![Tektronix DG2020A User Manual Download Page 231](http://html1.mh-extra.com/html/tektronix/dg2020a/dg2020a_user-manual_1077764231.webp)
Appendix B: Performance Verification
BĆ20
DG2020A, P3410, & P3420 User Manual
2. Press the START/STOP button on the front panel so that the button’s LED
indicator lights.
3. Confirm that the digital multimeter display reads 0.1 V (the low level) or
lower.
4. Press the front panel FORCE TRIGGER button.
5. Confirm that the digital multimeter display reads 4.4 V (the high level) or
higher.
6. Move the pin header cable connected to the P3410 CH0 output pin to the
CH1 pin and repeat steps 3 through 5. Confirm the output levels for CH2
through CH11 in the same manner.
This test confirms that accuracy of the delay relative to CH0 in the P3410 is
within the range
2 ns. This test is for the channels that support variable delay
times, i.e., channels CH8 to CH11.
Required Equipment
DG2020A
Pod connection cable
Oscilloscope
Two SMB-to-pin-header cables
Two SMB to BNC adapters
Performance check disk
Connections.
Connect the P3410 CH0 and CH8 data output pins to the oscillo-
scope CH1 and CH2 inputs, respectively, using the SMB-to-pin-header
cables and the SMB to BNC adapters.
Figure BĆ17: Variable delaytest connections
Variable Delay
Summary of Contents for DG2020A
Page 2: ...User Manual DG2020A Data Generator With P3410 P3420 Pods 071 0053 50 www tektronix com...
Page 5: ......
Page 15: ...Table of Contents DG2020A P3410 P3420 User Manual...
Page 19: ...Service Safety Summary DG2020A P3410 P3420 User Manual...
Page 50: ...Operating Basics DG2020A P3410 P3420 User Manual 2 15 Figure 2 10 Data structure image...
Page 69: ...Basic Menu Operation DG2020A P3410 P3420 User Manual...
Page 181: ...APPLICATION Menu DG2020A P3410 P3420 User Manual...
Page 211: ...Appendix A Performance Characteristics DG2020A P3410 P3420 User Manual...