Appendix A: Specifications
CSA7000 Series, TDS7000 Series, & TDS6000 Series Instruments User Manual
A- 29
Table A- 4: Trigger specifications (Cont.)
Characteristic
Description
Trigger position error, typical
Edge trigger, DC coupling, for signals having a slew rate at the trigger point of
≥
0.5
division/ns
Acquisition mode
Error
CSA7404, CSA7154, TDS7404,
TDS7254, TDS7154,
TDS6604, & TDS6404
Sample, Average
±
(1 waveform in 200 ps)
Peak Detect, Envelope
±
(2 waveform in 200 ps)
TDS7104 & TDS7054
Sample, Average
±
(1 displayed point + 1 ns)
Envelope
±
(2 displayed 1 ns)
Trigger holdoff range
250 ns to 12 s,
minimum resolution is 8 ns for settings
≤
1.2
s. A dither of
4 ns is added to the
holdoff setting
Table A- 4: Serial Trigger specifications (optional on TDS7000 Series & TDS6000 Series)
Characteristic
Description
Serial trigger number of bits
32
Serial trigger encoding types
NRZ
n
Serial trigger baud rate limits
Up to 1.25 GBd
Serial trigger, serial word recognizer position
accuracy
Edge trigger, DC coupling, for signals having a slew rate at the trigger point of
≥
0.5
division/ns
Acquisition mode
Error
Sample, Average
±
(1 waveform in 200 ps)
Peak Detect, Envelope
±
(2 waveform in 200 ps)
Clock recovery frequency range
1.5 MBd to 2.5 GBd. Above 1250 MHz the clock is only available internally as a
trigger source. Below 1250 MHz the clock is also available at the Recovered Clock
output along with regenerated data (CSA7000 Series only).
Summary of Contents for CSA7154
Page 4: ......
Page 22: ...Preface xviii CSA7000 Series TDS7000 Series TDS6000 Series Instruments User Manual ...
Page 42: ...Installation 1 20 CSA7000 Series TDS7000 Series TDS6000 Series Instruments User Manual ...
Page 58: ...Incoming Inspection 1 36 CSA7000 Series TDS7000 Series TDS6000 Series Instruments User Manual ...
Page 78: ...Rear Panel I O Map 2 12 CSA7000 Series TDS7000 Series TDS6000 Series Instruments User Manual ...
Page 458: ...Glossary Glossary 18 CSA7000 Series TDS7000 Series TDS6000 Series Instruments User Manual ...