Quick guide for Implantmed SI-1010 / SI-1015 / SI-1023

Hygiene and maintenance – Motor with cable

Start-up – General

Connect the mains cable
and foot control.

Connect motor cable.

Insert the irrigant support. Attach the universal 

support and lock it.

Open the pump cover.
Insert the irrigation tubing.

Close the pump cover.

Plug the mains cable into
a earthed power socket.

Switch on the control unit
at the power switch.

Control unit operation – Main menu

16

max. 50 Ncm

1:1

rpm

  40 000

Implantology 1

My favorites

Set program

Documentation (DOCU)

Set speed / torque

Setup

Progress display mode

Forward/reverse operation mode

Programs 1 to 5

Set coolant volume

 Pre-disinfection

>

  If heavily soiled, clean first with disinfectant cloths.

 Motor with cable

Do not twist or kink the motor cable! Do not coil it 
too tightly!

 Manual cleaning

>

  Rinse and brush off with demineralized water (< 38°C)

 Mechanical cleaning and disinfection

>

  Use only thermo washer disinfectors with a drying 

  program for preparation.

 Sterilization

W&H recommends sterilization according to EN 

13060, class B. Other sterilization methods may 

reduce the life span of the motor.

 storage

>

  Store sterile goods dust-free and dry.