![Tektronix DG2020A User Manual Download Page 220](http://html1.mh-extra.com/html/tektronix/dg2020a/dg2020a_user-manual_1077764220.webp)
Appendix B: Performance Verification
DG2020A, P3410, & P3420 User Manual
BĆ9
Setup
Oscilloscope
Displayed channel
CH1
Vertical axis
500 mV/div
. . . . .
Horizontal axis
1
m
s/div
. . .
Trigger mode
Auto
. . . .
Trigger level
500 mV
. . . . .
Input coupling
DC
. . . .
Input impedance
50
W
. .
Function generator
Waveform
Square wave
. . . . . . .
Frequency
1 MHz
. . . . . . .
Amplitude
1 V (50
W
termination)
. . . . . . .
Offset
500 mV (50
W
termination)
. . . . . . . . . .
Characteristics Confirmation Procedure
1. Load the TP2EXCLK.PDA test pattern file from the performance check
disk.
2. Press the START/STOP button on the front panel so that the button’s LED
indicator lights.
3. Confirm that the clock pulse waveform can be observed on the oscilloscope.
This test confirms that pattern data is output from the pod data outputs (CH0
through CH11).
Required Equipment
Oscilloscope
P3410 or P3420 pod
Pod connection cable
SMB-to-pin-header cable (when a P3410 pod is used)
SMB to BNC adapter (when a P3410 pod is used)
SMB-to-BNC cable (when a P3420 pod is used)
Performance check disk
Connections.
Connect the DG2020A rear panel pattern data output-connector
(POD A) to the data input-connector on the rear panel of a P3410 or P3420
pod using a pod connection cable.
Digital Output
Summary of Contents for DG2020A
Page 2: ...User Manual DG2020A Data Generator With P3410 P3420 Pods 071 0053 50 www tektronix com...
Page 5: ......
Page 15: ...Table of Contents DG2020A P3410 P3420 User Manual...
Page 19: ...Service Safety Summary DG2020A P3410 P3420 User Manual...
Page 50: ...Operating Basics DG2020A P3410 P3420 User Manual 2 15 Figure 2 10 Data structure image...
Page 69: ...Basic Menu Operation DG2020A P3410 P3420 User Manual...
Page 181: ...APPLICATION Menu DG2020A P3410 P3420 User Manual...
Page 211: ...Appendix A Performance Characteristics DG2020A P3410 P3420 User Manual...