Incoming Inspection
1- 24
CSA7000 Series, TDS7000 Series, & TDS6000 Series Instruments User Manual
NOTE
. Do not make changes to the front-panel settings that are not called out in
the procedures. Each verification procedure will require you to set the instrument
to certain default settings before verifying functions. If you make changes to
these settings, other than those called out in the procedure, you may obtain
invalid results. In this case, redo the procedure from step
1
.
When you are instructed to push a front-panel button or screen button, the button
may already be selected (its label will be highlighted). If this is the case, it is not
necessary to push the button.
Equipment
required
One BNC cable
CSA7404, CSA7154, TDS7404, TDS7254, TDS7154, TDS6604, &
TDS6404:
One P7240 probe (P7260 probe for TDS6604)
One probe calibration and deskew fixture
Prerequisites
None
1.
Initialize the instrument:
Push the front-panel
DEFAULT SETUP
button.
2.
Hook up the signal source:
Connect the equipment as shown in Figure 1--7
to the channel input you want to test (beginning with CH 1).
3.
Turn off all channels:
If any of the front-panel channel buttons are lighted,
push those buttons to turn off the displayed channels. See Figure 1--8.
4.
Select the channel to test:
Push the channel button for the channel you are
currently testing. The button lights and the channel display comes on.
Check Vertical Operation
Summary of Contents for CSA7154
Page 4: ......
Page 22: ...Preface xviii CSA7000 Series TDS7000 Series TDS6000 Series Instruments User Manual ...
Page 42: ...Installation 1 20 CSA7000 Series TDS7000 Series TDS6000 Series Instruments User Manual ...
Page 58: ...Incoming Inspection 1 36 CSA7000 Series TDS7000 Series TDS6000 Series Instruments User Manual ...
Page 78: ...Rear Panel I O Map 2 12 CSA7000 Series TDS7000 Series TDS6000 Series Instruments User Manual ...
Page 458: ...Glossary Glossary 18 CSA7000 Series TDS7000 Series TDS6000 Series Instruments User Manual ...