pTopSigma pBaseMean 3 pBaseSigma...

Page 314: ...ignal amplitude to the noise of the top or base of the signal as specified by you S N Ratio Ptop Pbase PTopsigma or PBasesigma Duty Cycle Distortion Duty cycle distortion is the peak to peak time vari...

Page 315: ...able on digital channels TIP Turn on Fast Acquisitions to enhance your measurements In Fast Acquisitions mode measurements are calculated on an array of acquired waveforms As the waveforms accumulate...

Page 316: ...asurements are calculated after each acquisition You can display up to eight automatic measurements A new control window opens when you click Phase Enter the parameters for the phase measurement and t...

Page 317: ...the amount of time that one waveform leads or lags another waveform The measurement is expressed in degrees where 360 comprise one waveform cycle Counter measurement setup NOTE This online help docum...

Page 318: ...verages Trigger Source Frequency The Trigger Source Frequency Application uses trigger hardware rather than the acquired waveform to measure the frequency of the signal on the trigger source to a very...

Page 319: ...ick in the Weight entry box and then use the multipurpose knobs to set the statistics weight Click Reset to restart the statistics calculations TIP Select the individual statistics directly from the M...

Page 320: ...in the entry boxes associated with the desired reference level and use the multipurpose knobs to adjust the value If you want the reference levels to be measured by volts click the Absolute button un...

Page 321: ...her it is defining the high or low reference level This setting is best for examining eye patterns and optical signals Mid reference The Mid reference defines the middle reference level of a waveform...

Page 322: ...Track Mode These buttons are only displayed when you click Cursor To define the gating area from a Zoom area click one of the four Zoom buttons To turn off measurement gating click the Off button Clic...

Page 323: ...trol window overview Learn more about using vertical bar cursors Learn more about the Measurement controls Learn about setting the Reference Level controls Learn about using Measurement Statistics Set...

Page 324: ...ed waveforms As the waveforms accumulate with persistence on your measurement accuracy may increase However only one or two channels can be displayed when you take automatic measurements using Fast Ac...

Page 325: ...u specify the histogram box in terms of location the Y location control adjusts the bottom edges of the histogram while the X location control adjust the left edge of the histogram box TIP Here are se...

Page 326: ...its horizontal 1 Select a mode 2 Specify the adjustment 3 Enter limits Set histogram box limits vertical 1 Select a mode Measurement setups 294 DPO70000SX MSO DPO70000DX MSO DPO70000C DPO7000C and MSO...

Page 327: ...fy the adjustment 3 Enter limits Set histogram box location horizontal 1 Select a mode 2 Specify the adjustment Measurement setups DPO70000SX MSO DPO70000DX MSO DPO70000C DPO7000C and MSO DPO5000B Ser...

Page 328: ...Enter limits Set histogram box location vertical 1 Select a mode 2 Specify the adjustment 3 Enter limits Measurement setups 296 DPO70000SX MSO DPO70000DX MSO DPO70000C DPO7000C and MSO DPO5000B Serie...

Page 329: ...ther green or white triangles on screen Green triangles represent locations placed automatically by the instrument white triangles represent locations placed manually by the user Behavior Use the Mark...

Page 330: ...t Your instrument or instrument setup may not include some of these features For DPO70000SX instruments see Features not available To use For information on the controls click the buttons Type Source...

Page 331: ...struments see Features not available To use Click a button to identify the type of search to perform The search Type will appear under Type in the Type Source table Behavior Clicking a button places t...

Page 332: ...ck OK Behavior Use the single Source Selection window to set the Channel parameters for an Edge Glitch Runt Transition Width Window or Timeout search Use the dual Source Selection window to set the Cl...

Page 333: ...nstrument to automatically search and place marks NOTE Search and mark operations are not available on digital channels To use For information on the controls click the buttons Type Source Copy Settin...

Page 334: ...y Settings button Overview Use these controls to copy settings from one search to another You can also copy a trigger setting to a search To use 1 Set up the desired search s in the Search Select wind...

Page 335: ...evel When auto is selected the midlevel of the chip select source is automatically calculated and values above or below this voltage are used to gate mark generation In custom mode the specified level...

Page 336: ...trol Window Pattern Search for a logic pattern AND OR NAND or NOR across multiple waveforms with each input set to either High Low or Don t Care Search for when the event goes true goes false or stays...

Page 337: ...on source waveforms Bus Trigger Control Window for Parallel Bus Trigger Control Window for I2C Serial Bus Trigger Control Window for SPI Serial Bus Trigger Control Window for RS232 Serial Bus Trigger...

Page 338: ...search counts results From the Analyze menu select Search then open the Results tab and click the Count button Overview Use the controls in this window to view the number of times the search criteria...

Page 339: ...From the Analyze menu select Search then open the View tab Overview Use the controls in this window to display marks To use Click the Show Marks button On or Off to display the marks placed on a wavef...

Page 340: ...any of the tabs Overview Use these controls to find points of interest and place or edit marks on the waveform The Search and Mark controls are also available through front panel buttons DPO7AFP requi...

Page 341: ...rk Cursor 1 Cursor 2 Click one of these buttons to position a cursor at the active mark Zoom 2 Zoom 3 Click one of these buttons to zoom the area around the active mark Zoom 1 is the default used by t...

Page 342: ...where there is a search mark Clear Click to remove the currently selected user mark If a user mark is not selected you are asked to first select a user mark Clear All Click this button to remove all...

Page 343: ...ab Overview Use these controls to export the Mark Table to a file To use Export Click this button to open the Save Marks Table dialog box and export the table to a file in CSV format Analyzing wavefor...

Page 344: ...settings as the main trigger The results of a visual search are displayed as marks with a color distinct from those of an Analyze search To use Do the following steps to set up a visual search 1 Set u...

Page 345: ...Active search areas only appear in the zoom window Use the other Analyze Search control windows to control or view the results of the Visual Search What do you want to do next Learn about bus setups L...

Page 346: ...dual search controls similar to the one shown above These controls vary depending on the search Type Setting the search conditions is similar to setting the parameters for an Event Trigger NOTE No inf...

Page 347: ...u can also set the component threshold levels through the Logic Thresholds Setup button Click the Search For selections Error Pattern Character Symbol or Any Character to view their parameters What do...

Page 348: ...ol to search for a symbol Click Setup to adjust the logic thresholds Search for any character Set up RS 232 search parameters After setting up an RS 232 bus go to the Analyze menu select Search and th...

Page 349: ...search conditions is similar to setting the parameters for an Event Trigger NOTE No information will be displayed in this window until you specify at least one search Type in the Search Select Contro...

Page 350: ...the Edit button Optionally you can also set the component threshold levels through the Logic Thresholds Setup button Click the Search For selections Start Parity Error or Data to view their parameter...

Page 351: ...Set up SPI search parameters After setting up an SPI bus go to the Analyze menu select Search and then select the Bus trigger type then open the Configure tab Overview Use the controls to set up the...

Page 352: ...Bus Search Selection List Select the type of bus cycle or activity to search for from the Search For drop down list For some Search For selections you must set additional fields to define other parame...

Page 353: ...Set up I2C search parameters After setting up an I2C bus go to the Analyze menu select Search and then select the Bus trigger type then open the Configure tab Overview Use the controls to set up the...

Page 354: ...he type of bus cycle or activity to search for from the Search For drop down list For some Search For selections you must set additional fields to define other parameters such as for a Data In this ex...

Page 355: ...ng ack Search for address The address mode options available in Search For Address are Address Format Address Value Addressing Mode Special Address and Direction Analyzing waveforms DPO70000SX MSO DPO...

Page 356: ...Edit to enter the data value Click Setup to adjust the logic thresholds Search for address and data The address and data mode options available in Search For Address are Address Format Address Value...

Page 357: ...l search controls similar to the one shown above These controls vary depending on the search Type Setting the search conditions is similar to setting the parameters for an Event Trigger NOTE No inform...

Page 358: ...t threshold levels through the Logic Thresholds Setup button Click the Search For selections Sync Reset Suspend Resume End of Packet Token Address Packet Data Packet Handshake Packet Special Packet or...

Page 359: ...Search for reset Search for suspend Search for resume Analyzing waveforms DPO70000SX MSO DPO70000DX MSO DPO70000C DPO7000C and MSO DPO5000B Series 327...

Page 360: ...er the address and endpoint values Click Setup to adjust the logic thresholds Search for data packet The data packet options available in Search For Data Packet are Data Packet Type When data packet i...

Page 361: ...he special packet option available in Search For Special Packet is Special Type Select the Special Type from the drop down list Click Setup to adjust the logic thresholds Search for error The Error op...

Page 362: ...symbol The options available in Search For Character Symbol are Character Symbol RD Symbol RD and Disparity Select the Character Symbol RD Symbol RD and Disparity Search for long ordered set The opti...

Page 363: ...t options available in Search For Packets are types of Packets Select the packet types Search for LFPS Search for any control character Analyzing waveforms DPO70000SX MSO DPO70000DX MSO DPO70000C DPO7...

Page 364: ...individual search controls similar to the one shown above These controls vary depending on the search Type Setting the search conditions is similar to setting the parameters for an Event Trigger NOTE...

Page 365: ...t click the Edit button Optionally you can also set the component threshold levels through the Logic Thresholds Setup button Click the Search For selections Stop Start of Transmission End of Transmiss...

Page 366: ...of transmission Search for bus turn around Search for escape mode The escape mode command options available in Search For Escape Mode are as follows Analyzing waveforms 334 DPO70000SX MSO DPO70000DX...

Page 367: ...ows Search for error The error type options available in Search For Error Type are as follows Search for short packet The data type options available in Search On Short Packet are as follows Analyzing...

Page 368: ...Channel Direction Packet Data Format and the Packet Data Value Search for long packet The data type value options available in Search For Long Packet are as follows Analyzing waveforms 336 DPO70000SX...

Page 369: ...et window allows you to set the Virtual Channel Direction Data Type Value Word Count Number of Bytes Payload Format Offset in bytes and Payload Value Analyzing waveforms DPO70000SX MSO DPO70000DX MSO...

Page 370: ...al search controls similar to the one shown above These controls vary depending on the search Type Setting the search conditions is similar to setting the parameters for an Event Trigger NOTE No infor...

Page 371: ...lick the Search For selections Long Ordered Set Any Control Character or Error to view their parameters What do you want to do next Bus setup control window Config tab on page 107Bus setup control win...

Page 372: ...Search and then select the Bus trigger type then open the Configure tab Overview Use the controls to set up the CAN Bus parameters To use For information on the controls click the buttons Type Source...

Page 373: ...Select the type of bus cycle or activity to search for from the Search For drop down list For some Search For selections you must set additional fields to define other parameters such as for Ident Dat...

Page 374: ...earch for data The data type options available in Search For Data are Number of Bytes Data and Data is Search for end of frame Analyzing waveforms 342 DPO70000SX MSO DPO70000DX MSO DPO70000C DPO7000C...

Page 375: ...and then select the Bus trigger type then open the Configure tab Overview Use the controls to set up the LIN Bus parameters To use For information on the controls click the buttons Type Source Copy S...

Page 376: ...ow Bus Search Selection List Select the type of bus cycle or activity to search for from the Search For drop down list For some Search For selections you must set additional fields to define other par...

Page 377: ...earch for ident data The data type options available in Search For Ident Data are Number of Bytes Identifier Data is and Data value Select the Data from the drop down list Click Edit to enter the data...

Page 378: ...d then select the Bus trigger type then open the Configure tab Overview Use the controls to set up the FLEXRAY Bus parameters To use For information on the controls click the buttons Type Source Copy...

Page 379: ...ection List Select the type of bus cycle or activity to search for from the Search For drop down list For some Search For selections you must set additional fields to define other parameters such as f...

Page 380: ...cycle count The data type options available in Search For Cycle Count are Cycle Count and Qualifier Select the Quallifier from the drop down list Click Edit to enter the Cycle Count value Click Logic...

Page 381: ...e in Search For Ident Data are Identifier Number of Bytes Qualifier Offset and Data Select the Identifier from the drop down list Click Edit to enter the data values Click Logic Thresholds to adjust t...

Page 382: ...t the Bus trigger type then open the Configure tab Overview Use the controls to set up the MIL 1553 Bus parameters To use For information on the controls click the buttons Type Source Copy Settings Se...

Page 383: ...earch For selections you must set additional fields to define other parameters such as for Command Word In this example you need to set an Search when RT Address is Parity T R bit RT Address Subaddres...

Page 384: ...en Remote Terminal Address is and RT Address click the Edit button Search for data The data type options available in Search For Data are Data and Parity Search for RT IMG Time The data type options a...

Page 385: ...arameters After setting up an Ethernet bus go to the Analyze menu select Search and then select the Bus trigger type then open the Configure tab Overview Use the controls to set up the Ethernet Bus pa...

Page 386: ...search conditions is similar to setting the parameters for an Event Trigger NOTE No information will be displayed in this window until you specify at least one search Type in the Search Select Contro...

Page 387: ...esses Q Tag Control Information MAC Length Type IPv4 Header TCP Header MAC Data TCP IPv4 Client Data End of Packet Idle or FCS CRC Error to view their parameters What do you want to do next Learn abou...

Page 388: ...length type The data type option available in Search For MAC Length Type is MAC Length Type IPv4 header The data type options available in Search For IPv4 Header are IP Protocol Source Address and De...

Page 389: ...a The data type options available in Search For MAC Data are Data Number of Bytes Offset and Qualifier TCP IPv4 client data The data type options available in Search For TCP IPv4 Client Data are Data...

Page 390: ...e Ethernet End of Packet Idle Select Idle to search for an Ethernet Idle character FCS CRC error Select FCS CRC Error to search for an Ethernet FCS error Analyzing waveforms 358 DPO70000SX MSO DPO7000...

Page 391: ...e Configure tab Overview Use the controls to set up DDR DDR2 DDR3 DDR4 DDR3L LPDDR LPDDR2 LPDDR3 GDDR3 GDDR5 bus parameters To Use For information on the controls click the buttons Type Source Copy Se...

Page 392: ...vels from the Input Levels drop down list If you selected the Custom input level enter the High Mid Nd Low levels using the keypad or attached keyboard Select the source of the Data DQ from the Data D...

Page 393: ...ry depending on the search Type Setting the search conditions is similar to setting the parameters for an Event Trigger NOTE No information will be displayed in this window until you specify at least...

Page 394: ...Copy Settings Search Parameters Search Mark Controls Behavior Clicking an entry in the Type column activates the individual search controls similar to the one shown above These controls vary depending...

Page 395: ...earn about Custom Serial Bus Setup View the results table From the Bus Setup window select the Display tab and click Results Table or in the Analyze menu select Results Table Overview Use this dialog...

Page 396: ...g icon is displayed if the packet contains an error red X or warning yellow triangle If the packet row or bus column contains collapsed detail the Expand icon is displayed Click on the icon to expand...

Page 397: ...ults table To resize the docked table click the splitter above the docked window and drag it to the size you want the table Analyzing waveforms DPO70000SX MSO DPO70000DX MSO DPO70000C DPO7000C and MSO...

Page 398: ...window you can move columns select the columns you want displayed or restore the defaults You can also drag columns to new locations and resize them using a mouse Right clicking on a column heading i...

Page 399: ...Results Table Click the Copy button to open the Select the columns to Copy Export dialog box From the dialog box click the Copy button To use To save to a file enter a Name or use the default file nam...

Page 400: ...programs What do you want to do next Learn about selecting the columns to Copy Export Select columns to copy or export From the Analyze menu select Results Table Click the Settings button to open the...

Page 401: ...Results Table Click the Export button to open the Select the columns to Copy Export dialog box From the dialog box click the Export button Overview Use this dialog box to export save the contents of t...

Page 402: ...fault folder is C Users Username Tektronix TekScope data NOTE Any characters that you enter in the file name box become part of the file name The instrument does not use any user defined file extensio...

Page 403: ...Overview MyScope allows you to create custom control windows that include only the controls you use regularly Instead of switching between several control windows put the controls you use into a cust...

Page 404: ...dentical control in the user interface Behavior Clicking on a control in the list allows you to preview the control in the pane at the right Selecting or clearing the check boxes for the controls in t...

Page 405: ...nd assigns the same measurements to the next channel as were assigned to the previous channel This is a quick way to assign the same measurements to multiple channels Each page of the measurement cont...

Page 406: ...directory select a location in the Save in drop down list Click Save Behavior This dialog box opens in the default directory C Users Username Tektronix TekScope MyScope If you want to save MyScope co...

Page 407: ...TekScope MyScope If you saved MyScope control window files in another directory click the Look in drop down list to go to that directory Clicking Open will open the selected MyScope control window al...

Page 408: ...MyScope setups 376 DPO70000SX MSO DPO70000DX MSO DPO70000C DPO7000C and MSO DPO5000B Series...

Page 409: ...ria Depending on the trigger type the instrument will update the graphic as you define the trigger parameters The trigger expression near the top right side of the Trigger window provides a one line s...

Page 410: ...of posttrigger data with the Horizontal Position control Select the trigger type From the Trig menu select A Event Trigger Setup or B Event Trigger Setup To use In the Trigger Type drop down list box...

Page 411: ...ts Window Use the Window trigger to trigger the instrument when the input signal rises above an upper threshold or falls below a lower threshold level NOTE Window trigger on B Events is not available...

Page 412: ...r to set the trigger level to 50 To use Click Normal or Auto to set the Trigger Mode Click Enhanced Triggering to more accurately place the trigger on the display Click Optimized for Noisy Signals to...

Page 413: ...It is also useful when looking for activities on signal nodes such as oscillations when you first set up the instrument The length of time it waits for a trigger event depends on the time base setting...

Page 414: ...the display of complex waveforms such as digital pulse trains that may have several possible trigger points The Level control can be either the main trigger level when there is no B Trigger event or...

Page 415: ...s are unsupported and the front panel control is inoperative This control also works differently for certain Comm triggers A event triggers Set up a parallel bus trigger From the Trigger menu select t...

Page 416: ...he Pattern Click the Format box and select a Binary ASCII Decimal Hex or Symbolic display format What do you want to do next Learn how to define a pattern for a bus Learn how to set the voltage thresh...

Page 417: ...some Trigger On selections you must set additional fields to define other parameters such as for a Pattern In this example you need to set the Character or Symbol values and select the Disparity You c...

Page 418: ...ption SR EMBD From the Trigger menu select the Bus setup Overview Use the controls to set up the I2C Bus parameters To use For information on the controls click the buttons Trigger Type Mark All Trigg...

Page 419: ...ameters such as for an Address In the following example you need to set an Address Format Address Value click the Edit button Data Direction and Addressing Mode Optionally you can also set the compone...

Page 420: ...10 XXX X CBUS 0000 001 X Not recommended for use with iCapture inputs What do you want to do next Learn about bus setups Set up an SPI bus trigger NOTE The Bus Trigger Control Window for SPI Serial op...

Page 421: ...levels through the Logic Thresholds Setup button For some Trigger On selections you need to set additional fields to define other parameters such as for a Data In the following example you need to set...

Page 422: ...use For information on the controls click the buttons Trigger Type Mark All Trigger Events Bus Logic Thresholds Trigger setups 390 DPO70000SX MSO DPO70000DX MSO DPO70000C DPO7000C and MSO DPO5000B Ser...

Page 423: ...On selections you must set additional fields to define other parameters such as for a Data In this example you need to set the Data Format click the drop down list Number of Bytes and Data Value click...

Page 424: ...use For information on the controls click the buttons Trigger Type Mark All Trigger Events Bus Logic Thresholds Trigger setups 392 DPO70000SX MSO DPO70000DX MSO DPO70000C DPO7000C and MSO DPO5000B Ser...

Page 425: ...ections you must set additional fields to define other parameters such as for a Data Packet In this example you need to set a Data Packet Type Trigger when data packet is Data Format Number of Bytes B...

Page 426: ...et the Trigger On condition to Special Packet Here you have the option to trigger on ANY specifically on ERR high only SPLIT high only PING high only or Reserved special types The Trigger on SPLIT tok...

Page 427: ...he options available in Start Complete are as follows Start and end bits The options available in Start and End bits are as follows Endpoint type The options available in Endpoint Type are as follows...

Page 428: ...option SR PCIE From the Trigger menu select Bus setup Overview Use the controls to set up the PCIe Bus parameters To use For information on the controls click the buttons Trigger Type Mark All Trigge...

Page 429: ...some Trigger On selections you must set additional fields to define other parameters such as for a Pattern In this example you need to set the Character or Symbol values and select the Disparity You c...

Page 430: ...option SR AUTO From the Trigger menu select Bus setup Overview Use the controls to set up the CAN Bus parameters To use For information on the controls click the buttons Trigger Type Mark All Trigger...

Page 431: ...is example you need to set the Trigger When Identifier When Direction values and select the And when data is values You can also set the component threshold levels through the Logic Thresholds Setup b...

Page 432: ...t byte with don t care values X To trigger on a two byte message you must enter don t care values explicitly up to the desired size For example a binary entered value of 01 is equivalent to XXXX XX01...

Page 433: ...use For information on the controls click the buttons Trigger Type Mark All Trigger Events Bus Logic Thresholds Trigger setups DPO70000SX MSO DPO70000DX MSO DPO70000C DPO7000C and MSO DPO5000B Series...

Page 434: ...to define other parameters such as for a Ident Data In this example you need to set the Trigger when Identifier and the And When Data is values You can also set the component threshold levels through...

Page 435: ...use For information on the controls click the buttons Trigger Type Mark All Trigger Events Bus Logic Thresholds Trigger setups DPO70000SX MSO DPO70000DX MSO DPO70000C DPO7000C and MSO DPO5000B Series...

Page 436: ...rigger On selections you must set additional fields to define other parameters such as for a Ident Data In this example you need to set the Identifier Data Number of Bytes Offset and Qualifier values...

Page 437: ...th option SR AERO From the Trigger menu select Bus setup Overview Use the controls to set up the MIL STD 1553 Bus parameters To use For information on the controls click the buttons Trigger Type Mark...

Page 438: ...s to define other parameters such as for a Ident Data In this example you need to set the Trigger When Address is RT Address type and values the T R bit Subaddress Mode type and value and select the W...

Page 439: ...or are as follows Set up an ethernet bus trigger NOTE The Bus Trigger Control Window for Ethernet Serial option is available only on DPO MSO5000B Series instruments with option SR ENET From the Trigge...

Page 440: ...use For information on the controls click the buttons Trigger Type Mark All Trigger Events Bus Logic Thresholds Trigger setups 408 DPO70000SX MSO DPO70000DX MSO DPO70000C DPO7000C and MSO DPO5000B Ser...

Page 441: ...his example you need to set the Source Port Destination Port Seq Number and set the Acq Number Click the Trigger On selections Start Frame Delimiter MAC Addresses Q Tag Control Information MAC Length...

Page 442: ...gger on MAC Length Type For Trigger on MAC Length Type you need to specify the MAC length and type Trigger on IPv4 Header For Trigger on IPv4 Header you need to enter the IP Protocol Source Address an...

Page 443: ...MAC Data For Trigger On MAC Data you need to specify the Data Number of Bytes Offset and the Qualifier Trigger on End of Packet Select End of Packet to trigger on an Ethernet End of Packet Trigger on...

Page 444: ...automatically uses Communication triggers You must install the Serial Communications Mask Testing Option on some instruments before you can access the Comm triggers Overview Use the controls in this...

Page 445: ...tandard drop down list depend on the coding format that you selected The coding format also determines whether upper lower single or clock threshold levels or pulse forms are used The standard determi...

Page 446: ...on the controls click the buttons Trigger Type Mark All Trigger Events Settings Source Set to 50 Level Coupling Slope Force Trigger Envelope The following Trigger Edge screen appears on MSO70000C DX...

Page 447: ...es Return to the Trigger Setup control window Trigger slope The trigger slope determines whether the instrument finds the trigger point on the rising edge falling edge or both edges of the signal Set...

Page 448: ...ication option has Occurs and Logic In other instruments the source selection includes a combo list and a Select button The Trigger if Period Qualification option has Occurs Logic and Occurs And The f...

Page 449: ...ols the trigger graphic shows the trigger sequence NOTE You have the option to add user defined labels for analog reference bus and digital sources What do you want to do next Learn more about other t...

Page 450: ...0C DX SX Series instrument except for the source selection and Trigger if Glitch Qualification options In MSO70000C DX Series the source selection includes a combo list and a Select button The Trigger...

Page 451: ...o be qualified by the logical state of other channels or by a bus pattern MSO70000C DX Series only When you click on the trigger window controls the trigger graphic shows the trigger sequence What do...

Page 452: ...ttings Pattern Type Pattern Edit Button Pattern Format Logic Thresholds Trigger if Logic State Behavior Use the Logic Pattern trigger to trigger the instrument when the logic inputs cause the selected...

Page 453: ...s must be satisfied for a specific amount of time before the instrument triggers The instrument input channels represent the data inputs You can set each channel to be logic high logic low or don t ca...

Page 454: ...e instruments the source selection is a drop down list and the Trigger if Runt Qualification option has Occurs and Logic In other instruments the source selection includes a combo list and a Select bu...

Page 455: ...r types Return to the Trigger Setup control window Trigger graphic The trigger graphic displays a visual indicator of the trigger criteria Depending on the trigger type the instrument will update the...

Page 456: ...e or if the clock source is R Clk only one level is available for the data source When you click on the trigger window controls the trigger graphic shows the trigger sequence Trigger graphic The trigg...

Page 457: ...some instruments the source selection is a drop down list and the Trigger if Setup Hold Qualification option has Occurs and Logic In other instruments the source selection includes a combo list and a...

Page 458: ...bus pattern MSO70000C DX Series only The minimum setup and hold time is 2 ns Negative setup and hold times skew the violation region to locate it before or after the clock edge The instrument can dete...

Page 459: ...Your instrument may not have every feature discussed in this online help depending on its configuration Overview Use the controls in this window to set up the SPI trigger parameters To use For informa...

Page 460: ...eas a two byte message would be entered as XXXX XXXX XXXX XX01 When you click on the trigger window controls the trigger graphic shows the trigger sequence Trigger graphic The trigger graphic displays...

Page 461: ...ion to be True or False when the clock input changes state The Pattern Type buttons set the Boolean logic function for the trigger Not recommended for use with iCapture inputs What do you want to do n...

Page 462: ...n set each data channel to be logic high logic low or a don t care value You can set the channel 4 xa0 input to a rising or falling clock edge or channel 2 on 2 channel models Use the multipurpose kno...

Page 463: ...lification options In some instruments the source selection is a drop down list and the Trigger if Timeout Qualification option has Occurs and Logic In other instruments the source selection includes...

Page 464: ...creen appears on MSO DPO5000B Series instruments NOTE The Logic qualification is not available on DPO5000B and MSO5000B Series instruments Trigger setups 432 DPO70000SX MSO DPO70000DX MSO DPO70000C DP...

Page 465: ...ument The time out limit is measured at the voltage level specified by the Level control NOTE You have the option to add user defined labels for analog reference bus and digital sources What do you wa...

Page 466: ...e source selection is a drop down list and the Trigger if Transition Qualification option has Occurs and Logic In other instruments the source selection includes a combo list and a Select button The T...

Page 467: ...control specifies the target transition time What do you want to do next Learn more about other trigger types Return to the Trigger Setup control window Trigger graphic The trigger graphic displays a...

Page 468: ...rmation on the controls click the buttons Trigger Type Settings Source Format Polarity Graticule Trigger On Holdoff Autoset Trigger setups 436 DPO70000SX MSO DPO70000DX MSO DPO70000C DPO7000C and MSO...

Page 469: ...l window Scan controls These controls are available for the Bilevel and Trilevel video format types Interlaced and Progressive are the only available Scan Types The Scan Rate enables you to trigger on...

Page 470: ...Width Qualification options In some instruments the source selection is a drop down list and the Trigger if Width Qualification option has Occurs and Logic In other instruments the source selection i...

Page 471: ...Level control The trigger can also be qualified by the logical state of other channels or by a bus pattern MSO70000C DX Series only When you click on the trigger window controls the trigger graphic sh...

Page 472: ...he Trigger Window screen on DPO and MSO instruments is the same except for the source selection and Trigger if Window Qualification options In a DPO Series instrument the source selection is a drop do...

Page 473: ...terms of time by using the Inside t or Outside t selections or by the logical state of other channels using the Trigger When Logic option The Trigger When controls change as you select different combi...

Page 474: ...he buttons Trigger Summary A Only A B Trig after Time A B Trig on nth Event Horizontal Delay Mode A B Seq Delay Counter Measurements The following screen appears on DPO70000SX Series instruments The f...

Page 475: ...e trigger graphic shows the trigger sequence What do you want to do next Learn about the A Event trigger types Learn about B Event triggers Learn about Trigger Mode and Holdoff Return to the Trigger S...

Page 476: ...al Delay Mode Horizontal Position A Trigger Level Acquisition Delay Trigger Reset Counter Measurements The following screen appears on MSO DPO5000B Series instruments To use 1 From the A Event tab sel...

Page 477: ...acquiring the data When you click on the trigger window controls the trigger graphic shows the trigger sequence What do you want to do next Learn more about horizontal delay Learn more about other tr...

Page 478: ...h Horizontal Delay off For information on the controls click the buttons Horizontal Delay Mode Horizontal Position A Trigger Level Trigger Reset Counter Measurements The following screen appears on DP...

Page 479: ...ode off the instrument looks for the trigger conditions specified by the A event When the instrument recognizes the A event it acquires the posttrigger data TIP Use the front panel Horizontal Position...

Page 480: ...ot available with MultiScope Triggers For information on the controls click the buttons Trigger Level Trigger Delay Horizontal Delay Reference Point Trigger Reset Counter Measurements The following sc...

Page 481: ...trigger conditions specified by the A event After the A event arms the trigger system the instrument delays the trigger for the period of time specified by the Trig Delay control After the trigger de...

Page 482: ...riggers For information on the controls click the buttons Trigger Level Trigger Delay Horizontal Position Trigger Reset Counter Measurements The following screen appears on MSO DPO5000B Series instrum...

Page 483: ...e trigger conditions specified by the A event After the A event arms the trigger system the instrument triggers on the next B event that occurs after the time specified by the Trig Delay control What...

Page 484: ...tons Trigger Level Trigger Events Horizontal Delay Reference Point Trigger Reset Counter Measurements The following screen appears on MSO DPO5000B Series instruments To use 1 Set up the A Event and B...

Page 485: ...instrument looks for a specific number of B events before triggering The instrument then delays the acquisition for the amount of time specified by the Horiz Delay control After the delay time has ex...

Page 486: ...Events Horizontal Position Trigger Reset Counter Measurements The following screen appears on MSO DPO5000B Series instruments To use 1 Set up the A Event and B Events from their individual tabs NOTE...

Page 487: ...the instrument looks for a specific number of B events before triggering and displaying the data What do you want to do next Learn more about other trigger types Return to the Trigger Setup control w...

Page 488: ...buttons Trigger Level Trigger Events B Event Scan Horizontal Position Trigger Reset Counter Measurements To use 1 Set up the A Event and B Events from their individual tabs 2 Select the A B Seq tab th...

Page 489: ...Event value and is advanced to the End Event value by one of three Advance Methods Sequential Random and Toggle What do you want to do next Learn more about B Event scan setup Learn more about other t...

Page 490: ...the B event and adjust the Ref Point and B Trig Level controls as necessary 4 Set the B Event Scan button to On Click Setup button to set the values for B Event Advance mode Start Event End Event and...

Page 491: ...ig menu select B Event Delayed Trigger Setup To use Use the controls is this window to define the parameters for the B Event trigger For information on the controls click the buttons NOTE On MSO DPO50...

Page 492: ...s Learn about Trigger Mode and Holdoff Return to the Trigger Setup control window Trigger graphic The trigger graphic displays a visual indicator of the trigger criteria Depending on the trigger type...

Page 493: ...ge trigger The following screen capture shows the same waveform acquired using edge trigger and visual trigger with two visual trigger areas controlling which triggered waveforms are displayed Use the...

Page 494: ...reas on and off Check the Trig Expression check box to toggle display of the trigger expression on and off Check the Area Settings check box to toggle display of the visual trigger area settings on an...

Page 495: ...ng a box on the display Then select Visual Trigger Add Visual Trig Area In the same menu you can hide or show all visual trigger areas and toggle visual triggering on and off Trigger setups DPO70000SX...

Page 496: ...dicating whether the channel is included or excluded in the area Creating a visual trigger area adds the area to the visual trigger expression Clicking a waveform handle before creating a visual trigg...

Page 497: ...rigger area and select Visual Trigger Area from the menu Overview Use the right click menu to control visual trigger areas qualification expressions and visual triggering Trigger setups DPO70000SX MSO...

Page 498: ...qualification expression Must be inside specifies which channel the area must be inside Must be outside specifies which channel the area must be outside Edit Equation opens the Qualification Expressi...

Page 499: ...ut recalling a visual trigger setup Learn about visual trigger qualification expressions Define an area In the Visual Trigger window click Visual Setup Overview Use these control to change the setting...

Page 500: ...rotate the area enter the number of degrees in the Flip and Rotate Area text box To flip the area 180 press the Flip Horiz or Flip Vert buttons To move the active vertex enter a Horizontal or Vertica...

Page 501: ...nly if matches this condition graphic displays the triggering conditions required to display triggered waveforms on the instrument Check the Display Visual Trigger on Screen check box to toggle displa...

Page 502: ...t controls to add additional numbers and operators to complete the math expression Connect an external keyboard and use it to enter the expression Create an expression using an external text editor or...

Page 503: ...about recalling a visual trigger setup Learn about visual trigger qualification Edit the visual trigger expression From the Visual Trigger menu double click the Visual Trigger expression To use Click...

Page 504: ...l trigger setup Learn about recalling a visual trigger setup Learn about visual trigger qualification Text edit controls Use the following controls to edit the expression Control Description Moves the...

Page 505: ...2 Click Save to save the visual trigger setup or cancel to close the dialog box without saving What do you want to do next Learn about visual trigger setup Learn about creating a visual trigger area...

Page 506: ...stored from an internal file on your hard drive as soon as you click the Open button What do you want to do next Learn about visual trigger setup Learn about creating a visual trigger area Learn about...

Page 507: ...ance trigger positioning click in the option boxes To set the Trigger Holdoff time manually click Time and then use the multipurpose knob to set the holdoff time The Holdoff range is 250 ns minimum ho...

Page 508: ...ed Holdoff is usually specified in units of time but other units may be used Trigger graphic The trigger graphic displays a visual indicator of the trigger criteria Depending on the trigger type the i...

Page 509: ...pplied NOTE You can use Trigger Position Correction in any combination for A Event B Event or Sequence triggers including Trigger After Time and Trigger on nth Event It is fully compatible with all ac...

Page 510: ...Trigger setups 478 DPO70000SX MSO DPO70000DX MSO DPO70000C DPO7000C and MSO DPO5000B Series...

Page 511: ...parameters for live channel waveforms The example windows below do not show controls for attached probes Click for DPO7000C and MSO DPO5000B Series instruments Click for MSO DPO70000C DX SX Series ins...

Page 512: ...ts and DPO70000DX instruments with option MSOU installed Termination Voltage is available only on MSO DPO70000DX SX Series instruments Input Setup iCapture Setup Viewing the Analog Characteristics of...

Page 513: ...nt to change The controls for each channel are independent What do you want to do next Continue to learn about the Vertical Setup controls Set up input channels NOTE iCapture is available only on MSO7...

Page 514: ...uments TIP You have the option to add user defined labels for the analog reference bus and digital sources Behavior When the Analog Input Normal button is selected the remaining display probe and zoom...

Page 515: ...on DPO7AFP required on DPO70000SX instruments Overview Use the Zoom control window to select the Zoom Area and Source assign and to lock and scroll the zoomed areas The following screen appears on MSO...

Page 516: ...n and Zoom windows from the front panel Selecting Zoom 1 4 as the Zoom Source allows you to zoom on zoom TIP You can easily zoom in on a waveform by clicking and dragging a box around the area of inte...

Page 517: ...Waveform controls Set the Position Offset From the Vertical or Math menus select Position Scale To use Use the Position Offset control window to set the vertical position and offset of the selected wa...

Page 518: ...en by clicking and dragging it to the desired location Behavior The label applies only to the selected waveform Each waveform channel math or reference can have its own label If you do not enter a lab...

Page 519: ...in purpose of the Offset control is to avoid clipping the waveform The Offset control adjusts the DC bias level that is subtracted from the input signal of a channel and establishes the level at the v...

Page 520: ...to the incoming waveform Offset removes clipping By setting an offset to vertically shift the acquisition positive relative to the waveform the clipped portion can be acquired Termination controls Fr...

Page 521: ...ation The instrument reduces the maximum vertical scale setting for the channel to 1 V from 10 V to 10 V from 100 V with a 10X probe attached since the amplitudes appropriate for the higher settings w...

Page 522: ...ve Probes Active voltage probes often contain preamplifiers that provide high resistance and low capacitance at the probe tip Some active voltage probes contain differential amplifiers that provide hi...

Page 523: ...ols to change the settings after you set up the Vertical Setup Control Window Deskew DPO70000SX series 1 To deskew TekConnect channels connect the channel inputs through the cables you will use to tak...

Page 524: ...he Set To 0 0s button Each input channel has its own deskew settings Deskew compensates individual channels for probes or cables of different lengths The instrument applies the delay values after each...

Page 525: ...t box to select the range of frequencies to acquire Behavior Bandwidth refers to the range of frequencies that the instrument can acquire and display accurately with less than 3 dB attenuation Each in...

Page 526: ...the probe tip for some high performance probes and tips To use To enable Bandwidth Enhancement for a vertical input channel 1 Select a Channel tab in the Vertical Setup window 2 Click Digital Filters...

Page 527: ...ll appear in reverse video if the user requested bandwidth cannot be achieved Force Constant Sample Rate The instrument takes the constant sample rate that you set in the Horizontal control window and...

Page 528: ...status of the Auxiliary trigger input This input is located on the instrument front panel The following screen appears on MSO DPO5000B Series instruments NOTE Some instruments include a Bandwidth cont...

Page 529: ...o compensate the entire signal path This will improve the probe gain and offset accuracy Probe compensation is one of several features you can use to optimize measurement accuracy NOTE The example win...

Page 530: ...p above it is recommended that you do the following to assure measurement quality Use the correct probe tip for your probe This will ensure that the correct bandwidth filter is applied and performance...

Page 531: ...u want to do next Learn more about the Probe Status indicator Learn more about the Probe Compensation controls Go to a step by step procedure to calibrate active probes Go to a step by step procedure...

Page 532: ...screen appears on MSO DPO5000B Series instruments For information on the Probe Setup controls click the button Controls Vertical setups 500 DPO70000SX MSO DPO70000DX MSO DPO70000C DPO7000C and MSO DP...

Page 533: ...al menu select Probe Controls click the Setup button and then open the D15 D0 tab NOTE You must attach a logic probe to the logic probe port on the front panel to activate this setup window Overview U...

Page 534: ...lect Probe Cal Overview Use the Probe Setup control window to compensate a probe this will improve the gain and offset accuracy Probe compensation is specific to the channel to which it is connected N...

Page 535: ...th active probes or older passive probes You can perform a probe compensation at any time to ensure measurements are taken with the highest possible accuracy Perform a probe compensation if you have c...

Page 536: ...do the probe compensation What do you want to do next Go to a step by step procedure to compensate active probes Go to a step by step procedure to compensate passive probes Learn more about the Vertic...

Page 537: ...p control window to compensate for propagation delays of different length probes connected to the vertical input channels Click for DPO7000C and MSO DPO5000B instruments Click for MSO DPO70000C DX SX...

Page 538: ...for MSO Series instruments to deskew the logic probe trigger path The fixtures may require a user supplied pulse generator To use 1 Connect the probes to a signal source 2 Use the horizontal and verti...

Page 539: ...ange the probe you restore a saved setup or you recall the factory setup What do you want to do next Go to a step by step procedure to deskew the probes Learn more about the Vertical controls Probe se...

Page 540: ...e instrument You can change the external attenuation by entering a multiplier or ratio in the top control or by entering a dB value in the lower control When you return the attenuation to the default...

Page 541: ...example the window below contains controls for a Tektronix current probe The following window contains controls for Tektronix TAP1500 TAP2500 type probes This window contains controls for Tektronix T...

Page 542: ...yed for compatible probes Errors may include the following types Tip fault Line open Self test failure of tip and Tip broken Tip temperature change Tip overheating Tip thermal shutdown NOTE When the p...

Page 543: ...ertical input channel Behavior The instrument automatically detects the type of probe that you have installed and displays the probe status information Default The instrument has no compensation data...

Page 544: ...ion on page 513 procedure before it will perform a probe compensation procedure What do you want to do next Learn more about the Probe Compensation controls Go to a step by step procedure to compensat...

Page 545: ...nals with AC components Use this procedure to optimize the acquisition system If the temperature has changed more than 5 C 9 F since the last signal path compensation SPC If using the instrument to me...

Page 546: ...to 15 minutes NOTE Before running the SPC calibration remove all channel input signals NOTE If you use a drive that has not had SPC run while in the current instrument you will see a no prior SPC war...

Page 547: ...n run Signal path compensation status The following table lists the possible status messages for signal path compensation and the recommended action Status Meaning and recommended action Compensated T...

Page 548: ...Vertical setups 516 DPO70000SX MSO DPO70000DX MSO DPO70000C DPO7000C and MSO DPO5000B Series...

Page 549: ...Images From the Edit menu select Copy Setup then open the Images tab Overview Use this control window to copy images to the clipboard for use with other applications DPO70000SX MSO DPO70000DX MSO DPO7...

Page 550: ...to those in the Page Setup dialog box and in the Screen Capture Save Options dialog box The copy operation takes place after you click Copy Clicking OK saves the changes and closes the dialog box but...

Page 551: ...cursors are already active If they are not the cursors will be turned on at the positions they were in the last time they were used Select the Include waveform scale factors check box if you want to i...

Page 552: ...next Learn about copying images Learn about copying measurements Learn about saving waveforms to a file Copy setup Measurements From the Edit menu select Copy Setup then open the Measurements tab Ove...

Page 553: ...Select Graticule to export a bitmap of only the graticule area Select Waveform to export waveform data Select Measurements to export measurement data 4 From the Edit menu select Copy Setup 5 Open the...

Page 554: ...can now paste the bitmap image into other applications 5 If you want to copy another image to the clipboard click Copy on the Edit menu or press Ctrl C on the attached keyboard NOTE If you try saving...

Page 555: ...eview at the bottom of the window Click the Print button when you have selected all of your desired settings Behavior The Type of printing should be Screen copy for all screen shots Print preview dial...

Page 556: ...nting General Purpose Controls Use the following front panel controls DPO7AFP required on DPO70000SX instruments to perform general purpose tasks such as Autoset the instrument send items to the print...

Page 557: ...at a later time Push the FastAcq DPX button to turn Fast Acquisition On or Off The FastAcq button is lighted when Fast Acquisition is turned on Use the Intensity knob to adjust the brightness of the...

Page 558: ...Copying and printing 526 DPO70000SX MSO DPO70000DX MSO DPO70000C DPO7000C and MSO DPO5000B Series...

Page 559: ...to access the Recall dialog box and recall a waveform from memory Behavior Clicking Save or Recall buttons activates the respective dialog boxes Typing a name in the Label text box places the label on...

Page 560: ...To save to a reference location select the Source and then click the reference location in Oscilloscope Memory To save to a file enter a Name use the default file name or click Ato increment file name...

Page 561: ...spreadsheets such as Microsoft Excel txt used for standard text file format dat used for Mathcad and MATLAB NOTE Any characters that you enter in the file name box become part of the file name The ins...

Page 562: ...ing the waveform data is written to C Users Username Tektronix TekScope Waveforms The file is named TDS_REF n WFM where n is the reference waveform number This file is deleted when the reference wavef...

Page 563: ...ory 1 Click on one of the setup location icons under Save in Oscilloscope Memory 2 Enter a Name for the setup by clicking on it once more and modifying the existing text 3 Click the Save button To sav...

Page 564: ...o save the reference waveform data save the reference waveform separate from the instrument setup The instrument preference settings are not saved with the setup They are saved when the instrument is...

Page 565: ...asurements If you want to save the reference waveform data save the reference waveform separate from the instrument setup You can then recall the saved waveform to the setup from the Reference Wavefor...

Page 566: ...Action On Event Overview Use this control window to set up the action save or e mail to perform when an event trigger limit test failure or mask test failure occurs To use 1 From the On the event of...

Page 567: ...parameters for the file 5 In the File Name radio buttons select whether to a Use Date and Timestamp or Custom Base Filename 6 In the Event Limits box enter the number of events that will save files u...

Page 568: ...ct Auto increment file name if you want to save a series of files without typing in a new name each time Select a file extension from the Save as type drop down list The file type will be automaticall...

Page 569: ...ck on one of the setup location icons under Save in Oscilloscope Memory 2 Enter a Name for the setup by clicking on it once more and modifying the existing text 3 Click the Save button To save the set...

Page 570: ...ve the reference waveform data save the reference waveform separate from the instrument setup The instrument preference settings are not saved with the setup They are saved when the instrument is powe...

Page 571: ...xt or csv to save from the Save as type list By default the file will be saved to the C Users Username Tektronix TekScope Data folder 8 Click the Save button to accept any changes and close the dialog...

Page 572: ...Overview Use this dialog box to save screen captures to a file for use with other applications File menu 540 DPO70000SX MSO DPO70000DX MSO DPO70000C DPO7000C and MSO DPO5000B Series...

Page 573: ...the Screen Capture Format drop down list 6 Click OK 7 In the Save As dialog box assign your own file name and then save the file to a specified folder By default the image will be saved to C Users Us...

Page 574: ...ame click the keyboard icon and use the pop up keyboard to enter a unique file name The extension specified in the Save as type list will automatically be appended to the file name Behavior The Save A...

Page 575: ...To save to a reference location select the Source and then click the reference location in Oscilloscope Memory To save to a file enter a Name use the default file name or click Ato increment file name...

Page 576: ...lar files without retyping the entire file name each time 2 Enter a Base file name 3 Click Save For the initial save the default count is 000 but you can overwrite this number if you choose Your first...

Page 577: ...want to save 3 Click the Save button to accept any changes and close the dialog box Behavior Digital data is saved in C Users Username Tektronix TekScope data unless you select a different location S...

Page 578: ...to accept any changes and close the dialog box Behavior Timestamp data is saved in C Users Username Tektronix TekScope data unless you select a different location TIP You can conveniently save time s...

Page 579: ...rms and instrument setups To use Click OK to erase Click Cancel to exit the dialog box Behavior Tek Secure Erase removes for the current user all reference waveform and instrument setup memory locatio...

Page 580: ...the instrument clock is out of sync Behavior The controls in this window operate the same as the standard Windows Date Time Properties dialog box After you click Set time and date now the system date...

Page 581: ...ller button to use your instrument to control other instruments Click the Off Bus button to make the GPIB port unavailable NOTE The Address controls are only available when you are in Talk Listen mode...

Page 582: ...nectivity Learn about GPIB Configuration Control Window External signals control window From the Utilities menu select External Signals or click the Ext Ref button on the Acquisition tab of the Horizo...

Page 583: ...py of the internal reference clock out the connector NOTE Ref Out is not available on MSO DPO70000C DX Series instruments 2 Click Pos or Neg to set the polarity To select the oscilloscope clock refere...

Page 584: ...N BNC connector Touch screen From the Utilities menu highlight Touch Screen then select Calibrate Overview Use the Touch Screen submenu to calibrate the touch screen To use 1 Select Calibrate 2 Follow...

Page 585: ...due to be run again 1 Follow the set up instructions 2 Click the Run SPC button to begin the procedure Behavior The only user accessible calibration is the signal path compensation SPC All other calib...

Page 586: ...on the selected test s or to halt looping as soon as a failure occurs 3 Click the Run button to initiate the tests The Error Log button displays the number of entries in the error log if the number i...

Page 587: ...problems with the instrument NOTE Remove the input signals before running any diagnostics If any diagnostic failures occur record the failure information and then contact your local Tektronix service...

Page 588: ...mail on If you do not have a SMTP server address or e mail address set up you cannot select an event to send e mail on The screen capture waveform and measurement types for the e mail attachments are...

Page 589: ...rement Snapshot option includes the information on measurements snapshot and the Histogram Data CSV option includes the histogram data in CSV format for the e mail The following measurement file forma...

Page 590: ...command The mapping is deassigned until you perform that function again Turn the knob s or scroll with the mouse wheel until the desired selection appears in the readout mapped to the knob Push the F...

Page 591: ...Behavior The confirmation messages give users the chance to change their minds before taking action The messages generally explain the implications of the action All preference settings are saved when...

Page 592: ...is fixed at 12 divisions from the center For example at 100 mV the trigger range is 1 2 V smaller than the standard TTL and ECL levels All preference settings are saved when you turn the instrument po...

Page 593: ...t User Preferences then open the Measurement tab Overview Use the Measurement tab to set the user preference for measurement annotation Annotations show the exact segment of the waveform from which th...

Page 594: ...user preference is set to 1 2 3 the vertical knobs will increase or decrease the scale in a 1 2 3 10 20 30 100 200 300 sequence If the user preference is set to 1 2 5 the vertical knobs will increase...

Page 595: ...on is not selected the function of the front panel Print button is to print When you select the alternate save function for the Print button the oscilloscope updates the File Save As menu under the Fi...

Page 596: ...button in the initial dialog box 2 Use either floating license or fixed license and follow the on screen instructions to install the options Utilities 564 DPO70000SX MSO DPO70000DX MSO DPO70000C DPO70...

Page 597: ...installation 1 In the Menu bar click Analyze 2 Select the application name from the submenu NOTE New versions of software for your instrument are available at our Web site To find application software...

Page 598: ...Option Installation to enable upgrades that you have purchased for your instrument from Tektronix For the most current list of upgrades go to www tektronix com or contact your local Tektronix represen...

Page 599: ...d click Open 3 On successful installation a dialog box appears indicating that the Option is enabled Reboot the instrument for the changes to take effect For example after installing DPOFL PTD TDSPTD...

Page 600: ...license You must minimize the oscilloscope application to see the WEB page Install the fixed license From the Utilities menu select Option Installation Overview Use Option Installation to enable upgra...

Page 601: ...r the option installation key provided by Tektronix in the Options dialog box 3 Click OK On successful installation a dialog box appears Reboot the instrument for the changes to take effect Utilities...

Page 602: ...es menu select Option Installation Overview Use the Option Installation to return the floating license for the installed Option s that you purchased for your instrument from Tektronix Utilities 570 DP...

Page 603: ...TD TDSPTD in the Tekscope Option Installation dialog box and click the Return button when the license is still valid or click Exit Key button when the license has expired A dialog box appears Click OK...

Page 604: ...rs on completing the return of a floating license Click Close 6 A dialog box appears indicating that the selected Option is disabled Reboot the instrument for the changes to take effect For example af...

Page 605: ...Utilities DPO70000SX MSO DPO70000DX MSO DPO70000C DPO7000C and MSO DPO5000B Series 573...

Page 606: ...Utilities 574 DPO70000SX MSO DPO70000DX MSO DPO70000C DPO7000C and MSO DPO5000B Series...

Page 607: ...y result in the instrument not meeting warranted performance levels at those volts div settings Compensate active probes NOTE This online help file supports many oscilloscope models from Tektronix Thi...

Page 608: ...bes to ensure maximum distortion free inputs to the instrument 1 Connect the probe to Ch 1 on the instrument 2 Attach the probe tip and reference lead to the probe compensation terminals 3 Press the A...

Page 609: ...sing proper probing connecting techniques 3 Click here for details on setting up the Vertical input controls NOTE Some probes automatically set their termination and other values 4 Push a front panel...

Page 610: ...its waveform record Equivalent This mode starts when the time base setting horizontal scale is too fast to let the instrument get enough samples with which to create a waveform record using real time...

Page 611: ...tion display settings Click Variable then click in the Time entry box and use the multipurpose knobs to set the persistence time Click Reset to clear the persistence Click Controls to activate the Per...

Page 612: ...shold you want to change click the channel threshold entry box and use the keypad or multipurpose knobs to set the threshold level 7 Click OK What do you want to do next Learn about digital setups Lea...

Page 613: ...e Bus Setup window click Parallel as the Bus Type 2 To add sources to the parallel bus click the Add Sources Select button The Bus Source Selection dialog appears 3 In the Bus Source Selection dialog...

Page 614: ...d configure the bus display Set up an RS 232 serial bus NOTE This online help file supports many oscilloscope models from Tektronix This feature is only available on some models First follow the commo...

Page 615: ...ow click Serial as the Bus Type and select MIPI CSI 2 from the drop down list 2 To set the channel type click the Channel Type entry box and select Analog or Digital from the drop down lists 3 To sele...

Page 616: ...ollow the common steps to set up any bus To continue to set up an CAN serial bus follow these steps 1 To define the bus type in the Bus Setup window click Serial as the Bus Type and select CAN from th...

Page 617: ...and select the rate from the drop down list 6 Click the Display tab and configure the bus display Set up an ethernet serial bus First follow the common steps to set up any bus To continue to set up an...

Page 618: ...e click the Threshold entry box and use the keypad or the multipurpose knobs to change the voltage reference threshold level 8 Click the Input box of the SDA component and select the channel connected...

Page 619: ...you see the on off status of the signals and decodes the parallel bus signals for you 1 Connect a logic probe to the logic probe port on the front panel 2 Connect the logic probe tips for data to test...

Page 620: ...box Rising Falling or Either 13 Click the Display tab and set the bus display parameters The instrument acquires decodes and displays data from the bus customized for you to readily interpret and ana...

Page 621: ...roll to the areas of interest to analyze the results What do you want to do next Learn about digital setups Learn about bus setups Learn about bus configuration Learn about symbol tables for parallel...

Page 622: ...the Digital menu select Bus Setup 8 Click an unused Bus from the list B1 B16 9 Click the Bus On button NOTE You can double click the Label box and use the keypad to enter a name for the bus and adjust...

Page 623: ...nt panel button and scroll to the areas of interest to analyze the results What do you want to do next Learn about digital setups Learn about bus setups Learn about bus configuration How to DPO70000SX...

Page 624: ...rol window 12 Enter a unique name for your MyScope control window and click Save This name will display in the upper right corner of the control window when you open it The default location to save My...

Page 625: ...sers Username Tektronix TekScope MyScope 13 From the MyScope menu click Open Control Window 14 Select the MyScope control window that you want from the Open MyScope Control Window dialog box and then...

Page 626: ...vel V Level Cursor 2 Level Cursor 1 Level is cursor displacement from the source ground multiplied by the source in volts div The two cursors may have different sources and therefore can have differen...

Page 627: ...ocedures to quickly take a measurement based on the default settings for High and Low reference levels The instrument must be installed and operating with a signal connected to an input channel and th...

Page 628: ...dow 2 From any tab on the Measurement Setup control window click Ref Levs to open the Reference Levels Setup control window 3 To set how the instrument determines the base and top of the waveform clic...

Page 629: ...Under Setups click Gating from any tab The Measurement Gating control window opens 5 To select how to define the gated area click one of the following Cursor Sets the gated area to the area between t...

Page 630: ...format in the Option dialog 9 Click Save Creating and using math waveforms Using math waveforms Use the same techniques to work with math waveforms that you use with channel waveforms This procedure...

Page 631: ...Time Spectral Var Meas or Filter tabs to display the available functions 6 Click a Primitive button to enter the function in the math expression Click an operand button to insert an operand 7 Use the...

Page 632: ...may be clipped resulting in errors in the derivative waveform Use vertical position and vertical offset to position your source waveform The vertical position and offset will not affect your derivativ...

Page 633: ...ect a Measurements type for more information see Taking Automatic Measurements 4 To display the measurement click Display to toggle it to on NOTE Automatic measurements are not available on digital ch...

Page 634: ...ath tab for the math waveform you want to create then click Display Off On to toggle it on TIP If the waveform that you select already exists its math expression appears in the window You can still us...

Page 635: ...ting Option on some instruments before you can access the Comm triggers TIP Selecting a mask type and standard adjusts the instrument horizontal vertical and trigger settings to those appropriate for...

Page 636: ...y mask hits If there are hits autofit repositions the waveform to minimize hits The number of hits reported is the number after autofit has minimized hits The autofit Config button lets you do the fol...

Page 637: ...y buttons to set mask and waveform polarity Positive tests the positive waveform pulses Negative inverts the mask and tests the negative waveform pulses Both test the first half of the waveforms in po...

Page 638: ...selected vertex vertical position 10 To add a vertex select the closest vertex that is clockwise from where you want to place a new vertex Click Add to add a vertex midway between the selected vertex...

Page 639: ...5 Click the Controls button to open the Mask Edit control window 6 Click in the Segment entry box and use the up down arrows multipurpose knob or keypad to select a segment to edit The selected activ...

Page 640: ...ndustry standard masks are available with the Mask Testing feature The Comm Trigger is designed to work with communications mask and standard testing Mask testing automatically uses Communication trig...

Page 641: ...Utilities menu select E mail on Event Setup 2 Enter the recipients e mail addresses in the entry box 3 Click the Config button and then enter your SMTP server address then click OK 4 Select the events...

Page 642: ...e button to insert the magnitude function into the math expression 7 Click the channel that is related to the time domain signal that you want to view as an FFT waveform The respective channel is inse...

Page 643: ...ed spectral waveform click Avgs to set the number of averages See the step Apply Averaging Creating a Math Waveform for details 7 Under Source select the tab and channel number for the source waveform...

Page 644: ...formation on setup for phase displays Click Unwrap to specify whether to unwrap phase in a spectral analysis phase waveform To set the level in dB that a magnitude in the spectrum must exceed to have...

Page 645: ...n seconds The gate position and gate duration must be within the acquisition To set the gate position click in the Gate Position entry box and use the multipurpose knobs or pop up keypad to adjust the...

Page 646: ...rough a suitable probe 2 Press Default Setup 3 Press Autoset 4 From the toolbar click Vertical and select the Chan 1 tab 5 Click in the Offset entry box and then use the multipurpose knobs or keypad t...

Page 647: ...you want to save as a reference 4 Under Save in Oscilloscope Memory click Ref 1 4 to save the waveform to nonvolatile memory You can also select a location from the Save in drop down list to save the...

Page 648: ...E When you recall a reference waveform from a file a new file containing waveform data is created in C Users Username Tektronix TekScope Waveforms The file is named TDS_REFn WFM where n is the referen...

Page 649: ...t NOTE Options that are grayed out are not available with the currently selected data destination 11 Under Waveform Detail select the Include waveform scale factors check box if you want to include wa...

Page 650: ...Measurements tab b Select either Displayed Measurements make sure that measurements are displayed in the graticule via the Measurement Control window or Histogram Data c If you clicked the Displayed M...

Page 651: ...from each other by a fixed amount of time Using roll mode Set up roll mode Use the following procedure to set up roll mode acquisitions 1 Power on the instrument and set up the horizontal and vertica...

Page 652: ...lloscope has recognized a valid trigger and is filling the posttrigger portion of the waveform If Ready is on the oscilloscope can accept and is waiting for a valid trigger If Arm is on the trigger ci...

Page 653: ...evel knob The instrument sets the trigger level to the halfway point between the peaks of the trigger signal This function has no effect for the other trigger types You can also click Set 50 in the Tr...

Page 654: ...ue The holdoff range is 250 ns to 100 s Select a preset trigger 1 Push the front panel Advanced button 2 In the Trigger control window select the A Event tab 3 Click a Trigger Type button to select a...

Page 655: ...on has no effect if the acquisition system is stopped Trigger on a sequence Use the following procedure to set up the instrument to trigger on a sequence 1 Connect a signal to an input channel 2 Set t...

Page 656: ...n the Trigger control window 2 Open the A B Seq tab then click the A Only button Trigger on an event after a specified delay Use the following trigger sequence to trigger the instrument on the B event...

Page 657: ...he Trigger if Glitch drop down list 9 You can also Trigger on a logic qualified pulse using the Trigger if Glitch Logic selection When the instrument detects that a glitch condition has occurred it ch...

Page 658: ...it checks the logical state of two other available channels and triggers only if their conditions are met Each channel can have a value of high H low L or don t care X A value is considered high if th...

Page 659: ...el from the drop down list 5 To set the trigger timeout level click in the Level entry box then use the multipurpose knob or pop up keypad to enter the voltage level NOTE You can set the level to a va...

Page 660: ...definitions of the logic functions for pattern triggers 7 From the Pattern drop down list select True to trigger the instrument when the logic patterns are true Select False to trigger the instrument...

Page 661: ...logic input logic high low or don t care the logic inputs are the instrument channels The Boolean logic function AND NAND OR or NOR The condition for triggering the Boolean function becomes True logic...

Page 662: ...POS rising or NEG falling edge for the Clock Channel Channel 4 6 To set the Input Threshold voltages click one of the Threshold Presets buttons to define the level type If you selected USER click in t...

Page 663: ...ion time slew rate of the positive going edges of pulses The edge must first cross the lower threshold and then cross the upper threshold Neg Monitors the transition time slew rate of the negative goi...

Page 664: ...r all standard trigger types Refer to trigger mode and set holdoff to learn more about trigger mode and holdoff Trigger on setup hold time violations Use the following procedure to detect and trigger...

Page 665: ...to maintain the limit In most cases you will enter positive values for both setup and hold time Positive values set the instrument to trigger if the data source is still settling inside the setup time...

Page 666: ...he values 6 To qualify the window trigger select from the Window Event drop down list combinations The Window Event controls change as you select different combinations of options Enter Window and Occ...

Page 667: ...t the scan rate that includes your signal 6 Select a value from the Trigger On drop down list to specify where the trigger will occur on the video signal Field Use the drop down list to select either...

Page 668: ...following procedure to set up the instrument to trigger on a parallel or on a serial bus 1 Push the front panel Advanced button 2 In the Trigger control window open the A Event tab 3 Select Bus from t...

Page 669: ...n as marked events on screen and they can also be viewed in the Results tab of the Mark Table 7 From the Analyze menu select Search and then select the Results tab The results of the search are displa...

Page 670: ...p 9 The results of the edge trigger and the visual trigger are also displayed in the search results NOTE Mark All Trigger Events in Record is not available with some trigger types How to 638 DPO70000S...

Page 671: ...he screen What do you want to do next Taking Automated Measurements on Histogram Data Take automated measurements on histogram data Use the following steps to take automatic measurements on histogram...

Page 672: ...e display of the selected histogram on and off TIP To clear histogram data right click in the display area and then select Clear Data from the short cut menu Using display modes Display a waveform Thi...

Page 673: ...it the screen and add a zoom graticule 4 To zoom on multiple areas repeat steps 2 and 3 5 Push the front panel MultiView Zoom button to select the axis you want to focus on in the zoom graticule The Z...

Page 674: ...he display of an analog oscilloscope Areas of the waveform with the highest sample density appear in lighter shades while the areas of lower sample density appear in darker shades Gray Use this palett...

Page 675: ...o blue tones Decreasing the hue value changes the color to red tones For best viewing set the lightness to 50 Higher numbers tend to wash out the color and lower numbers can darken the color and make...

Page 676: ...the setup then click Recall NOTE The data in reference waveforms is not saved with instrument setups However the setup information does include which reference waveforms are turned on and which are us...

Page 677: ...r software and the basic Windows instructions to install and remove your application NOTE When adding software to the instrument avoid making changes to system files See Windows Interface for more inf...

Page 678: ...select Extended Desktop NOTE Do not change the display settings for the internal monitor The display settings must be 1024 x 768 pixels and colors must be set to 32 bit color Ensure that the second m...

Page 679: ...e screen To restart the instrument application launch the TekScope icon from the Windows taskbar Start menu or your desktop Minimize the instrument application To temporarily close the TekScope applic...

Page 680: ...contrast to your preference You need to have a mouse attached to perform this procedure 1 Right click on the Windows desktop 2 Select Properties 3 Select the Settings tab 4 Click Advanced 5 Select th...

Page 681: ...more general terms synchronizes the capturing of digital samples in the Acquisition system to the trigger events generated from the Trigger system Acquisition system This system converts and stores th...

Page 682: ...part of the multi acquisition record that these modes produce The process loops back to step 3 to acquire additional records until the number of acquisitions required for the acquisition mode current...

Page 683: ...s to characterize a device The connection to the instrument depends on your application Up to four acquisition channels are available You can display each channel as a waveform or use the channel data...

Page 684: ...next Learn more about Autoset Learn about horizontal acquisition interrelated parameters Bandwidth enhancement Bandwidth enhancement is a DSP implemented FIR filter that flattens the oscilloscope freq...

Page 685: ...he instrument determines the coupling and termination required by the probe either implicitly because of the TekProbe TekVPI interface or through performing a probe compensation the instrument sets th...

Page 686: ...tical acquisition window and the displayed waveform The vertical volts per division you set determines the vertical size of the acquisition window allowing you to scale it to contain all of a waveform...

Page 687: ...Figure A Figure B Oscilloscope reference DPO70000SX MSO DPO70000DX MSO DPO70000C DPO7000C and MSO DPO5000B Series 655...

Page 688: ...define the horizontal acquisition window parameters that determine which segment of an incoming signal that becomes the waveform record following acquisition For additional information see Waveform Re...

Page 689: ...is the Sample Interval that changes to accommodate the window Time Duration and its scale setting and the Record Length setting The elements behave as follows If the Record Length or Time Duration ch...

Page 690: ...earn more about the horizontal acquisition window Incompatible acquisition features The following table shows which acquisition features and modes are incompatible with other features or modes Incompa...

Page 691: ...fail No signal present Signals with extreme or variable duty cycles Signals with multiple or unstable signal periods Signals with too low amplitude No recognizable trigger signal Signals with a freque...

Page 692: ...mpling an analog signal converting it into digital data and assembling it into a waveform record which is then stored in acquisition memory Sampling is the process that provides a waveform record per...

Page 693: ...hin the data stream These parameters include the following Sample Interval This is the time between sample points taken during acquisition Record Length This is the number of samples required to fill...

Page 694: ...n use the instrument will not be able to get enough samples to create a waveform record At that point the instrument will switch from real time to equivalent time to obtain additional samples Learn ab...

Page 695: ...need to set up the trigger conditions The next figure shows how triggers fit into the overall instrument operation Triggers create meaningful waveforms from signal displays This instrument has simple...

Page 696: ...gital channels These sources are available on MSO instruments You can select any combination of digital channels Bus This source is used to trigger a parallel bus or a serial bus You can include any c...

Page 697: ...lloscope models from Tektronix This feature is only available on some models Glitch trigger Transition trigger Pattern trigger Runt trigger Window Trigger State trigger Width trigger Timeout trigger a...

Page 698: ...es not acquire a waveform and the last waveform record acquired remains frozen on the display If no last waveform exists no waveform is displayed Auto trigger mode enables the instrument to acquire a...

Page 699: ...ut trigger modes Trigger coupling Trigger coupling determines what part of the signal is passed to the trigger circuit Edge triggering can use all available coupling types AC DC Low Frequency Rejectio...

Page 700: ...uncover information that helps you find the source of the glitch Learn about trigger coupling Learn about trigger slope Learn about glitch triggers Delayed trigger system You can trigger with the A Ma...

Page 701: ...a Value click the Edit button You can set the instrument to trigger on an 8b10b bus when the instrument detects a Pattern Character Symbol Error or Any Control Character For some Trigger On selections...

Page 702: ...ny of the preconditions selected for the logic inputs5 are TRUE then the instrument triggers NOR Clocked NOR If none of the preconditions selected for the logic inputs6 are TRUE then the instrument tr...

Page 703: ...ve to the clock When you use setup hold triggering you define The channel containing the logic input the data source and the channel containing the clock the clock source The direction of the clock ed...

Page 704: ...mes skew the setup hold violation zone to locate it before or after the clocking edge as shown above in the center and bottom waveforms The instrument can then detect and trigger on violations of a ti...

Page 705: ...ecified fields or lines of a composite video signal You can select from several preset video signal formats or set a custom format NOTE Only composite signal formats are supported Graphic display form...

Page 706: ...curs after the trigger delay time You can set the trigger delay time with the keypad or the multipurpose knobs Trigger on nth Event After the A trigger arms the trigger system the instrument triggers...

Page 707: ...an be used with any trigger setup You can turn horizontal delay on and off from the front panel the Horizontal Acquisition control window and many of the Trigger control windows The next figure compar...

Page 708: ...w a summary of triggering and horizontal delay Triggering and horizontal delay summary The next figure shows all combinations of triggering and horizontal delay Oscilloscope reference 676 DPO70000SX M...

Page 709: ...form display Display overview This instrument includes a flexible customizable display that lets you control how waveforms appear The next figure shows how the display features fit into the overall in...

Page 710: ...eadout Adjust and display the parameters controlled by the multipurpose knobs 4 Display Live reference and math waveforms are displayed here along with cursors 5 Waveform Handle Click and drag to chan...

Page 711: ...ar modes and select toolbar extensions 2 Drag cursors to measure waveforms on screen 3 Drag the horizontal reference marker to reposition a waveform 4 Click the waveform handle to assign the multipurp...

Page 712: ...hese elements through the Display menu Colors command Display menu Display Palette submenu or Display Setup control window Colors tab Disp toolbar button Normal Displays hues and lightness levels for...

Page 713: ...one of the acquisition display settings Use for displaying points that may occur outside the normal acquisition envelope Variable persistence Accumulates record points on the waveform for a specified...

Page 714: ...intouts or for other instrument users Display Turns the text display on and off Properties Opens the Text Properties control window for positioning of the text on the display Clear Erases the entire t...

Page 715: ...4 Channels are predefined Push the Vertical CH n button to toggle the channel display on or off Logic D 15 0 Channels are predefined From the Digital Setup control window click the D15 D8 or D7 D0 tab...

Page 716: ...remember about how horizontal operations relate to the waveform types Reference waveforms are shown with the horizontal settings in effect at the time that they are saved You cannot adjust these sett...

Page 717: ...cquisition waveform or one of the zoomed areas When zooming either horizontally or vertically Zoom expands or contracts the waveform by the Zoom Factor What do you want to do next Go to a step by step...

Page 718: ...record This setting is best for examining waveforms that have no large flat portions at a common value such as sine waves and triangle waves almost any waveform except for pulses See the next figure R...

Page 719: ...histogram based High and Low values as follows 1 It makes a histogram of the record with one bin for each digitizing level 256 total 2 It splits the histogram into two sections at the halfway point b...

Page 720: ...use gated measurements it is the location of the right vertical cursor Hysteresis is the hysteresis band 10 of the waveform amplitude It is used in MCross1 MCross2 and MCross3 calculations For example...

Page 721: ...the polarity of first crossing no default It can be rising or falling StartCycle is the starting time for cycle measurements It is a floating point number with values between 0 0 and Record Length 1 0...

Page 722: ...litude voltage measurement The area over one waveform cycle For data not cyclical you might prefer to use the Area measurement If StartCycle EndCycle then return the interpolated value at StartCycle D...

Page 723: ...ctually a group of measurements To get a specific delay measurement you must specify the target and reference crossing polarities and the reference search direction Delay the time from one MidRef cros...

Page 724: ...his sample continue the search to find the first negative crossing of HighRef The time of this crossing is THF Use linear interpolation if necessary Figure 22 Fall Time 3 From THF continue the search...

Page 725: ...tart Otherwise Details of the integration algorithm are given later Integration Algorithm Mean dBm The normalized mean If the waveform source is from an optical probe this can give average optical pow...

Page 726: ...then NegativeWidth MCross2 MCross1 else NegativeWidth MCross3 MCross2 Waveform Start Optical power See Mean dBm Peak to peak Amplitude measurement The absolute difference between the maximum and minim...

Page 727: ...e waveform crossing in the same direction polarity as that found MCross1Target for the target waveform is found 4 The phase is determined by the following If the target waveform leads the reference wa...

Page 728: ...of this crossing is the low rise time or TLR Use linear interpolation if necessary 3 From TLR continue the search looking for a crossing of HighRef Update TLR if subsequent LowRef crossings are found...

Page 729: ...en 0 0 and RecordLength 1 0 If A and B are integers then where s is the sample interval Similarly W t is the sampled waveform t is the continuous function obtained by linear interpolation of W t A and...

Page 730: ...etermined in the following manner 1 Step through the waveform from Start to End until the sample min and max pair do not straddle MidRef Figure 24 Choosing Minima or Maxima to Use for Envelope Measure...

Page 731: ...button and the instrument is waiting for a trigger LTGT This is a measurement warning that s only generated in some cases where null samples are present in a waveform such as an ET waveform or a math...

Page 732: ...measurement Waveform does not close ref These errors can occur when the measurement system cannot correctly find edge crossings for the specified edge crossing and waveform No second ref crossing No...

Page 733: ...y X and Y position in the waveform display area represented by the intersection of a vertical and horizontal cursor line Screen cursors measure both vertical parameters typically volts and horizontal...

Page 734: ...here Ch1 is the waveform shown above Avg is the average function And get this math waveform Offset position and scale The settings that you make for offset scale and position affect the math waveform...

Page 735: ...orm Or you can integrate a single waveform into an integral math waveform as shown below With spectral analysis you can analyze waveforms in the frequency domain The interface is similar to a dedicate...

Page 736: ...as Math1 Math2 Ch3 If you do the Math1 definition is rejected with an error because a circular definition is not allowed Fast Acquisitions Math is not allowed in Fast Acquisitions mode Roll Mode Math...

Page 737: ...not available on digital channels What do you want to do next Learn more about cursor measurements Go to a step by step procedure for creating math waveforms Go to a step by step procedure for using...

Page 738: ...he Predefined Expressions or Equation Editor control window To help you create valid math waveforms the following tools will block most illegal entries by disabling any window element that would creat...

Page 739: ...suring power and energy such as in switching power supplies Characterizing mechanical transducers as when integrating the output of an accelerometer to obtain velocity The integral math waveform deriv...

Page 740: ...he libraries provides two functions MyAdd and MyMultiply each of which takes two vector inputs the following math definitions are valid Math1 MyAdd Ch1 Ch2 Math1 MyMultiply Ch1 Ch2 Ch3 Math2 MyAdd Ref...

Page 741: ...e is built on top of INormalizedVector An IFastFrame is a grouping of INormalizedVectors IFastFrames are generated when FastFrame is enabled on the instrument The number of frames is stored in the mem...

Page 742: ...mDB will be zero displays as clear The advanced user can change the vertical scale and position as well as the horizontal scale and spacing by setting Vertical Scale Vertical Position Horizontal Scale...

Page 743: ...ffset HOffset Horizontal position HPosition Horizontal units HUnits Additionally the following information about the target math is available LPCT MPCT HPCT HYSTPCT The recommended method for looking...

Page 744: ...rames If the plugin produces IFastFrame or IWaveformDB it is called once per FastFrame acquisition When the plugin takes as input an IFastFrame it is given all of the frames at once If the plugin take...

Page 745: ...he class with the math attribute We call our plugin Add and it takes two INormalizedVectors as input and produces an INormalizedVector namespace MyMathPlugins class MyMath Add wfm wfm This plugin adds...

Page 746: ...as been called as part of the math expression firstTime is true In this case the varargin array consists of the record length and the sample rate recordLength varargin 1 sampleRate varargin 2 and the...

Page 747: ...process obj varargin result varargin 1 end end end The constructor is called the first time the analysis function is used in a specific math expression on the instrument Here anything that requires s...

Page 748: ...Displays log data in dB dBm or linear mode You can display the real or imaginary parts of the spectral magnitude only Reference level offset and reference level controls give complete control over the...

Page 749: ...the end of the acquired waveform You may set duration using the record length control or the sample rate control Resolution determines the time between samples Duration is kept constant as resolution...

Page 750: ...of all spectral math controls Learn more about spectral math controls Go to a step by step procedure for defining a spectral math waveform Oscilloscope reference 718 DPO70000SX MSO DPO70000DX MSO DPO...

Page 751: ...f the gate affects the resolution bandwidth of the spectral analyzer Gate width has units of seconds The resolution bandwidth directly controls the gate width but the numerical value is entered in uni...

Page 752: ...f the sample rate Therefore if the span is set to full the center frequency control will not have any adjustment range and will remain at one half of the sample rate It is necessary to decrease the sp...

Page 753: ...of all spectral math controls Learn more about spectral math controls Go to a step by step procedure for defining a spectral math waveform Oscilloscope reference DPO70000SX MSO DPO70000DX MSO DPO7000...

Page 754: ...ex data point in the spectrum and Ref is the Reference Level Offset value dBm This selects dB as described in the above equation but also sets the Reference Level Offset to a value that is equivalent...

Page 755: ...el to the equivalent of 1 mW into 50 ohms Real and imaginary magnitudes You can display the linear magnitude of the real data or the imaginary data in the spectrum This is useful if you process the sp...

Page 756: ...ap The spectral analyzer produces phase values from to radians or 180 to 180 degrees However when you perform impulse response testing and the phase is continuous then phase values outside these range...

Page 757: ...panel trigger level control Adjust the front panel horizontal Position control Spectral analyzer windows In the time domain a window is a bell shaped function equal in length to the gate duration For...

Page 758: ...ussian 2 0 0 76 dB 79 dB 50 a 3 75 not cosine series Flattop2 3 8 0 0065 dB 90 dB 50 0 213348 0 206985 0 139512 0 043084 0 003745 Tek Exponential 1 42 0 60 dB 67 dB 20 not applicable 3 dB BW in bins T...

Page 759: ...um Zero phase reference This is the position in the time domain gate that is the reference point for phase in the output spectrum That is if a sine wave input has its peak at the zero phase reference...

Page 760: ...e waveform be stationary within the gated region This means that waveform parameters such as frequency and amplitude should not change significantly as a function of time within the gate region that i...

Page 761: ...in hertz may be computed by dividing the BW in bins by the gate duration in seconds This is also referred to as resolution bandwidth RBW Coherent gain The gain factor normally associated with differe...

Page 762: ...ce position then it reads out as zero phase in the spectrum If the phase is to be correct when doing impulse response testing the impulse in the time domain must be located at this position in the gat...

Page 763: ...s unique because the time domain shape has negative values NOTE The Flattop2 window is useful for high accuracy magnitude measurements for signals that do not require very narrow bandwidth What do you...

Page 764: ...n function transforms into a Gaussian exponential shape in the frequency domain NOTE This window provides optimal localization in both the time and the frequency domain What do you want to do next Lea...

Page 765: ...at do you want to do next Learn about using spectral analyzer windows Hanning Kaiser Bessel and Blackman Harris windows These windows have various resolution bandwidths and scallop losses see the figu...

Page 766: ...Hanning window Kaiser Bessel window Oscilloscope reference 734 DPO70000SX MSO DPO70000DX MSO DPO70000C DPO7000C and MSO DPO5000B Series...

Page 767: ...ext figure This means the data samples in the gate are not modified before input to the spectral analyzer NOTE This window has the narrowest resolution bandwidth of any of the windows but it also has...

Page 768: ...hape is triangular NOTE Use this window for impulse response testing where the 20 position is the zero phase reference point More of the acquired data record length is used to capture the impulse resp...

Page 769: ...increase the sample rate and observe whether any of the harmonics unwrap to different frequency locations Another way to recognize aliasing is to realize that higher order harmonics usually have decre...

Page 770: ...ting persistence data If Autofit makes frequent waveform movements there may be little or no displayed waveform persistence data Segments and mask hits Each mask can have a maximum of 16 segments Segm...

Page 771: ...the included standard telecommunications masks or you can define your own custom masks Selecting a mask automatically sets the instrument communications triggers to properly display most communication...

Page 772: ...iagram measurements are based on the power level the voltage level or the time locations of edges within each acquisition The next figure shows the areas of an eye diagram used to calculate measuremen...

Page 773: ...ps for later recall limited only by the space that you have to store the setups By saving and recalling different setups you can switch from setup to setup without having to first manually record your...

Page 774: ...e reference waveform files The Reference Setup control window includes a pop up keyboard that you can use with your mouse or the touch screen to enter the waveform path name file name and comments The...

Page 775: ...es of which contain the following header information The first header value holds the record length The second header value holds time in seconds between samples The third header value holds the trigg...

Page 776: ...an Inserts the text ATAN into the math expression Auto trigger mode Auto trigger mode causes the instrument to trigger if a valid trigger event does not occur within a set amount of time Auto trigger...

Page 777: ...Use the multipurpose knob to set the bit rate for the trigger CAN format In the CAN Format drop down list specify the format CAN signal type In the drop down Signal Type list select the type of data...

Page 778: ...y box and use the multipurpose knob and to set the threshold on which you want to trigger Different levels are available depending on the code selected The upper level controls the upper level trigger...

Page 779: ...s S400b 491 5 Mb s S800b 983 0 Mb s S1600b 1 966 Gb s XAUI 3 125 Gb s RapidIO 500M 500 Mb s RapidIO 750M 750 Mb s RapidIO 1 0G 1 0 Gb s RapidIO 1 5G 1 5 Gb s RapidIO 2 0G 2 0 Gb s SFI SPI 5 2 5G 2 5 G...

Page 780: ...neutral if there was no change Depending on the code used the pulse form can also indicate that triggering is to occur on an eye diagram To use Click the pulse form you want the instrument to use for...

Page 781: ...e text COS into the math expression Cosh Inserts the text COSH into the math expression The COSH function computes the hyperbolic cosine of x Crossing percent Crossing percent measurement is the eye c...

Page 782: ...ck the Adj buttons to assign the multipurpose knobs to the X axis and the Y axis controls What do you want to do next Learn about cursor types Learn about cursor styles Learn about Cursor Readouts in...

Page 783: ...elay This timing measurement is the time between the mid reference default 50 amplitude point of two different waveforms Trigger delay Use the Trig Delay control to trigger the instrument a specified...

Page 784: ...lignment NOTE This online help supports many oscilloscope models from Tektronix This feature is only available on some models The exact trigger location is used in alignment process to avoid jitter To...

Page 785: ...ls in the Logic Threshold Setup Window What do you want to do next Continue to learn about the Vertical Setup controls Learn about the Logic Threshold Setup menu Set the logic thresholds NOTE The Logi...

Page 786: ...OK button What do you want to do next Learn about the digital channel setup Learn about bus setups Select the display style From the Display menu select Display Setup then open the General tab To use...

Page 787: ...ta stream that shows signal levels and transition regions typically over one cycle The eye diagram has the appearance of an opened eye Compared to a normal instrument display the eye diagram generally...

Page 788: ...color as the original waveform What do you want to do next Learn about changing the overall display appearance Learn about changing the appearance of objects in the display Learn about using text wit...

Page 789: ...oss2mean TCross1mean Edit a user mask Use the Mask Edit Setup to alter a user mask 1 From the Masks menu select Mask Edit Setup to display the Mask Edit control window 2 Under Mask Elements use the up...

Page 790: ...alid when the acquisition mode is set to WfmDB FastAcq is enabled or the measurement is taken on a reference waveform saved in FastAcq mode ExtdB 10 Log pTopMean pBaseMean Eye base Eye Base is the bas...

Page 791: ...in seconds Eye Width tCross2Mean 3 pTopSigma tCross1Mean tCross1Sigma Set up e mail configuration From the Utilities menu highlight E mail on Event then Select Setup from the submenu In the E mail on...

Page 792: ...Server Address box for your instrument to send e mail You can customize your e mail by selecting the number of attempts to make entering a From address setting the time out length entering a login an...

Page 793: ...time out for sending e mail to the server If your mail server requires a login or password enter the login in the Auth Login entry box and the password in the Auth Password entry box Entering a netwo...

Page 794: ...th the CD ROM 020 2513 XX For online information about the VXIplug play driver access the driver online help in from the help system of your software development environment in which you install the d...

Page 795: ...measured over the entire waveform or gated region High impedance probes Most general purpose probes have high input impedance The high input resistance of passive probes typically 10 M provides neglig...

Page 796: ...3 Enter limits Histogram box limits vertical 1 Select a mode 2 Specify the adjustment 3 Enter limits Oscilloscope reference 764 DPO70000SX MSO DPO70000DX MSO DPO70000C DPO7000C and MSO DPO5000B Series...

Page 797: ...al 1 Select a mode 2 Specify the adjustment 3 Enter limits Histogram box location vertical 1 Select a mode 2 Specify the adjustment Oscilloscope reference DPO70000SX MSO DPO70000DX MSO DPO70000C DPO70...

Page 798: ...histograms Hits in box Displays the number of points in or on the histogram box Maximum This voltage measurement is the maximum amplitude It is typically the most positive peak voltage and is measured...

Page 799: ...ram box Half of all acquired points within or on the histogram box are less than this value and half are greater than this value Set the horizontal delay Use the Horizontal Delay Mode control to enter...

Page 800: ...sition Scale control window Overview The horizontal position control moves the trigger point within the acquired waveform To use Click in the Position entry box and use the multipurpose knob or keypad...

Page 801: ...When Horizontal Delay Mode is On use the Acquisition Delay control to enter the delay time You can also enter the delay time with the front panel Horizontal Position knob TIP The Horizontal Delay Mode...

Page 802: ...tion Scale control window Overview To use Click in the Scale entry box and use the multipurpose knob or keypad to change the setting Behavior Use the Scale control to adjust the time base This control...

Page 803: ...number of waveforms for that mode using the multipurpose knob If you use Waveform Database mode use the multipurpose knob to select the number of samples Oscilloscope reference DPO70000SX MSO DPO70000...

Page 804: ...s that you can obtain up to 15 significant bits with Hi Res mode Note that the resolution improvements are limited to speeds slower than 40 ns div Additional resolution bits 2 5 GS s 8 5 bits 1 1 GHz...

Page 805: ...uisition tab Overview Use the Acquisition tab to set the acquisition parameters for all live waveforms The Acquisition controls are common to all channels This means that you cannot acquire in Sample...

Page 806: ...s only appear when the Delay Mode is On Horiz Delay enables the instrument to look for a trigger and then wait a specified time before acquiring data This is useful when you know that the data you are...

Page 807: ...cquisition samples per second The faster the sample rate the more accurately the instrument can represent fine details in a fast signal Duration Readout This readout displays the time to cross the 10...

Page 808: ...ord Length Limit dialog box This feature is available in Automatic mode only Behavior When you change the record length you can choose to hold either the sample rate or the horizontal scale constant T...

Page 809: ...he instrument application to restore proper operation Before installing other desktop applications exit the instrument application To do this click Exit on the File menu Interactions of roll mode and...

Page 810: ...ine to form a logic pattern Each channel can have a value of high H low L or don t care X A value is considered high if the channel input voltage is greater than the specified threshold voltage A valu...

Page 811: ...The hexadecimal pattern uses the symbols 0 1 2 3 4 5 6 7 8 9 A B C D E F and X Five hexadecimal characters will convey all 20 digital inputs The characters show up in the binary hexadecimal ASCII and...

Page 812: ...do you want to do next Learn how to define a pattern for the instrument to use to detect a bus value Learn about bus setups Logic clock inputs For the supported instruments you need to connect one of...

Page 813: ...Use the controls to set up the Logic Pattern or Logic State trigger pattern Click the Edit button to access the Pattern Editor dialog box NOTE The Logic Pattern or Logic State trigger setups include t...

Page 814: ...Although that DC loading degrades amplitude accuracy the lower input capacitance reduces high frequency loading to the circuit under test That makes low input impedance passive probes ideal for timing...

Page 815: ...o align the signal within the mask when autoset is done What do you want to do next Learn about vertical autoset Learn about trigger level autoset Learn about autofit for autoset Select the mask autos...

Page 816: ...dow click the Autoset Config button Overview Use this control to choose whether to have autoset affect the trigger level To use Click the Trigger Level button to turn it off or on Behavior When the Tr...

Page 817: ...eform crosses over an area where two segments overlap both segments will count the waveform hit When you set Hit Count to on the counts will appear in thePass Fail Test Summary Math arbitrary filters...

Page 818: ...40 and 0 45 Stop band rejection is typically between 50 and 60 dB Figure 25 Frequency response of the available low pass filters The following table lists the available low pass filters File name Nor...

Page 819: ...nd 0 45 Stop band rejection is typically between 50 and 60 dB Figure 26 Frequency response of the available high pass filters The following table lists the available high pass filters File name Normal...

Page 820: ...filters File name Normalized bandwidth Normalized center frequency bandpass_0 05bw_0 05center flt 0 05 0 05 bandpass_0 05bw_0 10center flt 0 05 0 10 bandpass_0 05bw_0 15center flt 0 05 0 15 bandpass_0...

Page 821: ...floor This is possible because the FFT is an average calculation internally and the averaging function increases the vertical bits of resolution Figure 27 Frequency response of the available band sto...

Page 822: ...ring is well within the pass band of the filter you choose That will insure that only noise is removed Lengths of 3 5 10 20 50 100 and 200 are provided in the library The red trace is for filter lengt...

Page 823: ...from its desired behavior in the frequency range of 0 to 0 025 times the sample rate and also in the range of about 0 475 to 0 5 times the sample rate This type of filter can be used to create quadrat...

Page 824: ...s allows the same set of filter coefficients to operate at all sample rates The ASCII file format is specified as follows Comments are preceded by symbol sampleRate coef1 coef2 coefN sampleRate coef1...

Page 825: ...windows To use Select a Math 1 4 waveform from the drop down list Click in the readout field to enter the math expression Click a button to enter its character or operator into the expression Click t...

Page 826: ...ch channel or math waveform Behavior These controls only affect the selected math waveform The Position control increases or decreases the vertical position of the waveform one pixel at a time The Sca...

Page 827: ...m measurement to the Measure list Histogram measurements WfmCt Median Pk Pk U 1S Hits in Box Max Mean U 2S Peak Hits Min St Dev U 2S 3 Click Display On in the upper right corner of the Measurement Set...

Page 828: ...am box are less than this value and half are greater than this value Histogram Max Displays the voltage of the highest nonzero bin in vertical histograms or the time of the rightmost nonzero bin in th...

Page 829: ...and off Behavior This list contains your selected automatic measurements You can add up to eight automatic measurements to the list The measurements appear on the main screen when the Display button i...

Page 830: ...g measurement is the distance time between the mid reference default 50 amplitude points of a negative pulse The measurement is made on the first pulse in the waveform or gated region Noise peak to pe...

Page 831: ...locks the zoomed areas in their relative horizontal positions Behavior If the zoom areas are unlocked the Horizontal Position and Factor controls affect only the selected zoom area The Horizontal Pos...

Page 832: ...e Time entry box and use the multipurpose knobs to set the persistence time Click Reset to clear the persistence Click Controls to activate the Persistence Control window Click Setup to open the Displ...

Page 833: ...time stamps You can view the time stamps by turning on the readouts Time delta between frames You can calculate the Time Delta differences between Frames and display those differences These time stamp...

Page 834: ...ou change the oscilloscope controls while Fastframe is running the controls will be returned to the user and the maximum number of frames may not be captured Behavior The Frame Size is the record leng...

Page 835: ...Overlay is selected two additional fields are displayed Start Frame and Number of Frames These fields allow you to select a range of frames that are displayed in the color of the Source channel NOTE I...

Page 836: ...rame For example if channels and maths are set to frame 2 and references are set to frame 3 then after selecting Lock Channel Math Ref an increment of 1 frame will cause channels to display frame 3 an...

Page 837: ...rview Use the Waveform Save Options dialog box to set the parameters for the waveform you want to save Oscilloscope reference DPO70000SX MSO DPO70000DX MSO DPO70000C DPO7000C and MSO DPO5000B Series 8...

Page 838: ...scale factors check box if you want to include waveform details like header information record length sample interval trigger point information and so on If you only want to include the waveform volt...

Page 839: ...strument The values in the first to fourth columns are same for all the waveforms So you need to export these columns only once Set mask test controls From the Mask menu select Mask Controls You can a...

Page 840: ...the delay between two waveforms To use 1 Select the source waveforms from the Channels Selector 2 Use the other controls in the control window to select the Delay Edges Mid Reference point Mid2 Refer...

Page 841: ...this level click the control and then set the level with the multipurpose knob Mid2 reference The Mid2 reference defines the middle reference used on the second waveform specified in delay or phase me...

Page 842: ...where 360 comprise one waveform cycle The measurement is taken at the voltage level specified by the Mid Reference control Mid reference The Mid reference defines the middle reference level of a wave...

Page 843: ...interface What do you want to do next Learn about the MyScope Setup and Edit control windows Select timestamp save options From the File menu highlight Save As then select Timestamp Table from the sub...

Page 844: ...on to allocate half of the available display for the zoomed graticule and half of the available display for the acquisition window Click 80 20 to allocate 80 of the available display for the zoomed gr...

Page 845: ...idth controls Offset and clipping Clipping The circled portion of the incoming waveform is clipped and therefore is not acquired The clipped portion cannot be processed or displayed Offset can move th...

Page 846: ...rity Select the polarity from the drop down list Click Positive to trigger on a positive going pulse Click Negative to trigger on a negative going pulse Click Either to trigger the instrument on eithe...

Page 847: ...spectral phase waveform The math waveform is based on the last active waveform on the screen A spectral waveform is created from a channel waveform you cannot create a spectral waveform from math or...

Page 848: ...ty Factor PTopmean PBasemean PTopsigma PBasesigma Recall a template waveform Use this procedure to recall a template waveform to a reference location 1 From the File menu click Recall 2 Under Recall W...

Page 849: ...saving masks is C Users Username Tektronix TekScope Masks If you saved masks in a different folder click the Look in drop down list to find the folder where your masks are saved Horizontal modes From...

Page 850: ...iting takes place when you change the time per division setting You can use the Sample Rate control to temporarily exceed the record length limit by any amount The default maximum limit depends on the...

Page 851: ...l analyzer filters You can adjust the resolution bandwidth in units of hertz The resolution bandwidth is measured at the 3 dB points on each side of the peak spectral lobe of the filter Resolution ban...

Page 852: ...e a mask 1 From the Mask menu select Mask Edit Setup 2 In the Mask Edit control window click Save 3 In the Save As dialog box select User Mask You may need to click More to access the User Mask select...

Page 853: ...an Harris Gaussian Tek Exponential Flattop2 The windows are listed in the order of their ability to resolve frequencies resolution bandwidth You can easily observe the shape of a window in the frequen...

Page 854: ...is in use frequency domain interpolation occurs and there is minimal scallop loss Zero fill cannot be directly controlled it is affected by changing settings of resolution bandwidth or gate width Whe...

Page 855: ...the B Event trigger to capture complex data For information on the controls click the buttons Trigger Summary A Only A B Trig after Time A B Trig on nth Event Horizontal Delay Mode A B Seq No Delay C...

Page 856: ...box and use the keypad or the multipurpose knob to enter a new bit rate Behavior The Standard you select determines the bit rate range Adjusting the bit rate will change the standard to Custom Select...

Page 857: ...controls appear Use these controls to set the clock polarity and clock level threshold Select the serial data standard From the Trig menu select Serial Pattern Setup To use In the Standard drop down...

Page 858: ...t Click the Clear button to clear the text box 3 Click Enter to complete the edit Behavior If you select the hexadecimal format and the binary pattern cannot be displayed in hexadecimal format then th...

Page 859: ...clipping 3 Adjust the Resolution controls to set the number of samples you want to acquire with each acquisition 4 From the Horiz Acq menu select FastFrame Setup 5 Configure the number of frames Fram...

Page 860: ...etup control windows To use Select a Math 1 4 waveform from the drop down list Click in the readout field to enter the math expression Click a button to enter its character or operator into the expres...

Page 861: ...ment to the function This function creates an FFT magnitude waveform Phase Inserts the text SpectralPhase into the math expression Select one of the waveforms as an argument to the function This funct...

Page 862: ...eful for example when Math1 is the spectral magnitude and Math2 is the spectral phase of the same signal Locking forces the magnitude and phase waveforms to track if you change the spectral controls W...

Page 863: ...eforms Click one of the Scale buttons to specify the vertical units in degrees radians or as group delay Group delay unwraps and differentiates the phase All phase values out of the spectral transform...

Page 864: ...ls to specify the vertical scale for magnitude and phase waveforms Click either the Magnitude or Phase button Click one of the Vertical Control buttons to specify the vertical units Click in an entry...

Page 865: ...et the threshold voltage levels for the trigger SPI data source In the drop down list box specify the MISO and or MOSI data source Supported ADEs TekVISA supports these Application Development Environ...

Page 866: ...This coupling attenuates signals above 50 kHz before passing the signal to the trigger circuitry LF Rej This coupling attenuates signals below 80 kHz before passing the signal to the trigger circuitry...

Page 867: ...e Adjust the trigger level with the front panel Trigger Level knob Click in the Level entry box in a Trigger control window and use the multipurpose knob to adjust the level Select the trigger on opti...

Page 868: ...type occurs and the bus pattern occurs on the selected bus Set the trigger occurs and qualification NOTE This online help supports many oscilloscope models from Tektronix This feature is only availab...

Page 869: ...he envelope shape of the signal to the trigger The high frequency burst data is seen as a single pulse The envelope function works with edge glitch pulse width or timeout trigger modes The envelope fu...

Page 870: ...ommon across all input channels Changing the trigger level of one channel changes the trigger levels of all input channels to the same level Independent Allows you to set a unique trigger level for ea...

Page 871: ...an the time specified by Width Click POS to trigger on waveforms with a positive polarity exiting the window Click NEG to trigger on waveforms with a negative polarity exiting the window Click EITHER...

Page 872: ...stics reset with each new single sequence acquisition On a reference waveform statistics are derived from all frames in that reference waveform when FastFrame is on If FastFrame is off the statistics...

Page 873: ...rol window before copying items to the clipboard You can also use the Edit menu to clear data and undo the last Autoset File menu Use the File menu for basic file operations such as saving recalling a...

Page 874: ...ssible from the menu bar only None of the utilities are directly accessible from the front panel controls or from the toolbar Vertical menu Use the Vertical menu options to set vertical parameters for...

Page 875: ...a value with the on screen pop up keypad Use an attached keyboard to type a value directly into the position field What do you want to do next Learn more about offsetting channel waveforms Learn more...

Page 876: ...ront panel for the selected waveform Enter a value with the on screen pop up keypad Use an attached keyboard to type a value directly into the position field What do you want to do next Learn more abo...

Page 877: ...instrument triggers Bus selection For the supported instruments use the control to select the Bus on which to trigger from the list of bus setups Bus time condition For the supported instruments click...

Page 878: ...specific parallel bus value When you select a serial bus and click the Edit button use the Pattern Editor controls to set up the bus trigger pattern for a serial bus value For example the Data SPI bus...

Page 879: ...d to define a pattern Learn about bus setups Bus pattern editor Bus tab Use the Pattern Editor controls to set up the bus trigger pattern for the instrument as follows Oscilloscope reference DPO70000S...

Page 880: ...For parallel bus Select the parallel bus and click Edit button Select a bus and click Edit button Oscilloscope reference 848 DPO70000SX MSO DPO70000DX MSO DPO70000C DPO7000C and MSO DPO5000B Series...

Page 881: ...type of bus selected address data hub or port For example for data bus B1 the field would display B1Data Select a bus I2C with Trigger On set to Address or Addr Data or USB with Trigger On set to Toke...

Page 882: ...he controls to set the bus trigger pattern of the channels the instrument uses as logic waveforms to detect a parallel or serial bus value Click a channel button to select the value for that channel 0...

Page 883: ...er pattern of the channels the instrument uses as logic waveforms to detect a parallel or serial bus value Click a channel button to select the value for that channel 0 1 or X For convenience you can...

Page 884: ...ou can change the format of the bus pattern after you define a pattern Hex causes the instrument to display the bus pattern as a hexadecimal value Binary causes the instrument to display the bus patte...

Page 885: ...attern or the Logic State trigger Click the Logic Thresholds button to access the Threshold Setup dialog box Use the controls to set up the Threshold Setup parameters for a Pattern or a State trigger...

Page 886: ...Threshold Setup dialog box for a 4 bit parallel bus This is an example of the Threshold Setup dialog box for an I2C serial bus Oscilloscope reference 854 DPO70000SX MSO DPO70000DX MSO DPO70000C DPO700...

Page 887: ...Threshold Setup dialog box for an SPI serial bus This is an example of the Threshold Setup dialog box for an RS232 serial bus Oscilloscope reference DPO70000SX MSO DPO70000DX MSO DPO70000C DPO7000C an...

Page 888: ...old Setup dialog box for an USB serial bus What do you want to do next Learn about the Global Threshold field Learn about bus setups Oscilloscope reference 856 DPO70000SX MSO DPO70000DX MSO DPO70000C...

Page 889: ...play waveforms with AC and DC components Select AC to display waveforms with AC components Select GND to display a zero volt waveform Use this selection to establish the ground reference point on the...

Page 890: ...y displayed channel waveform can be turned off without impacting any math functions measurements or histograms done on that waveform You may want to turn some waveform displays off to make the remaini...

Page 891: ...ontrol that affects all waveforms including math and reference waveforms What do you want to do next Continue to learn about the Vertical Setup controls Set the vertical position scale From the Vertic...

Page 892: ...nterface map for the location of the readouts What do you want to do next Continue to learn about the Vertical Setup controls Multipurpose knobs The multipurpose knobs can be used to set screen interf...

Page 893: ...ay the Probe Setup control windows Use Setup to configure the attached probe type The setup will vary depending on the individual probe you connect Use Cal to check the status of the attached probe an...

Page 894: ...use with high impedance passive probes Select 50 for use with most active probes and low impedance Zo probes NOTE Termination is automatically set for TPP0500 and TPP1000 Series probes on the MSO DPO...

Page 895: ...that DC loading degrades amplitude accuracy the lower input capacitance reduces high frequency loading to the circuit under test That makes low input impedance passive probes ideal for timing and phas...

Page 896: ...720p 50 720p 60 875i 60 1080 24sF 1080p 24 1080p 25 1080p 30 1080p 50 1080p 60 Specify the video trigger holdoff 1 Click in the Holdoff field 2 Specify Holdoff By Time or By Fields 3 Use the multipurp...

Page 897: ...g Logic Waveforms MSO Series Using Waveform Math Using Spectral Analysis Characterizing a Signal Adding Application Specific Measurements Windows environment Connecting to a Network Printing a Hard Co...

Page 898: ...Oscilloscope reference 866 DPO70000SX MSO DPO70000DX MSO DPO70000C DPO7000C and MSO DPO5000B Series...

Page 899: ...4 576 653 744 Add sources parallel bus 149 Adjust 246 648 684 Advanced trigger button 76 Aliasing recognizing 737 Amplitude measurements 91 272 744 Analog oscilloscopes 65 174 177 642 signal 186 simul...

Page 900: ...bus and clock polarity 153 parallel bus and clock source 152 parallel bus and remove a channel 152 parallel bus setup 147 148 parallel bus source 149 serial bus channel threshold 140 serial bus data r...

Page 901: ...OUT 81 COM 1 81 COM 2 81 Connectors 12 5 GHz clock 81 DisplayPort 81 DVI I 81 EXT REF 81 front panel 69 GPIB 81 LAN 81 parallel port 81 PS 2 keyboard 81 REF OUT 81 serial port 81 side panel 81 signal...

Page 902: ...de 380 trigger mode holdoff 474 trigger setup 377 user preferences keypad defaults 559 user preferences prompt before action 559 vertical acquisition 654 vertical offset 487 vertical setup 479 wavefor...

Page 903: ...774 Delay measurement setup 808 Delay mode 185 768 769 Delete setups 530 533 waveforms 530 533 Deskew control window 505 interaction with fast acquisition 191 505 procedure 45 576 861 Diagnostics 554...

Page 904: ...522 copy setup 517 518 520 select for copy 522 Undo last autoset 189 Edit MyScope 375 Editing a user mask 757 Editing MyScope control windows 592 Effective number of bits enhanced 190 Eliminating ali...

Page 905: ...features 255 locking waveforms 830 magnitude waveforms 610 828 832 phase waveforms 828 830 process 255 window characteristics 258 windows 257 821 File 742 File menu options display on off 484 exit 64...

Page 906: ...ndow defined 733 Hardware acquisition 659 hardware applications adding 645 removing 645 harmonics 737 Help about tekscope 21 Hi Res mode 770 817 high input impedance probes 74 763 862 863 High low met...

Page 907: ...76 Initializing the oscilloscope 45 Ink saver 26 Input resistance 653 termination 653 Input amplifier 659 Input channel acquisition hardware 659 conditioning 652 trigger sources 664 input connector pr...

Page 908: ...ration autofit controls 223 autoset controls 222 display control 220 display controls 221 Mask editing setup window 225 Mask hits per segment 235 Mask margin controls 226 Mask pass fail define source...

Page 909: ...s 715 custom functions 714 plugins 248 maximum input voltage 35 Maximum signal level 653 Mean 67 Measure amplitude 91 Measurement annotation 561 gating 597 localizing 597 phase 285 snapshot 269 Measur...

Page 910: ...639 horizontal 92 93 horizontal parameters 92 interaction with fast acquisition 191 list 266 797 mean 287 miscellaneous 282 reference levels 288 standard deviation 287 statistics 287 statistics contro...

Page 911: ...cquisition 148 setup for an unclocked acquisition 147 trigger 362 383 Parallel bus options 147 149 152 153 Pass fail mask test 227 Pattern editor for a bus Bus setup control window 846 Pattern editor...

Page 912: ...sation procedure passive probes 576 Probe controls 861 Probe setup compensation 502 control window 499 status 511 tip selection window 504 Probe Setup Control Window 497 Probe tip selection 504 Probes...

Page 913: ...al bus trigger 316 RS 232 Serial Bus search for data 318 search for parity error 318 search for start 318 Run stop button 78 Run stop control window 192 Runt pulse triggering 625 Runt trigger control...

Page 914: ...L 1553 trigger 350 MIL STD 1553 trigger 350 RS 232 trigger 316 SPI trigger 319 338 340 343 346 USB trigger 325 Serial bus channel threshold setup 140 Serial bus data rate setup 138 Serial bus input se...

Page 915: ...me 716 Spectral math waveforms example 614 frequency domain 716 gating 716 overview 716 set up 611 time domain 716 Spectral setups control window advanced advanced 250 basic basic 249 frequency 251 ga...

Page 916: ...attern condition MSO DPO5000B and MSO70000C DX Series 845 bus pattern editor 846 bus pattern editor DPO MSO5000B Series 847 bus selection 845 bus time condition 845 codes 746 747 considerations 656 in...

Page 917: ...rigger data 377 pretrigger data 377 sequential 441 624 823 setup control window 377 shared independent 838 types 377 378 Troubleshooting autoset 659 support information 16 Windows interface 22 Turn of...

Page 918: ...2 vertical offset control window 487 Vertical controls vertical setup control window aux tab aux tab 496 Vertical menu options attenuation 507 bandwidth 493 coupling 490 deskew 505 display on off 857...

Page 919: ...59 spectral magnitude 66 255 610 830 832 spectral phase 66 255 830 WfmDB mode 770 Width trigger defined 674 Width trigger control window 437 Window bin width 720 Window threshold violations triggering...

Page 920: ...Index 888 DPO70000SX MSO DPO70000DX MSO DPO70000C DPO7000C and MSO DPO5000B Series...

Reviews: