Appendix B: Performance Verification
DG2020A, P3410, & P3420 User Manual
BĆ19
This test confirms the pod output voltages into a 1 M
W
load.
High level
4.4 V or higher
. . .
Low level
0.1 V or lower
. . .
Required Equipment
DG2020A
Pod connection cable
BNC to dual banana adapter
SMB-to-pin-header cable
SMB to BNC adapter
1 M
W
resistor
Performance check disk
Connections.
Connect the CH0 data output pin for the pod connected to the
DG2020A to the digital multimeter input through an SMB-to-pin-header
cable, and SMB to BNC adapter, and a BNC to dual banana adapter. Connect
the 1 M
W
resistor across the BNC to dual banana adapter terminals.
Figure BĆ16: Output voltage test connections
Setup
Digital multimeter
Function
DCV
. . . . . . . .
Range
Auto
. . . . . . . . . .
Characteristics Confirmation Procedure
1. Load the TP6DCTTL.PDA test pattern file from the performance check
disk.
Output Voltage Levels