Appendix B: Performance Verification
DG2020A, P3410, & P3420 User Manual
BĆ15
6. Confirm that square waves with 1
m
s and 500 ns periods are displayed on the
oscilloscope alternately each time the front panel STEP/EVENT button is
pressed.
7. Connect the pod data output CH1 to the pod EVENT INPUT. Use an
SMB-to-pin-header cable for the P3410 pod and an SMB to SMB cable for
the P3420 pod. See Figures B-9 and B-10.
Figure BĆ9: P3410 event input connections
Figure BĆ10: P3420 event input connections
8. Confirm that a 1
m
s square wave and a 500 ns square wave are displayed on
the oscilloscope alternately for 1 cycle and 2 cycles respectively.
The following three steps confirm the event output signal operation.
9. Connect the DG2020A EVENT OUTPUT to the oscilloscope’s CH2 with a
BNC cable.