![Tektronix DG2020A User Manual Download Page 221](http://html1.mh-extra.com/html/tektronix/dg2020a/dg2020a_user-manual_1077764221.webp)
Appendix B: Performance Verification
BĆ10
DG2020A, P3410, & P3420 User Manual
When connecting the DG2020A to a pod with a pod connection cable,
turn off the DG2020A power before connecting the cable. Connecting the cable
with the power in the on state can damage the DG2020A and the P3410 pod.
When attaching the pod cable, ensure that the plug and socket are aligned
correctly.
Make sure that you have correctly inserted the cable plug in the DG2020A and
the pod before turning on power. The yellow wire end of the connector must be
aligned with the triangular yellow index mark on the DG2020A or pod.
Incorrectly connected cables will damage the DG2020A and the pod.
For more details, refer to page 1-9.
Next, connect the pod CH0 signal output to the oscilloscope CH1 input
using an SMB-to-pin-header cable and an SMB to BNC adapter (for a P3410
pod) or an SMB-to-BNC cable (for a P3420 pod). See Figure B-5 for the
general method of connection. Also, see Figures B-6 and B-7 for detailed
views of the pod data output connectors.
"" " #"#"
"
"
"
!!
$"$$
$"$ "
$"$
Figure BĆ5: Pod connection
Summary of Contents for DG2020A
Page 2: ...User Manual DG2020A Data Generator With P3410 P3420 Pods 071 0053 50 www tektronix com...
Page 5: ......
Page 15: ...Table of Contents DG2020A P3410 P3420 User Manual...
Page 19: ...Service Safety Summary DG2020A P3410 P3420 User Manual...
Page 50: ...Operating Basics DG2020A P3410 P3420 User Manual 2 15 Figure 2 10 Data structure image...
Page 69: ...Basic Menu Operation DG2020A P3410 P3420 User Manual...
Page 181: ...APPLICATION Menu DG2020A P3410 P3420 User Manual...
Page 211: ...Appendix A Performance Characteristics DG2020A P3410 P3420 User Manual...